October 10, 2023
S15LProIn-LineIndicatorWithModBus®InstructionManual
7
Chapter Contents
Holding Registers for Device Information
Address
without Offset
Address with
Offset Description Holding Register
Representation Default Value Saved Access
1000 1001 Low word model number Example:0x0002A734(hex)
=173876(dec)
Highword=0x0002
Lowword=0xA734
See Device
Yes RO
1001 1002 High word model number Yes RO
1002 1003 Model version (BCD) See Device Yes RO
1003-1018 1004-1019 Model name, string
S15L Pro
Indicator with
ModBus
Yes RO
1019 1020 Low word configuration number Example:0x00016D43(hex)
=93507(dec)
Highword=0x0002
Lowword=0xA734
See Device
Yes RO
1020 1021 High word configuration number Yes RO
1021 1022 Configuration version (BCD) See Device Yes RO
1022-1037 1023-1038 Serial number/date code, string See Device Yes RO
1038-1053 1039-1054 Serial number, string See Device Yes RO
Holding Registers to Configure Operation Mode
Address
without Offset
Address with
Offset Description Holding Register
Representation Default Value Saved Access
3200 3201 Operation Mode 0=NormalMode
1=DemoMode 0 No RW
Holding Registers to Configure Normal Mode
Address
without Offset
Address with
Offset Description Holding Register
Representation Default Value Saved Access
3300 3301 Animation
0=Off
1=Steady
2=Flash
3=TwoColorFlash
4=Reserved
5=Reserved
6=Reserved
7=IntensitySweep
8=TwoColorSweep
0 No RW
3301 3302 Reserved 0 0 No RW
3302 3303 Pulse Pattern
0=Normal
1=Strobe
2=ThreePulse
3=SOS
4=Random
0 No RW
3303 3304 Speed
0=Slow
1=Medium
2=Fast
3=CustomFlashRate
0 No RW
3304 3305 Reserved 0 0 No RW
3305 3306 Reserved 0 0 No RW
3306 3307 Reserved 0 0 No RW
Continued on page 8
ModBus® Register Map
© Banner Engineering Corp. All rights reserved.