ARJAY ENGINEERING 2852-CAP User manual

MODELS:
2852-CAP / 2851-CAP
REMOTE MOUNTED
CAPACITANCE MONITOR
USER MANUAL (REV: 3.0)
ARJAY ENGINEERINGLTD.
2851 Brighton Road
Oakville (Toronto), Canada
L6H 6C9
TEL: ++1 (905) 829-2418
FAX: ++1 (905) 829-4701
NORTH AMERICA: 1-800-387-9487
WEBSITE: www.arjayeng.com
EMAIL: arjay@arjayeng.com

THIS PAGE INTENTIONALLY LEFT BLANK

2852-CAP, 2851-CAP Revision 3.0 Page 3 | 29
TABLE OF CONTENT
NOTICE Please read the Installation Notes (4.0) prior to locating and mounting
the instrument.
1.0 SPECIFICATION .................................................................................................... 4
2.0 USE HAZARD INFORMATION............................................................................... 7
3.0 INSTRUMENT OVERVIEW..................................................................................... 8
3.1 Features ............................................................................................................................... 9
3.2 Model Number Table ............................................................................................................ 9
4.0 INSTALLATION.................................................................................................... 10
4.1 Remote Mounted Controller Installation ............................................................................. 10
4.2 Probe Installation ................................................................................................................ 11
5.0 CONTROLLER OVERVIEW................................................................................. 13
5.1 Input / Output Terminal Specification ................................................................................. 14
5.2 Permanent Power Connection (AC Powered Models Only) ............................................... 15
5.3 Glossary Of Symbols .......................................................................................................... 1 5
6.0 STARTUP AND CALIBRATION ........................................................................... 16
6.1 Startup ................................................................................................................................ 16
6.2 Menu Flow Chart Background Information ......................................................................... 16
6.2.1 Menu Short Form ................................................................................................................ 16
6.2.2 Menu Description ................................................................................................................ 17
6.2.3 Data Entry ........................................................................................................................... 17
6.3 285X Controller Setup ........................................................................................................ 1 7
6.3.1 Setup And Network ............................................................................................................. 17
6.3.2 Relay Setup Menu .............................................................................................................. 18
6.4 285X Controller Calibration ................................................................................................ 20
6.4.1 Auto Calibration .................................................................................................................. 20
6.4.2 Manual Calibration .............................................................................................................. 20
7.0 CONTROLLER NETWORK.................................................................................. 22
7.1 MODBUS Configuration ..................................................................................................... 22
7.2 2800 Series MODBUS Register Mapping .......................................................................... 23
8.0 MAINTENANCE.................................................................................................... 24
9.0 TROUBLESHOOTING.......................................................................................... 25
10.0 FLOW CHARTS.................................................................................................... 27
TABLE OF FIGURES
FIGURE 1 – PROBE INSTALLATION............................................................................... 12
FIGURE 2 – TYPICAL INSTALLATIONS.......................................................................... 12
FIGURE 3 – CONTROLLER OVERVIEW.......................................................................... 13

2852-CAP, 2851-CAP Revision 3.0 Page 4 | 29
1.0 SPECIFICATION
Specifications are subject to change without notice
Specification Details
Power Input:
(Specify voltage at time of
Order)
12VDC +15% /-10% or 24VDC +15% /-10%, 250mA maximum
100VAC – 240VAC +/- 10%, 50/60 Hz, 150mA maximum
Note: DC input models must be supplied by Limited Energy power source. Limited
Energy means compliance with one of the following requirements:
- Class 2 circuit according to Canadian Electrical Code, Part, I, C22.1;
- Class 2 circuit according to National Electrical Code, NFPA-70;
- Limited Power Supply (LPS) according to IEC 60950-1;
- Limited-energy circuit according to IEC 61010-1.
Connections to mains supply Permanent (for AC/DC model)
User Interface:
Display & Keypad Two line LCD display, Alarm status LEDs and Keypad for select
menu or enter values
Communication Interface: Modbus (RS485)
Relays / Analog Outputs:
Relay Outputs 2 SPDT relays, Dry Contacts are 3A @ 250 VAC (Resistive),
selectable failsafe or non-failsafe, selectable high or low acting alarm,
programmable time delay: 0 – 600 seconds , 0-100% differential
(2851: Isolated)**
mA Signal Output 4-20 mA, 900 Ohms max (24VDC Power) OR 450 Ohms max
(12VDC Power).
** Optional (Note: If Isolation is ordered then relays are not available)
Instrument Performance:
Measuring Range 0 - 5000 pF (Most applications are 100pF to 1000pF)
Accuracy 0.2% of full scale pF
Resolution 0.004% at 1000pF
Calibration Linear 2 point Auto calibration or Manual calibration
Environmental:
Operating conditions Continuous
Ambient Temperature -20 °C to +55 °C controller
-40 °C to +55 °C PMC2800 pulse card in probe junction box
Process Temperature -60 °C to +200 °C Standard or Heavy Duty Teflon Probe
-60 °C to +100 °C Flexible Tefzel Probe
Maximum Process Pressure 1500 PSI / 10342 kPa
Relative humidity 0 to 95% (non-condensing)
Altitude ≤2000 m

2852-CAP, 2851-CAP Revision 3.0 Page 5 | 29
Specification Details
Installation Category II
Pollution Degree 2
Equipment mobility Fixed
Mechanical Specification: Refer to Dimensional Drawing
Protection Degree –
Controller Enclosure
Type 4 / IP65 Painted Steel (Blue)
Type 4x / IP66 Polycarbonate (Gray)
Type 4x / IP66 316 Stainless Steel
Protection Degree –
Probe Junction Box
IP65
Options on Controller
Enclosure:
Buzzer
Pilot Light OR Strobe/Beacon
Intrinsic Safety Barrier
Custom alarms
2800 Series Controller
Certification
Ordinary Locations:
• UL file number: E343390
• UL / IEC 61010-1, 2nd Edition, 2008/10/28 (Electrical Equipment
for Measurement, Control, and Laboratory Use; Part 1: General
Requirements)
• CAN/CSA-22.2 No. 61010-1, 2nd Edition, 2008/10/01, (Electrical
Equipment for Measurement, Control, and Laboratory Use; Part 1:
General Requirements)
• CE: UL / IEC 61010-1; IEC / EN 61000-4; ISO 9001:2008
Remote Level Probe
Certification
(Hazardous Locations)
North America • CSA file number: 56812
• Class – 2258-02 – Process control equipment for Hazardous
Locations
• Class – 2258-82 – Process control equipment for Hazardous
Locations – Certified to US Standards
• Ex db IIC T5 Gb
• Class I, Zone 1, AEx db IIC T5 Gb
International IECEx/ATEX *• II 2G Ex db IIC T5 Gb
• Sira 17ATEX1303X
• IECEx CSA 17.0034X
Optional Intrinsic Safety *
(Zone 0, Div 1) for North
America, IECEx/ATEX
For certifications see PMC2800/PMC2800-TMP manual
(Addendum 1) for more detail.
Note: Must order Intrinsic Safety Barrier. Model: 28aaa-bbb-c-defg
* For ATEX / IECEx / CSA / UL and EU Declaration of Conformity certificates
refer to www.arjayeng.com

2852-CAP, 2851-CAP Revision 3.0 Page 6 | 29
SPECIFIC CONDITIONS OF USE FOR HAZARDOUS LOCATION
(denoted by X after the certificate number)
1. Process temperature for the standard and heavy duty probes are -60°C to +200°C and for
flexible probe is -60°C to +100°C. It is end-user’s responsibility to ensure that the ambient
around the Capacitance Controller and ARJAY Probe does not exceed the permitted ambient of
+55°C.
2. The maximum process pressure is 1500psi/ 10342KPa.
3. Only approved certified entry devices shall be used.
4. Flameproof joints are not permitted to be repaired in the field. If the flamepath is damaged, the
enclosure is to be removed from service and replaced with a new enclosure.
Model: 28aaa-bbb-c-defg
Where:
aaa bbb c d e f g
51 LT 1 = 100-240VAC,
50/60Hz 0 = No Housing 0 = No
Li
g
h
t
0 = No Buzzer 0 = No Intrinsic Barrier
52 LS 2 = N/A 1 = Painted Steel
blue
(
4
)
1 = Red
Pilot Li
g
ht 1 = Buzzer 1 = Intrinsic Barrier
OWS 3 = 12 VDC 2 = Polycarbonate
(
4X
)
2 = Amber
Beacon
OWI 4 = 24 VDC 3 = Brushed 316SS
(
4X
)
9 = other
CAP
FCM
FCS
IFA
ILA
2851 available for these models only.
*
*
*
*
*
*

2852-CAP, 2851-CAP Revision 3.0 Page 7 | 29
2.0 USE HAZARD INFORMATION
CAUTION Indicates a potentially hazardous situation that may result in minor or
moderate injury.
WARNING Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
DANGER Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
NOTICE Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.

2852-CAP, 2851-CAP Revision 3.0 Page 8 | 29
3.0 INSTRUMENT OVERVIEW
The Arjay Model 2852-CAP & 2851-CAP Capacitance Monitor provides a proportional analog
output. The 2852-CAP includes two independent alarm relays for monitoring and control of liquids
or solids in a tank or vessel. The Model 2851-CAP provides a single isolated analog output without
relays. The designations 2852 & 2851 will be used interchangeably throughout this manual with
285X.
The Arjay system has an active capacitance probe that is inserted into the vessel. As the material
level changes around the probe, the capacitance reading of the probe increases or decreases. The
controller % output and relays are calibrated against two user defined dielectrics or product phases.
The remote mounted Arjay 285X controller monitors the capacitance change of the probe and
activates the relay contacts for use with alarms, pumps, valves, etc. The analog output and RS-485
Modbus communication are available for control interface.
The complete 285X-CAP Series system consists of the probe, the PMC card and 2852 controller.
The standard probe is constructed of a Teflon coated probe with stainless steel rod. The probe
length is customer specified for the tank and depth of material to be measured. Probes can be
inserted vertically or angled into the tank. The importance is that the probe must be parallel to the
tank wall when inserted for best linearity. A concentric shielded probe is available as an option to
improve linearity and achieve a parallel ground reference. The PMC card is mounted in a junction
box at the top of the probe. The PMC card translates the capacitance signal from the probe into a
frequency pulse, which can be transmitted up to one kilometer to the 285X controller via 2-wire
shielded cable.
Calibration requires a known change of product in the vessel to provide two calibration points. The
Arjay processor will automatically extrapolate to determine the 0 to 100% span.
The two relays (Model 2852-CAP only) can activate alarm at different setpoints and are entered as
a % value through the keypad and display. Each relay also has a differential setpoint to allow the
relay to activate at one % and de-activate at a different %. This is similar to a latching format which
can cycle a pump or valve between two distinct % values using only one relay.

2852-CAP, 2851-CAP Revision 3.0 Page 9 | 29
3.1 Features
• Microprocessor based RF capacitance level transmitter
• Two Differential Alarm relays (SPDT, 3A @ 250VAC) - Model 2852-CAP only
• 0-20mA or 4-20mA field selectable output
• Modbus protocol via RS-485 for access by Arjay handheld, Central Access Panel or
compatible system
• Local 2 point Auto or Manual calibration or remote calibration via network
• User specified custom features may be added by contacting Arjay Engineering Ltd.
3.2 Model Number Table
MODEL DESCRIPTION
2851-CAP REMOTE MOUNTED 4-20mA (ISOLATED)
2852-CAP REMOTE MOUNTED 4-20mA (NON-ISOLATED) C/W 2 ALARM RELAYS
*The designation 2852 & 2851 will be interchangeable with 285X throughout this manual.

2852-CAP, 2851-CAP Revision 3.0 Page 10 | 29
4.0 INSTALLATION
NOTICE If any damage to the instrument is found, please notify an Arjay
Engineering representative as soon as possible prior to installation.
NOTICE Qualified Personnel must undertake all installations.
WARNING EXPLOSION HAZARD – DO NOT OPEN PROBE JUNCTION BOX
WHEN ENERGIZED.
WARNING DO NOT OPEN WHEN EXPLOSIVE ATMOSPHERE IS PRESENT
WARNING If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
4.1 Remote Mounted Controller Installation
Choose the mounting location in accordance with good instrument practice. Extremes of ambient
temperature and vibration should be avoided (see specifications and installation drawing).
The 285X controller may be mounted up to one kilometer from the PMC card using a minimum 18
gauge, 2-wire SHIELDED cable (Belden 8760 or Equivalent).
Check the polarity of the + and - wiring between the controller and the PMC prior to powering on the
unit; + to + and - to - to avoid damage. Shield of cable should be connected to secondary ground
terminal marked as (at controller side only). Incoming Earth ground should be connected to
primary ground terminal marked on sub plate.
NOTICE
The controller operates in a Failsafe Mode (factory default)
This means that the relays are in an energized state during normal operation.
The N.O. relay contact will be held closed and the N.C. relay contact will be
held open during a normal condition. This will allow the relay to return to its
non-energized (shelf) state during an alarm, fault or power failure condition.
Wire accordingly.
NOTICE Maximum Conduit size for installation size is ¾” FNPT.
NOTICE To ensure proper operation and electrical safety, make sure the 285X
enclosure and the PMC junction box are electrically grounded.
CAUTION
The Internal ground screw provided in this enclosure must be used for
equipment grounding connection.
The External ground screw is provided for use only as a supplemental
connection where required (or, permitted) by local codes or authorities.
This manual suits for next models
1
Table of contents
Other ARJAY ENGINEERING Measuring Instrument manuals

ARJAY ENGINEERING
ARJAY ENGINEERING 4200-IR-4 User manual

ARJAY ENGINEERING
ARJAY ENGINEERING 2880R-LT User manual

ARJAY ENGINEERING
ARJAY ENGINEERING 4200-IG-2 User manual

ARJAY ENGINEERING
ARJAY ENGINEERING 4100-HCF2 User manual

ARJAY ENGINEERING
ARJAY ENGINEERING HYDROSENSE 4410-OCM-X User manual

ARJAY ENGINEERING
ARJAY ENGINEERING GasAlert II User manual

ARJAY ENGINEERING
ARJAY ENGINEERING HYDROSENSE 2410 User manual

ARJAY ENGINEERING
ARJAY ENGINEERING HYDROSENSE 4420-OCM User manual

ARJAY ENGINEERING
ARJAY ENGINEERING 4500-RMS User manual

ARJAY ENGINEERING
ARJAY ENGINEERING 2852-FCM User manual