Aktek TRFM 1.0 User manual

Turbine Flowmeter
User Manual
www.aktek.com.tr
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
TRFM 1.0

CONTENT
I. General........................................................................................................................1
1.1 Introduction....................................................................................................... 1
1.2 Product Features................................................................................................1
1.3 Working Principle............................................................................................. 2
1.4 Major Technical Performance........................................................................... 3
1.5 Overall Dimension............................................................................................ 5
II. Installation Requirements.......................................................................................... 7
III. Operation..................................................................................................................9
3.1 Wiring................................................................................................................9
3.1.1TB3WE Three-Wire System Electric Wiring........................................ 10
3.1.2 TB2WE Three-Wire System Electric Wiring....................................... 10
3.2 Local LCD Operation Instruction....................................................................11
3.2.1 User Menu Operation............................................................................ 11
3.2.2 User Data Setting...................................................................................11
Appendix...................................................................................................................... 15
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr

1
I. General
1.1 Introduction
TRFM 1.0 turbine flow meter consists of turbine flow sensor and display instrument
and it is made by us using foreign state-of-the-art technologies, which is an ideal
gauge for measuring of liquid flow.
The flow meter is characterized by simple structure, high precision and easy
installation and repair. The product may be used in a wide range of industries,
including oil industry, chemical industry, metallurgy, water supply, paper-making,
environment protection and food industry.
It is applicable for use in closed pipes to measure flow of liquid which will not erode
stainless steel (1Cr18Ni9Ti), 2Cr13, Al2O3 and hard alloy and is free of impurities
such as fiber and granules. If this product is used in association with display
instruments with special functions, it can be used for purpose of automatic definite
quantity control and alarming in case of excessive amount.
1.2 Product Features
1. The sensor is of hard alloy bearing thrust type, which may guarantee the precision
and improve the wear resistance performance as well.
2.Simple and firm structure, easy for installation and dismantling.
3.Wide range of measuring with very low lower flow velocity limit.
4.Small loss of pressure,fine repeat ability and high precision.
5.High resistance to electromagnetic interference and vibration.
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr

2
1.3 Working Principle
When liquid flows through the casing of sensor, the impulse of fluid will provide the
blade with a rotation moment as there is an angle between the blade of impeller and
the flow direction. The blade will rotate as the friction moment and the fluid
resistance are overcome and it will reach a stable speed when the moments are at
balance. Under certain conditions, the rotation speed of blade will be in direct
proportion to the flow velocity. Due to the magnetic conductivity of blade, when
located in the magnetic field generated by signal detector (made of permanent magnet
steel and coils), the rotating blade will cut the magnetic lines and periodically change
the flux through the coil, thereby inducing electrical impulse signals at both ends of
the coil. The induced signals, after amplified and rectified by amplifier, will form a
continuous rectangular impulse wave with certain amplitude which may be remotely
transmitted to display instrument indicating the instant flow and the cumulative flow
of fluid. Within a certain range of flow, the impulse frequency f is in direct proportion
to the instant flow of fluid flowing through the sensor, which is shown in the equation
below:
k
f
Q 3600
Wherein:
f- Impulse frequency [Hz];
k- Instrument factor of sensor [1/m3], which is given by checklist. If [1/L] is used as
the unit, the equation will be:
k
f
Q 6.3
Q- Instant flow of liquid (in operation) [m3/h];
3600- Conversion factor.
Instrument factor of each sensor will be filled out in verification certificate by the
manufacturer. The instant flow and cumulative flow will be displayed when the value
of k is loaded into associated display instrument.
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr

3
1.4 Major Technical Performance
1. Nominal drift diameter:(4~200)mm, refer to Table 1 for the basic parameters;
2. Medium temperature:(-20~80)°C; Split type (-20~120) °C;
3. Ambient temperature:(-20~70)°C;
4. Precision: ±0.5%, ±1%;
5. Detector signal transmission wiring system: three-wire voltage impulse (three-core
shield wire);
6. Power supply:
1) TB3WE Three Wire
External Power:12~24VDC/30mA(-20%~+15%),if no output can be as low as 6V
Battery:Lithium battery 3.6V(2/13Ah)/0.4mA
2)TB2WE Two Wire
12~24VDC/4-20mA(-20%~+15%)
7. Transmission distance: the distance between the sensor and the display instrument
may be as far as 1000m;
8. Local display power supply: 3.6V (Lithium battery, may be used continuously for
more than 3 years);
9. Display mode: local LCD displays instant flow and cumulative flow;
10. Output Signal
a)TB3WE Three Wire
1)Pulse Output:High level voltage amplitude≥5V, low level<0.5V
2)Three sire 4-20mA linearity correction current output(need ground wire)(load
resistance≤800Ωat 24V)
3)RS485 communication:flowmeter with RS485 interface, communication distance≤
1200mm.
b)TB2WE Two Wire
1) Two wire 4~20mA linearity correction current output(need ground wire)(load
resistance≤600Ωat 24V)
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr

4
2) Origin Pulse output:High level≥5V(power supply voltage-1V),Low level<0.5V
Table 1
TRFM1.0□□□□ □ □ □ Description
Type
TRFM1.0-
A
Flow sensor pulse output three-wire system, +12V power
supply;
TRFM1.0-
BLocal display, powered by 3.6V battery;
TRFM1.0-
C
Local display with 4 ~20mA or pulse output, powered by
24V;
TRFM1.0-
DFlow transmitter 4~20mA output, powered by 24V;
Nominal drift
diameter
4
Normal
flow
range
m3/h
0.04~0.25
Extended
flow
range
m3/h
0.04~0.4
6 0.1~0.6 0.06~0.6
10 0.2~1.2 0.15~1.5
15 0.6~6 0.4~8
20 0.8~8 0.45~9
25 1~10 0.5~10
32 1.5~15 0.75~15
40 2~20 1~20
50 4~40 2~40
65 7~70 3.5~70
80 10~100 5~100
100 20~200 10~200
125 25~250 12.5~250
150 30~300 15~300
200 80~800 40~800
Explosion protection Not marked, without explosion protection
B Explosion protection type
Precision class A Precision: Class 0.5
B Precision: Class 1.0
Turbine type A Normal flow range
B Extended flow range
Note:
Sensors with pipe diameter of DN4~DN40 are of thread connections with maximum operating pressure of
6.3Mpa.
Sensors with pipe diameter of DN50~DN200 are of flange connections with maximum operating pressure
of 2.5Mpa.
Sensors with pipe diameter of DN4 ~DN10 are provided with front and rear straight pipe sections and
filters.
Please specify when placing an order if flange connections are required for pipe diameter of DN15~DN40.
Please specify when placing an order for high pressure type and special requirements.
Drawing 3 Insertion Vortex Flowmeter Outline Size and Installation Diagram
www.aktek.com.tr
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL

5
1.5 Overall Dimension
The installation types of sensors vary according to specifications, which may be
connected either by thread or flange. The installation types are shown in Fig. 1, Fig. 2,
Fig.3, Fig. 4 and Fig. 5. The installation dimensions are shown in Table 2.
Fig. 1 Structure of DN4~DN10 sensor and installation dimension diagram
Filter 2. Front straight pipe section 3. Impeller 4. Preamplifier 5. Casing 6. Rear
straight pipe section
Fig. 2 Filter structure diagram
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr

6
Clamp ring 2. Bolts 4×14 3. Washer 4. Sealing washer 5. Steel wire
1Cr18Ni9Ti-0.8×2.5 6. Filter screen 7. Base
Fig. 3 Structure of DN15~DN40 sensor and installation dimension diagram
Casing 2.Front guide part 3.Impeller 4.Rear guide part 5.Preamplifier
Fig. 4 Structure of TRFM1.0—50~200 sensor and installation dimension diagram
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr

7
Ball bearing 2.Front guide part 3.Expansion ring 4.Casing 5.Preamplifier
6.Impeller 7.Bearing 8.Shaft
Fig. 5 Wiring diagram for connection between sensor and display instrument
II. Installation Requirements
Flow meter may be installed horizontally or vertically. In the latter case the fluid shall
be flowing from downward and fulfill the pipe to avoid bubbles; the flowing direction
of liquid shall be consistent with the direction indicated by the arrow on casing of the
sensor; as far as front and rear straight pipe sections are concerned (see Fig. 6), at
upstream there shall be front straight pipe section at least 10 times of nominal drift
diameter in length and at downstream no less than 5 times of nominal drift diameter in
length. The internal wall of pipe sections shall be smooth and clean, free of defects
such as indent, fouling and peeling. The pipe axis of the sensor shall be aligned with
that of the neighboring pipe and the washers used for connection and sealing may not
be embedded into depth of the pipe cavity; the sensors shall be kept away from
foreign electric field and magnetic field, effective shielding measures shall be taken in
case of necessity to avoid external interference.
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr

8
In order that the normal transfer of liquid will not be affected by maintenance, it is
recommended that bypass pipes be installed at position of sensor.
In case of open air installation, water proof measures shall be taken for purpose of
amplifier and plug of the sensor. The wiring between sensor and display instrument is
shown in Fig. 5.
When fluid contains impurities, filter shall be additionally installed. The number of
filter screen meshes is determined in accordance with the flow and impurity, normally
20 to 60 meshes. When fluid is mixed with free gases, gas eliminator shall be
additionally installed. The complete pipe system shall be well sealed. The user shall
fully understand the erosion nature of the measured medium to protect the sensor
from being eroded.
Fig. 6 Requirements on straight pipe section for installation of flow meter
One
90°
elbow
Two 90°
elbows
for two
planes
Concen
tric
expand
er
Control
valve
half-ope
n
Concen
tric
shrinka
ge wide
open
valve
Two 90°
elbows
for one
plane
20×D
5×D
5×D
40×D
5×D
5×D
25×D
50×D
25×D
5×D
5×D
15×D
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
www.aktek.com.tr