Armstrong Steam QM-3 User manual

Steam QM-3 Steam Quality Monitor
Installation and Operation Manual
IOM-245-V2.1
Armstrong InternationalArmstrong International
Parc Industriel des Hauts-Sarts, 2ème avenue 4, 4040 HERSTAL - BELGIUM
Tél: +32(0)4 240 90 90 Fax: +32 (0)4 240 40 33 armstrongarmstronginternational.com
Keep this manual with equipment
for future reference.
Certified by

Armstrong InternationalArmstrong International
Parc Industriel des Hauts-Sarts, 2ème avenue 4, 4040 HERSTAL - BELGIUM
Phone: +32(0)4 240 90 90 Fax: +32 (0)4 240 40 33 armstrongarmstronginternational.com
Armstrong
Contents
Revision History . . . . . . . . . . . . . ..1
Safety . . . . . . . . . . . . . . . . . . ..2
Abbreviations and Acronyms. . . . . . . . . .3
General Description . . . . . . . . . . . ..4
Calorimeter Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....5
Cabinet Exterior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....6
Cabinet Interior . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....7
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....8
Installation . . . . . . . . . . . . . . . . . .9
General Considerations (Site Selection). . . . . . . . . . ....9
Typical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Start-Up Procedure . . . . . . . . . . . . .14
Software Navigation. . . . . . . . . . . . . 15
Standard Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Special Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Troubleshooting . . . . . . . . . . . . . . . 19
Software update. . . . . . . . . . . . . . . 24
Component and Parts List . . . . . . . . . . 25
Product Certifications . . . . . . . . . . . . 26
Appendix One: Wiring Diagram . . . . . . .27
Appendix Two: Principle Schematic . . . . .28
Appendix Three . . . . . . . . . . . . . . . 29
MODBUS Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Limited Warranty and Remedy. . . . . . . . 30

Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 1 of 30
Revision History
Version Release Date Description of Changes
1.0 28/11/2013 Initial
1.1 22/01/2013 Temperature transmitter T (T1, T2, T3) - Design Modification
2.0 17/03/2016 Tee and Adapter - Design Modification
page 5: update of the calorimeter assembly graphic according to new design.
page 6: update of the cabinet exterior graphic according to new design.
page 7: update of the cabinet interior graphic according to new design.
page 9: update of the calorimeter assembly and cabinet exterior graphics.
correction: dimension conversion lenght 3m => 118"
page 10: update of the typical installation for vertical or side of horizontal steam line graphic. Add
comment on the cable connecting Harting Plus to sensors.
page 11: update of the typical installation for top of horizontal steam line graphic.
page 12: Text modifications for point 5, 8 and 9
Add "Drain Hose Specifications" chart.
page 17: Special Screens. Advanced Setting Menu: updates of the Screen with adding of the Serial
Mode information. Special Screens. Calibration Menu Screen: update of the data into the screen
(V230).
page 18: Add drain menu procedure.
page 25: adding List of component and parts.
page 26: update of the ETL Electrical Safety Listings logo.
page 29, Appendix Three: update data logger (MODBUS) connection (master and slave modes).
All document:
SteamLog eWon data logger not available anymore, all references deleted from document.
Yokogawa information deleted from document.
2.1 31/07/2020 page 8: Electrical parmameter - add : (±10 %)
page 9: add Indore use and Pollution Degree 2 to General Consideration (Site Selection) §

Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 2 of 30
Safety
Icon Legend
Burn hazard! Uninsulated components upstream of cabinet may be hot.
• Do not touch when unit is working.
• Allow to cool before moving or servicing unit.
Live steam will cause burns; condensate water may cause them. Skin exposure
to 140 °F (60 °C) water for only five seconds may cause a second degree burn.
This device must be installed in accordance with appropriate local,
national, and international standards, codes, and practices.
Read this manual. It contains important information.
Service must be performed by a qualified person.
Installation should always be accompanied by competent
technical assistance.
Improper installation, start-up, operation, maintenance, or service
may void warranty.
You are encouraged to contact Armstrong International or its local
sales representative for additional information.
Shock hazard! High voltages present inside equipment.
• Electrical installation must be performed by qualified personnel.
• Disconnect power before performing any electrical service.
Equipment must be disposed of according to applicable
environmental requirements.
Indicates Power On
Indicates Power Off
Indicates important information concerning potential
for personal injury or damage to equipment
Indicates electrical hazard
Indicates hot surface
Keep unit away from heat-sensitive equipment and installations.

Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 3 of 30
Abbreviations and Acronyms
Term Meaning Explanation
∆P Differential Pressure ∆P1: Difference between water column in NCG vessel and atmosphere.
∆P2: Difference between water column in condensate vessel and atmosphere.
Al Alarm Indicates an out-of-limit situation, but has no impact on operation.
Al1: Dryness below user-defined set point longer than two seconds.
Al2: Four consecutive calculations of NCGs are over the user-defined limit. This calculated
value is displayed on the main screen and updated every 30 seconds.
Al3: T1 above 257 °F (125 °C) longer than two seconds.
C Celsius
cm Centimeter
Df Default Indicates failure. Turns off power to heating element and opens EV0to drain.
Df4: T3 above 185 °F (85 °C) longer than two seconds.
Df5: No condensate from condenser in last ten minutes.
Df6: T2 above 356 °F (180 °C) longer than two seconds.
DIN Deutsches Institut für Normung eV
dP Differential Pressure
EC European Community
EEC European Electrotechnical Commission
EN European Norm
EV Electronic Valve
F Fahrenheit
gal Gallon
h Hour
imp Imperial [measure]
in. Inch
kg Kilogram
L Liter
lb(s) Pound(s)
max Maximum
min Minimum
mm Millimeter
NCG Non-Condensable Gases NCGmax is the limit of the NCG rate. Alarm 2 indicates the limit has been exceeded. Range is
0–15%. Default is 3.5%.
P Pressure P is steam pressure upstream of calibrated orifice. Modbus sends data as bar even with
imperial measure selected.
ppm Parts per Million
psi(g) Pounds per Square Inch (gauge)
Q Steam Flow
QM Quality Monitoring
R Resistance Shown as watts.
R/O Reverse Osmosis
sec(s) Second(s)
SI International System of Units
STSuperheat
T Temperature T1: Temperature after pressure reduction to atmosphere.
T2: Temperature after heating resistance.
T3: Temperature after condenser.
X Dryness Fraction (sometimes called steam
quality or moisture content)
Xmin is the lower dryness limit. Alarm 1 indicates the limit has been exceeded. Range is
0.85–0.95. Default is 0.95.

Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 4 of 30
General Description
Steam QM-3 is intended to replace manual testing of
pure steam and provide real-time data proving that steam
quality meets applicable requirements.
Advantages over manual testing are:
• Improved safety
• Ease of use
• Reduced time per test
• Reduced cost per test
• More accurate and objective results
• Ability to trend data over time
Steam QM-3 is set up and calibrated to test for parameters
defined in EN285 standard. It performs three tests:
• Calculating dryness
• Calculating superheat
• Quantifying non-condensable gases (NCG)
(NCG measurement is performed first. If it is within range,
dryness and superheat measurements will be performed.)
Data from Steam QM-3 can be recorded using a data historian with Modbus
output (see appendix three on p. 29 for connection information.)
Armstrong strongly recommends that the Steam QM-3 unit be installed in one location and
not used for checking multiple steam outlets.
It is possible to use one unit for multiple locations and it could be installed for portability at
customer's preference. Note however that:
• Calorimeter assemblies and cabinets are matched sets and are not interchangeable.
• Moving both calorimeter and cabinet to alternate locations as a unit is preferred.
• If one cabinet is connected to a different calorimeter, it must be recalibrated prior to use.
Armstrong reserves the right to make design or
specification changes without notification.
Materials of construction comply with all
standards known at the time of manufacture.

2
Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 5 of 30 Calorimeter Assembly
3
4
Steam to cabinet (flexible
hose not shown)
Connection to
steam line
7
Alternate steam line connection
(if chamber must be rotated)
Calorimeter Assembly
Note: The assembly shown below is configured
for connection to a vertical steam line.
Calorimeter assembly weighs
approximately 11 lbs (5 kg).
1 Inlet tee
2 Sanitary clamp
3 P pressure transmitter
4 Calibrated orifice (inside connection)
5 Calorimeter
6 Sanitary clamp
7 Steam line from calorimeter to
superheating chamber (not visible)
8 Superheating chamber
9 Drain (6 mm)
10 Tee
11 T1 temperature transmitter
12 T2 temperature transmitter
13 Heating Element
11
5
1
8
10
12
9
6
13

12
Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 6 of 30 Cabinet Exterior
Cabinet Exterior
RS485 con-
nection (M12)
Handle
Wheel (back)/Foot (front)
Harting
connector
Steam from
calorimeter
Condensate drain
Power On
indicator light
Navigation
Pad
Display
Electrical
connection and
power switch
(not visible)
Auxiliary
(unused)
connections
Cooling water
inlet
Cooling water
outlet
Cabinet weighs approximately
42 lbs (19 kg).

Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 7 of 30 Cabinet Interior
9
7
10
5
3
4
2
1
8
Condensate drain
Steam from
calorimeter
Cabinet Interior
6
11
8 EV2 solenoid valve
9EV0solenoid valve
10 Condenser
11 Relay board with solenoid LED
12 Main screen (consult page 6)
13 Cabinet steam inlet
14 Cabinet condensate outlet
1 NCG pressure differential sensor
2 EV1 solenoid valve
3 NCG burette
4 Condensate pressure differential sensor
5 T3 temperature sensor
6 Condensate burette
7 EV3 solenoid valve
14
13

Armstrong InternationalArmstrong International
IOM-245-V2.1
Steam QM-3
Page 8 of 30 Specifications
Specifications
Parameter Specification
Calorimeter operating temperature range (T1) 0 –150 °C (32 – 302 °F)
Maximum allowable calorimeter temperature (T1) 180 °C (356 °F)
Calorimeter operating saturated steam pressure range (P) 0,5 – 4 barg (7 – 60 psig)
Superheated steam operating temperature range (T2) 0 – 180 °C (32 – 356 °F)
Dryness fraction 0,85 – 1
Dryness fraction accuracy ± 0,01 of display
Non-condensable gases content ≤ 15%
Condensate temperature range (T3) 0 – 80 °C (32 – 176 °F)
Operating condensate temperature (T3) 65 °C (149 °F )
Estimated steam consumption 1,5 kg/h (3,3 lbs/h ) @ 3 barg (45 psig)
Estimated water consumption 15 L/h (4 gal/h ) @ 10 °C (50 °F)
Electrical 115/230 VAC (±10 %)
50/60 Hz
100 W