AstroAI DT201D Assembly instructions

-1-
1. INTRODUCTION
This manual provides all safety information, operation
instruction, specifications and maintenance for the meter,
which is compact, handheld, and battery operated.
This instrument performs AC/DC voltage, AC Current,
Resistance, Audible Continuity, Diode and Temperature
measurements. It is a 3 5/6 digits, 6000 counts auto ranging
digital clamp multimeter.
DT201D digital clamp multimeter has been designed
according to EN61010-1 oncoming electronic measuring
instruments with an over voltage category (CAT III 600V
CAT II 1000V) and Pollution degree 2.
Warning
To avoid possible electric shock or personal injury, and
to avoid possible damage to the Meter or to the equipment
under test, adhere to the following rules:
Before using the Meter inspect the case. Do not use
the Meter if it is damaged or the case (or part of the
case) is removed. Look for cracks or missing plastic.
Pay attention to the insulation around the connectors.
Inspect the test leads for damaged insulation or
exposed metal. Check the test leads for continuity.
Do not apply more than the rated voltage, as marked
on the Meter, between the terminals or between any
terminal and grounding.
The rotary switch should be placed in the right

-2-
position and no any changeover of range shall be
made during measurement is conducted to prevent
damage of the Meter.
When the Meter working at an effective voltage over
60V in DC or 30V rms in AC, special care should be
taken for there is danger of electric shock.
Use the proper terminals, function, and range for your
measurements.
Do not use or store the Meter in an environment of
high temperature, humidity, explosive, inflammable
and strong magnetic field. The performance of the
Meter may deteriorate after dampened.
When using the test leads, keep your fingers behind
the finger guards.
Disconnect circuit power and discharge all
high-voltage capacitors before testing resistance,
continuity or diodes.
Replace the battery as soon as the battery indicator
appears. With a low battery, the Meter might
produce false readings that can lead to electric shock
and personal injury.
Remove the connection between the testing leads
and the circuit being tested, and turn the Meter power
off before opening the Meter case.
When servicing the Meter, use only the same model
number or identical electrical specifications
replacement parts.
The internal circuit of the Meter shall not be altered at
will to avoid damage of the Meter and any accident.

-3-
Soft cloth and mild detergent should be used to clean
the surface of the Meter when servicing. No abrasive
and solvent should be used to prevent the surface of
the Meter from corrosion, damage and accident.
The Meter is suitable for indoor use.
Turn the Meter power off when it is not in use and
take out the battery when not using for a long time.
Constantly check the battery as it may leak when it
has been using for some time, replace the battery as
soon as leaking appears. A leaking battery will
damage the Meter.
2. GENERAL CHARACTERISTICS
Display : LCD, 6000 counts updates 2/sec
LCD size : 30*49 mm
Polarity Indication : “-” displayed automatically
Over-range Indication : “OL” displayed
Low Battery Indication : “ ” displayed
Range select : auto range or manual
True-RMS : ACV ACAmeasure
Operation Temperature : 0°C to 40°C, less than 80%RH
Storage Temperature : -10°C to 50°C, less than 85%RH
Battery Type : 9V NEOA 1604,6F22equivalent

-4-
Dimension(H×W×D) : 245×66×45mm
Clamp open jaw : Max. 42mm
Weight : Approx 334g
3. ELECTRICAL SYMBOLS
DC (Direct Current).
AC (Alternating Current).
DC or AC
Important safety information.
Refer to the manual
Dangerous voltage maybe present.
Earth ground
Low battery
Diode
Continuity test
°C Centigrade
AUTO Auto range
Conforms to European Union directive

-5-
Double insulated
4. PANEL DESCRIPTION

-6-
1、Transformer Jaws
Pick up the AC Current flowing through the conductor.
2、Trigger
Press the level to open the transformer jaws when the finger
press on the level is released the jaws will close again.
3、Hz/DUTY BUTTON
Push this button to select Hz or DUTY CYCLE
measurement, at Hz positing. at AC curent position,can
Measure frequency .
4、RANGE BUTTON
AC/DC voltage, AC current and Resistance measuring
ranges can be selected manually or automatically by
pushing the range control button. Push this button as
follows to choose range control mode and needed ranges.
5、Display
3 5/6 digit LCD, with a max. reading of 5999
6、“ ” Input Connect
High input for all except voltage , measurement will accept
banana plugs.
7、COM Input Jack
Low input for all voltage, resistance, and continuity etc
measurement will accept banana plugs.
8、“ ”RANGE
resistance,continuty,capacitance,diode , Push selcet button

-7-
to choose needed ranges.
9、SELECTING BUTTON
Push this button to select Ω, , or measuring
function when the function switch is set at Ω range.
10、Function Switch / Range Switch
This switch can be used to select desired function and
range.
11、D.HOLD BUTTON
When this button is pushed, LCD will show the last reading,
and “H” symbol will appear till pushed again. Back light
function, push this button for more than 3 seconds, back
light will light ,after 15 seconds ,back light will turn off
automatically。

-8-
1、Duty Cycle is selected
2、Fahrenheit temperature test is selected
3、Celsius degree test is selected
4、Continuity test is selected
5、Diode test is selected
6、Autorange mode is selected
7、Automatic power-off mode is selected
8、high voltage wanning symbol
9、Data Hold is enabled
10、MAX - Maximum reading is being displayed
11 、MIN - Minimum reading is being displayed
12、DC
13、Negative sign
14、AC
15、Low Battery and replaced immediately

-9-
5. SPECIFICATIONS
Accuracy is guarantied for 1 year 23°C±5°C less than
80%RH
5-1. DC VOLTAGE (Auto ranging)
Input Impedance: 10MΩ
Range
Resolution
Accuracy
600mV
0.1mV
±(0.8% of rdg + 5dgts)
6V
1mV
±(0.8% of rdg + 3dgts)
60V
10mV
600V
100mV
1000V
1V
±(1.0% of rdg + 5dgts)
Overload Protection:
600V AC RMS CAT III, 1000V AC RMS CAT II
Max. Input voltage:
600V AC RMS CAT III, 1000V AC RMS CAT II
5-2. AC CURRENT (Auto ranging)
Measuring voltage drop: 200mV
Range
Resolution
Accuracy
60A
10mA
±(2.5% of rdg + 10dgts)
600A
100mA
1000A
1A
show True-RMS value
5-3. AC VOLTAGE (Auto ranging)
Range
Resolution
Accuracy
600mV
0.1mV
±(1.2% of rdg + 8dgts)
6V
1mV
±(1.2% of rdg + 6dgts)
60V
10mV
600V
100mV
750V
1V
±(1.2% of rdg + 8dgts)