Accutron Ultra PC % User manual

1
Ultra PC™ % Flowmeter User Manual

2
INDICATIONS FOR USE
To be used in nitrous oxide-oxygen sedation systems for delivering to a patient
a mixture of nitrous oxide and oxygen gases with a maximum nitrous oxide
concentration of 70%.
CONTRAINDICATIONS
Contraindications for use of nitrous oxide/oxygen inhalation may include:
1. some chronic obstructive pulmonary diseases;
2. severe emotional disturbances or drug-related dependencies;
3. first trimester of pregnancy;
4. treatment with bleomycin sulfate;
5. methylenetetrahydrofolate reductase deficiency.
Whenever possible, appropriate medical specialists should be consulted before
administering analgesic/anxiolytic agents to patients with significant underlying
medical conditions (e.g., severe obstructive pulmonary disease, congestive heart
failure, sickle cell disease, acute oritis media, recent tympanic membrane graft,
acute severe head injury.1)
1American Academy of Pediatric Dentistry. Clinical Guidelines: Guidelines on Use of Nitrous Oxide for
Pediatric Dental Patients. Pediatr Dent 2009;31(6):148-151. Available at:
“http://www.aapd.org/media/policies.asp”. Accessed ebruary 23, 2010.
WARNINGS
To be used only by a professional trained in the use of nitrous oxide.
Patient should always be closely monitored during nitrous oxide use. If patient has
an adverse reaction, reduce or stop the flow of nitrous oxide as needed. The O2
flush button can be used to rapidly purge the lines of N2O. If patient does not show
signs of quick recovery, remove nasal hood and treat with pure oxygen from either
the O2resuscitator fitting or an auxiliary oxygen tank using a demand valve, oxy-
gen assisted manual resuscitator, or equivalent. Call for emergency assistance if
rapid response is not achieved.
Do not use this device for the administration of general anesthesia or as part of, or
in conjunction with, a general anesthesia administration system.
Unit is calibrated at the factory to ±5% per industry recommendations.
Verify that the correct gases are being delivered to the correct ports on the analgesia
gas machine before initial use. This can be done by shutting off the oxygen cylin-
ders in the tank room. Gases should not flow when flowmeters are turned on.

3
CAUTIONS
ederal (U.S.) law restricts this device to sale by or on order of a dentist or physician.
Do not attempt to repair, alter, or calibrate this device. Unauthorized repair, alter-
ation or misuse of this device is likely to adversely affect the performance and will
void the warranty.
Safety features contained in this notice should be routinely checked to assure
proper function. If any of these safety features are not functioning properly, contact
your dealer or Accutron and arrange for the necessary repairs before reusing the
machine.
Always use clean, dry medical gases. Introduction of moisture or other contaminants
into Accutron analgesia gas machines may result in defective operation.

4
INDICATIONS D’UTILISATION
Vous ne devez utiliser cette machine que dans des systèmes de sédation oxygène-
oxyde nitreux qui livrent au patient un mélange de gaz d’oxyde nitreux et
d’oxygène ayant une concentration maximale d’oxyde nitreux de 70%.
CONTRE-INDICATIONS
Les contre-indications sur l’inhalation d’oxyde azoteux/oxygène peuvent inclure:
1. certaines maladies respiratoires obstructives chroniques;
2. des troubles émotionnels graves ou des dépendances liées à la toxicomanie;
3. le premier trimestre d’une grossesse;
4. le traitement au sulfate bléomycinique;
5. une déficience en réductase d’acide tétrahydrofoliqueméthylène.
Dans la mesure du possible, consulter un spécialiste médical avant d’administrer
des agents analgésiques/anxiolytiques aux patients qui présentent des troubles
médicaux sous-jacents (par ex.: maladies respiratoires obstructives graves, défail-
lance cardiaque, drépanocytose, otite moyenne aiguë, greffon récent de la mem-
brane du tympan, traumatisme crânien grave aigu.1)
1American Academy of Pediatric Dentistry. Lignes directrices cliniques: Lignes directrices sur l’utilisa-
tion de l’oxyde azoteux pour les patients dentaires pédiatriques. Pediatr Dent 2009;31(6): 148-151.
Disponible sur le site Web: “http://www.aapd.org/media/policies.asp”. Accès le 23 février 2010.
AVERTISSEMENTS
Doit être utilisé uniquement par un professionnel formé dans l’utilisation de
l’oxyde azoteux.
Le patient doit toujours faire l’objet d’une surveillance étroite pendant l’utilisation
de l’acide azoteux. Si le patient présente des effets indésirables, réduire ou arrêter
le débit de l’acide azoteux, selon le besoin. Le bouton de purge O2pour peut être
utilisé pour évacuer rapidement les lignes de N2O. Si le patient n’affiche aucun
signe de récupération rapide, enlever l’enceinte nasale de hood et le traiter avec
de l’oxygène pur, soit à partir du raccord du réanimateur O2ou d’une bouteille
d’oxygène auxiliaire utilisant un détendeur, d’un réanimateur manuel assisté par de
l’oxygène ou l’équivalent. Demander de l’aide d’urgence si une réponse rapide
n’est pas obtenue.
Ne pas utiliser ce dispositif pour l’administration d’une anesthésie générale ou
comme partie ou encore en conjonction avec un système d’administration
d’anesthésie générale.
Unité est calibré à l’usine de ± 5% selon les recommandations de l’industrie.
Vérifiez que les gaz sont correctes livré dans les ports corrects sur la machine de
gaz analgésique avant la première utilisation. Cela peut être fait en fermant les
cylindres d’oxygène dans la chambre du réservoir. Les gaz ne doivant pas cir-
culer lorsque les débimétres sont actirés.

5
ATTENTION
Conformément à la loi fédérale des États-Unis, cette machine ne peut être vendue
que par un dentiste ou par un médecin ou sur leur ordonnance.
N’essayez jamais de réparer, de modifier ou de calibrer cet instrument. Toute
réparation, modification non autorisée ou mauvais usage de cet instrument nuira
considérablement à son rendement et annulera la garantie.
Vous devez vérifier couramment les caractéristiques de sécurité de cette brochure
pour assurer un bon fonctionnement. Si l’une des caractéristiques de sécurité ne
fonctionne pas bien, appelez votre revendeur ou Accutron et prenez les mesures
nécessaires pour réparer votre appareil avant de l’utiliser à nouveau.
Utilisez toujours des gaz médicaux propres, secs. Toute introduction d’humidité ou
d’autre contaminant dans un appareil d’analgésie à gaz Accutron provoquera un
mauvais fonctionnement.

6

7
Table of Contents Page
Warnings and Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
I. Upright PC Unit
Basic Mechanics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Central System Mounting Options . . . . . . . . . . . . . . . . . . . . . 9
Portable System Installations . . . . . . . . . . . . . . . . . . . . . . . . . 10
II. Cabinet Mount Unit
Basic Mechanics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Sample Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
III. Newport Portable System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
IV. Safety eatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
V. General Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
VI. Directions for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
VII. Periodic Equipment Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
VIII. Rubber Goods Attachment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
IX. Emergency Oxygen Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
X. Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
XI. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
XII. Warranty & Returned Goods Policy . . . . . . . . . . . . . . . . . . . . . . . . .26
XIII. Repair Service Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
XIV. Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
XV. Ownership Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
© 2010 Accutron, Inc.
Part No. 25097 - Rev. 05 - 11/2010

8
I. PC UPRIGHT UNITS (For Portabl & C ntral Syst ms)
60%
%
-7-
-6-
-5-
-4-
-3-
-2-
-1-
N2O
O
2
Flush
U
L
T
R
A
PC
N O
2
-7-
-8-
-9-
-10-
-6-
-5-
-4-
-3-
-2-
-1-
O2
Off
F
L
O
W
+
-
13"
5"
8"
TECHNICAL SPECIFICATIONS
Physical 5"x 13"x 8"(12, 7 x 33 x 20, 3 cm)
Weight 5.1 lbs (2, 3 kg)
Bottom mount: 1/2 – 20 Thread
Top Mount: 3/4 Diameter Hole
Gas Supply
Oxygen: 50-55 PSI at 100 LPM
N2O: 50-55 PSI at 10 LPM
Gas Fittings
Oxygen Inlet: Male DISS (CGA 124)
N2O Inlet: Male DISS (CGA 1040)
Mixed Gas Outlet: O.D. .875"(2, 2)
Oxygen Resuscitator: 1/4"I.D. Quick Disconnect
Gas Delivery
Oxygen lush: Min 20 LPM
Oxygen low: 1 - 10 LPM (Accuracy, +/- 0.5 LPM)
Oxygen Resuscitator : Min 100 LPM 100% O2
Nitrous Oxide low: 0 - 7 LPM (Accuracy +/- 0.5 LPM)
Nitrous Oxide %: 0-70%*
*50% Max Units Available
Outspout
Total low
Knob
Oxygen
lush
Nitrous Oxide
low Tube
Oxygen
low Tube
%
N2O
Knob
Top
Mounting Bracket
(3/4"Ø)
Serial #
Oxygen
Resuscitator
Connection
O2Inlet DISS
N2O Inlet
DISS
Bottom Mount
Threads 1/2 – 20
Room Air Override
Downspout

9
O
2
Flush
O
O2
Resuscitator
-7-
-6-
-5-
-4-
-3-
-1 -
-2-
-9-
-8-
-7-
-8-
-7-
-6-
-5-
-4-
-3-
-2-
-1-
-7-
-6-
-5-
-4-
-3-
-1 -
-2-
F
L
O
W
PC UPRIGHT CENTRAL GAS SYSTEM MOUNTING OPTIONS
obile Stand (Bottom ount)
Under Cabinet Slide (Back ount)
Wall Arm ounts (Back ount)
2 Gas Supply
Lines Connect
To Wall Outlet
Mobile Stand
1/2-20 Threaded Post
Choice of :
O2
Resuscitator
Mounting
Cap
9
"
Straight
Telescoping
Bi-fold
Bolt
Wall Arm

10
PC UPRIGHT 2-CYLINDER PORTABLE
2-CYLINDER PORTABLE
SYSTE ASSE BLY
1. Mount the flowmeter to the portable manifold:
a. Remove hole cap from the upper hole (use
sharp edge knife or screwdriver) at rear of
manifold and release set screw (use 5/32"
Allen Wrench) enough to remove flowmeter
mount.
b. Remove protective cap from threads of
mount and remove from manifold. Thread
flowmeter mount into hole on the bottom of
flowmeter by hand and tighten firmly using
5/32" Allen Wrench placed in hole on side
of flowmeter mount.
c. Place flowmeter on manifold with mount in
pocket.
d. Tighten set screw firmly from rear while
keeping flowmeter oriented toward front of
manifold.
e. Place hole cap into hole behind the set screw
and push it all way in.
2. Attach the hoses to the flowmeter:
a. Run the nitrous oxide hose under the mani-
fold and attach it to the nitrous oxide DISS
fitting located on the back of the flowmeter.
b. Run the oxygen hose under the manifold
and attach it to the oxygen DISS fitting
located on the back of the flowmeter.
Note: To avoid gas leaks, make certain all hose
attachments are securely tightened.
3. Connect the gas cylinders to the manifold’s
“E” yokes:
a. Connect the nitrous oxide gas cylinders to
the “E” yoke on the N2O side
b. Connect the oxygen gas cylinders to the
“E” yoke on the O2side.
c. It is recommended to use new gasket every
time tank is replaced or removed.
4. (Keep wrench in handy location. It is used to
open and close tanks).
a. Open both gas cylinders and turn off
flowmeter.
b. Check hose connections for leaks using
soap solution.
c. Tighten fittings as needed to stop leaks.
5. Verify flowmeter operations.
Note: See page 23 for attaching rubber goods,
including scavenging circuit.
Flowmeter
ount
(make sure lock
washer is still in
position prior to
installation into
flowmeter)
Manifold
Set Screw
Hole Cap
O2Hose
Stand Post
N2O Hose
E-Yoke
(1 of 2)
Mobile Base
Locking Washer
(remove protective
cap from flowmeter
mount prior to
installation into
flowmeter)
Identify the following
components in the 2-cylinder
portable system package:
• lowmeter
• Portable Manifold (with
DISS Hoses and mobile
stand
attached)
• E Cylinder Wrench (with
chain attached)
14 5/8"
17 1/4"
39 3/8"
O2
Resuscitator
200
400 600
800
1000
200
400 600
800
1000 4000
3000
2000
1000
-7-
-6-
-5-
-4-
-3-
-2-
-1-
N2O
O
2
Flush
U
L
T
R
A
PC
NO
2
-7-
-8-
-9-
-10-
-6-
-5-
-4-
-3-
-2-
-1-
O2
Off
F
L
O
W
+
-
This manual suits for next models
2
Table of contents
Other Accutron Measuring Instrument manuals