INFORMATION
No. INF03/75 Date: June 2008 Product SM3000 Multipoint Videographic Recorder
Manual IM/SM3000
SM3000
Membrane Switch Sub-Assembly Replacement
1Introduction
This information sheet describes the procedure for replacing the
SM3000 instrument's Membrane Switch sub-assembly, which
is an integral part of the Display Molding.
The Membrane Switch Sub-assembly Pack, GR2000/3710,
includes the following items:
Continued…
Withdraw the instrument
from the case
Remove tamper-proof
seal, if fitted
Unscrew the jacking screw
securing the instrument to
the case
3
1
2
Fig. 3.1 Removing the Instrument from its Case
Information. Refitting is the
reverse of removal.
2Tools Required
•Medium flat-bladed screwdriver
•No.2 Pozidriv screwdriver
•Small-bladed screwdriver or pointed instrument
3Membrane Switch Sub-Assembly Replacement –
Figs. 3.1 and 3.2
Caution. The instrument is vulnerable to
electrostatic damage. Wear an anti-static strap or dismantle
the unit on an anti-static workbench.
1) Isolate the instrument from the power supply.
2) Remove the instrument from its case – see Fig. 3.1.