Bamo BAMOPHAR 107 User manual

BAMOPHAR 107
pH/ mV monitors
USER MANUAL
MES
23-11-2018 M-107.01-EN-AE 107-01/1
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

SUMMARY
1. 3DESCRIPTION
2. 3TECHNICAL FEATURES
3. 3DIMENSIONS
4. 4ELECTRICAL CONNECTIONS
4.1 4VIEWS OF TERMINALS
4.2 5TERMINAL STRIP ASSIGNMENTS
4.3 6CONNECTION OF pH CABLE 9060 TO WALL MOUNT BAMOPHAR
4.3.1 69060 coaxial cable preparation
4.3.2 7Connection of pH cable 9060 to terminal of BAMOPHAR
5. 8MAIN DISPLAY
5.1 8WITHOUT EXTENSION UNIT
5.2 8WITH AN EXTENSION UNIT
6. 9DESCRIPTION OF ICONS
7. 9DISPLAY SETTINGS
7.1 9SCREEN INFORMATION
7.2 9SCREEN SAVER
7.3 9LANGUAGE SELECTION
7.4 10DESIGNATION
8. 10CONSULTATION/ MODIFICATION
9. 10pH/ ORP SETTINGS
9.1 10ELECTRODE CALIBRATION
9.2 11SETTING THE RELAY OUTPUTS S1, S2 AND S3 AS THRESHOLDS
9.2.1 11COMMENTS ABOUT REGULATION WITH BAMOPHAR
9.2.2 11ADJUSTMENT OF THRESHOLDS S1, S2 AND S3 IN MODE ON/OFF REGULATION
9.2.3 11ASSIGNMENT OF S3 THRESHOLD TO EXTERNAL SIGNAL (E1)
9.3 12SETTING OF TEMPERATURE
9.4 12SETTINGS OF ANALOGUE OUTPUT FOR MEASUREMENT
9.5 12SETTINGS OF ANALOGUE OUTPUT FOR TEMPERATURE
9.6 12FORCED RELAY
9.7 13SETTING OF THE RELAY REGULATION MODE
9.8 14OUTPUT S4 SETTINGS
9.8.1 14DEFAULT SETTING
9.8.2 14ADJUST ALARM
9.8.3 14CLEANING PROBE
9.9 14PID REGULATION
9.10 15TYPE OF INSTRUMENT
9.11 15FORCED MEASURE
MES
23-11-2018 M-107.01-EN-AE 107-01/2
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

1. DESCRIPTION
BAMOPHAR 107 is designed for pH and ORP measurements. It is equipped with a color touch screen for the display of a multilingual menu
friendly and intuitive. The reading is easy at the same time for measurement, temperature and relay status. It displays a menu with all
parameters for configuration of analogue outputs, thresholds and regulation mode. In order to facilitate its commissioning, a programming
menu can simulate the measurement, acting on themeasurement analog outputs and P.I.D, as well as on the thresholds.
2. TECHNICAL FEATURES
End-user interface Color touch screen 4.3'', resolution 480x272 pixels
Display of measurements, menus, temperature, relay status
Configuration: password protected
Measuring scales 0... 14 pH or -1000 ... +1000 mV
Accuracy ± 0.03 pH or ± 3 mV
Input impedance > 10¹³ Ω
Sensor signal connection For coaxial connector (9054) on panel and rail mount devices
Screw terminals on wall mount instrument
Temperature compensation Automatic, input for sensor Pt 100 Ω at 0 °C, range 0 to 100 °C
Manual, between 0 and 100 °C, by configuration
Relay outputs 4 contacts N.O., voltage free
Configurable thresholds 2 independent thresholds, to set up for measurement or temperature
1 independant threshold to set up for measurement or temperature, or for an external signal (E1)
1 threshold for alarm (too long injection, measurement out of range, faulty Pt100 sensor or
cleaning mode)
Contact Initial resistance 100 mΩ max. (voltage drop 6 V DC 1 A)
Switching power 831 VA AC / 3 A / 277 V AC
90 W / 3 A / 30 V DC
Switching capacity (min.) 100 mA, 5 V DC (depending of frequency, ambient conditions)
ON/ OFF Regulation Adjustable cycle time from 0 to 9999 s,
High and low proportional bandwidths, high and low dead zones
PID Regulation Adjustable proportionality from 0 to 200%; Integrant and Derivative: 0 to 999 s
Calibration routine Relay outputs inhibited; Analogue outputs on standby at latest values
Auto-cleaning menu Settings of frequency and duration; Relay outputs inhibited, analogue outputs on standby at latest
values
Measurement output 0/4 ... 20 mA (max. 600 Ω) proportional to measurement value of pH or ORP
Temperature output or PID 0/4 ... 20 mA (max. 600 Ω), scaling within 0 … 100°C
Program testing Simulation through the menu on measurement output, temperature, PID and thresholds
Main power supply 230 V AC - 50/60 Hz - consumption 10 VA (other supplies on request)
Models Panel mount, 96x144 mm; Front IP 65, rear IP 40
Wall mount, IP65; Cable glands
Temperature limits Operating: -5 ... +50 °C; Storage: -10 ... +70 °C
OPTION (RS 422 + data logger)
Interface RS422 output, J-BUS link; Binary slave mode; 2400 to 9600 bauds
Data Logger Record of cycle average measurement; 150 000 records max. on memory card, reading on
external driver.
3. DIMENSIONS
MES
23-11-2018 M-107.01-EN-AE 107-01/3
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

4. ELECTRICAL CONNECTIONS
4.1 VIEWS OF TERMINALS
SIGNAL INPUT
(for coaxial connector 9054)
PANEL MOUNTING MODEL
WALL MOUNTING MODEL
OPTION: RS422 AND DATA LOGGER
Wall mounting model: Access to memory card
by opening the top cover
Measurement input
Screw connector for pH cable type 9060
Note: Signal input on screw connector for 9060 cable type.
(Only for wall mount models - codes 107520, 107521, 107523 and 107524)
MES
23-11-2018 M-107.01-EN-AE 107-01/4
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

4.2 TERMINAL STRIP ASSIGNMENTS
Description Terminal Connection
Output [mA], pH or Redox measurement 1+ mA
2- mA
Output [mA] of temperature or PID 3- mA / PID
4+ mA / PID
Input E1 (external contact) 5+ 24V
Whith voltage occurs between terminals 5 & 6: regulation is set on stand-by (see § 9.2.1) 60V
Power supply to 4-20 mA sensor (2-wire) 70V
8+ 24V
Input for Pt 100 sensor (2- or 3-wire)
9+
10 +
11 -
12 Shielding
Connections for an extension terminal (blind unit)
13 Y
14 Z
15 B
16 A
Relay S4 (N.O. contact) for Fault alarm/ Regulation/ Cleaning routine 17 S4
18
Relay S2 (N.O. contact) for measurement/ temperature 19 S2
20
Relay S1 (N.O. contact) for measurement/ temperature 21 S1
22
Relay S3 (N.O. contact) for measurement or temperature 23 S3
May be assigned to the external signal: becomes NC contact (see § 9.2.3) 24
25
Main power supply
26 PE = Earth (equipotential)
27 N = Neutral
28 L = Live
Example of connecting on E1 input a flow
controller, NPN type, Normally closed
(code 194831)
Brown (shunt)
Black
Blue
Brown
Brown (+)
Blue (-)
Black (output)
Inductive detector
NPN type, normally closed
Connection of an extension unit
Main unit Extension unit
Cable length max.. 500 m
Cable type Network cable or
4-wire shielded cable, ≥ 0,25 mm²
MES
23-11-2018 M-107.01-EN-AE 107-01/5
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

4.3 CONNECTION OF pH CABLE 9060 TO WALL MOUNT BAMOPHAR
Only for wall mount models - codes 107520/ 21/ 22/ 23/ 24
Measuring signal is safely transmitted through a special low noise cable (Ref. 9060)
The insulation must be perfect between the core and the shield. Humidity traces are prohibited.
IMPORTANT: a defect on insulation or a short-circuit will be displayed as pH 7.
4.3.1 9060 coaxial cable preparation
1. 4.
2.
5.
6.
3.
Remove the outer insulation by 20 mm.
Fold back the shield (copper mesh)
Remove the 2nd sheath (black layer) over 20 mm.
Strip the core of the coaxial cable for 10 mm
Recommended: Crimp a wiring tip (1 mm², length 8 mm,
red color)
Shorten the tip (length of 4 mm is necessary).
MES
23-11-2018 M-107.01-EN-AE 107-01/6
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

4.3.2 Connection of pH cable 9060 to terminal of BAMOPHAR
1°) Remove the BAMOPHAR lower cover to access to the terminal block.
2°) Remove the lid of the terminal, sliding it downwards or upwards.
Coaxial cable: Core terminal
Clamp for copper shieldLid
3°) Pass the prepared cable through the cable gland.
4°) Insert the tip in the screw connector
5°) Be sure that the electrical contact between the copper shield and the 2 brackets will be perfect.
6°) Press on the cable to insert it properly in the brackets.
7°) Screw on tightly to block the tip in its connector.
8°) Replace the lid of the terminal in its initial position.
10°) Fit back the lid of the BAMOPHAR.
9°) Tightly screw on the cable gland to assure a watertight connection.
MES
23-11-2018 M-107.01-EN-AE 107-01/7
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

5. MAIN DISPLAY
5.1 WITHOUT EXTENSION UNIT
Switching status of relays S1, S2, S3, S4 Measurements display
Parameter settings of BAMOPHAR
Display settings Measurements details
Designation - Identification of your
instrument
Date and time
(Opened or Closed)
(Only for BAMOPHAR with data
logger)
pH (or ORP) and temperature values
Status of E1
Actuated =
Closed; Not active
= open
5.2 WITH AN EXTENSION UNIT
Switching status of relays S1, S2, S3, S4
Designations - Identifications of your
instruments
Measurements display
Parameter settings of both BAMOPHAR
Display settings
Date and time
Measurements details
(Opened or closed)
Main A is shown in yellow
Extension unit B is shown in green
(Only for BAMOPHAR with data
logger)
pH (or ORP) and temperature values for main and extension units (e.g. flow rate)
MES
23-11-2018 M-107.01-EN-AE 107-01/8
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

6. DESCRIPTION OF ICONS
The touch screen allows you to navigate within the menus and set up the display according your application.
To each yellow icon dedicated to main unit A corresponds a green icon dedicated to the extension unit B.
HOME
DISPLAY SETTINGS
Access to parameters of display (language and identification of the monitor)
Return to main display
Language selection
LANGUAGES
MENU
Access to parameter settings of the monitor
INFORMATION
Access to serial number and version of your BAMOPHAR
PADLOCK
Open: means MODIFICATION mode
RETURN
To go back to the previous screen
ARROWS
Cursors to navigate within the menus
SELECTION
Scroll through the selection list
CONFIRMATION
Confirm setting and go to the next parameter
SAVE
To save the parameters
Closed: means CONSULTATION (review) mode
e.g. Green icon for extension unit
7. DISPLAY SETTINGS
Note : To access to the display settings, BAMOPHAR must be in MODIFICATION mode (See § CONSULTATION/
MODIFICATION).
7.1 SCREEN INFORMATION
Screen identification number and its version are available in this menu.
7.2 SCREEN SAVER
Screen saver brightness is adjusted by moving the cursor: decreased (to the left) or increased (to the right).
7.3 LANGUAGE SELECTION
Select the flag according to your choice.
The display automatically returns to previous view.
Press icon "HOME" to go back to the main display.
MES
23-11-2018 M-107.01-EN-AE 107-01/9
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL

7.4 DESIGNATION
it is recommended to name each channel (main unit A and Extension unit B) for an easiest reading:
1) Press on icon A or B according to the instrument you wish to rename.
2°) A keyboard appears, enter the new name.
3°) Save the new designation.
8. CONSULTATION/ MODIFICATION
CONSULTATION mode let the operator to check out all working parameters.
This mode is represented by the closed padlock icon.
To change the settings of the BAMOPHAR, you need to enter in the MODIFICATION mode.
This mode is protected by a password identical to the last 4 digits of the serial number.
On the main display, press MENU icon.
Press the padlock icon and type the 4 last digits of Serial Number.
To validate it, press "OK"; Device is now in MODIFICATION mode (padlock is open).
If the entered keyword is wrong, BAMOPHAR stays in the mode CONSULTATION.
After 30 minutes the mode MODIFICATION switches back to CONSULTATION mode.
Where can we find the serial number?
The serial number (SN) is written on the identification label of the device.
It appears as well in MENU, icon " i " (INFORMATION).
9. pH/ ORP SETTINGS
9.1 ELECTRODE CALIBRATION
For calibration routine, temperature displayed on main menu should be close to the real temperature of buffers.
• To avoid the mixing of buffer solutions, rinse the pH electrode with water each time you change the buffer.
• Never wipe the electrode.
• For a correct calibration, choose the buffer value in relation with the process (e.g. buffer pH 4.00 for a process with acids).
Press MENU and choose ADJUST ELECTRODE.
STANDARD 7.00 pH Plunge the tip of the electrode in the pH 7.00 buffer solution (Ref 9011), then confirm.
Do not forget to remove the electrode protection (translucent cap).
ASYM. +00.00pH Leave the solution for about 5 minutes in order to stabilize the pH 7 measurement, then confirm.
If the measurement for ASYM. (asymmetry) is too high, a message ERROR appears:
- Verify that the buffer in use is the right one.
- Verify all connections and cable.
- In case the trouble persists, change the electrode.
BUFFER 4 pH or 10pH Once the pH measurement with buffer 7.00 is correct, rinse the electrode and plunge the
electrode tip in other buffer solution, or pH 4.00 (Reference 9012) or pH 10 (reference 9013).
Type the pH buffer value, then confirm.
SLOPE 100.0 % Wait a moment for the slope measurement to stabilize, then confirm.
If the slope is weak (<70%), a message “ERROR“ appears.
In such a case, if the buffer is the right one and not obsolete, change the electrode.
DELAY 0000 Sec When the slope is valid (70 to 110%), enter the delay while measurement and outputs stay
blocked at values, previous to calibration routine; then confirm.
SAVE ? To record the calibration parameters, press the icon SAVE.
Do not forget to set up the temperature mode to its previous setting: manual mode (fluid
temperature) or automatic mode (Pt 100 sensor connected), see the menu TEMPERATURE.
MES
23-11-2018 M-107.01-EN-AE 107-01/10
pH/ mV monitors
BAMOPHAR 107
22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL
Tel
Fax
+33 (0)1 30 25 83 20
+33 (0)1 34 10 16 05
Web
E-mail
www.bamo.eu
INTERNATIONAL