VWR 10050-584 User manual

VWR Collection Manual ver 1, rel 1 20.02.2017
VWR®Thermometer
INSTRUCTION MANUAL
North American Catalog Numbers: 10050-584 / 10050-586 /
10050-588 / 10050-590 / 10050-592 / 10050-594 / 10050-596/
75784-084 / 75784-086
European Catalogue Numbers: 620-2334 / 620-2335 / 620-
2336 / 620-2337 / 620-2338 / 620-2339 / 620-2340
Version: 1
Issued: 7th March 2017

VWR Collection Manual ver 1, rel 1 20.02.2017
Legal Address of Manufacturer
United States Europe
VWR International VWR International bvba
Radnor Corporate Center Researchpark Haasrode 2020
Building One, Suite 200 Geldenaaksebaan 464
100 Matsonford Road, P.O. Box 6660 B-3001 Leuven
Radnor, PA 19087 +32 16 385011
800-932-5000 http://be.vwr.com
http://www.vwr.com
Country of origin - China
Intended use: This item is intended for general laboratory use to monitor temperature.
SPECIFICATIONS
Sampling
Rate: 30 Seconds
Range: –58.0 to 158.0°F / –50.0 to 70.0°C
Resolution: 0.1°
Accuracy: ±0.5°C (–30 to 50°C), ±1°C otherwise
Features: LCD thermometer, maximum/minimum memory display and high/low alarm
OPERATION
Memory Clear
Button: Press to clear all MIN/MAX memory of selected probe
Mode Button: Press to toggle between “MIN/MAX DISPLAY MODE” and “ALARM DISPLAY
MODE”.
Probe 1/2: Selects probe channel to display.
() Alarm
On/Off: Enables or disables the Alarm function.
Reset Button: Reset HI/LO AL set value, MIN/MAX Memories for both probes.
MIN/LO Button: In “MIN/MAX display mode” press to clear the minimum memory of selected
probe channel. In “Alarm display mode” press to increment the LO AL (Low Alarm) set
value.
MAX/HI Button: In “MIN/MAX display mode” press to clear maximum memory of selected probe
channel. In “Alarm display mode” press to increment HI AL (High Alarm) set value.
°C/°F Switch: Selects temperature unit of measure (Celsius or Fahrenheit).
MINIMUM AND MAXIMUM TEMPERATURE VALUES ARE NOT PROGRAMMABLE.

VWR Collection Manual ver 1, rel 1 20.02.2017
PROBE INSTALLATION
Plug probe sensors into jacks on side of unit, then slide probe channel switch to desired channel.
With probe sensor attached, unit displays current probe temperature & probe MIN/MAX temperatures.
With the probe sensor removed unit displays “--.-”.
PROBES
Model 10050-584 (620-2334)/10050-586 (620-2335)-- Supplied with 1 bottle probe (10050-584 (620-
2334)) or 2 bottle probes (10050-586 (620-2335)), designed for use in refrigerators and freezers.
Bottle probes are filled with a patented nontoxic glycol solution that is GRAS (Generally Recognized
As Safe) by the FDA (Food and Drug Administration) eliminating concerns about incidental contact
with food or drinking water.
The solution filled bottle simulates the temperature of other stored liquids. Velcro® and a magnetic
strip are provided to mount the bottle inside a refrigerator/freezer. The included micro-thin probe cable
permits refrigerator/freezer doors to close on it. (Do not immerse bottle probes in liquid).
Model 10050-588 (620-2336) / 10050-590 (620-2337)-- Supplied with standard plastic probe with
cable. Designed for use in air and liquids, the sensor and cable may be completely immersed.
Model 10050-592 (620-2338)– Supplied with 1 bottle probe with cable and 1 standard plastic probe
with cable.
Model 10050-594 (620-2339) / 10050-596 (620-2340)– Supplied with 1, 5ml vaccine bottle probe
(10050-594 (620-2339)) or 2, 5ml vaccine bottle probes (10050-596 (620-2340)). Designed to fit in
vaccine trays.
Velcro® and magnetic strip are supplied to mount the bottle to the inside of a refrigerator/ freezer and
to mount the display unit to the outside. Micro-cable permits refrigerator doors to close on it.
RESETTING THE UNIT
Use a pointed object to push the RESET button on the back of unit.
NOTE: HI/LO temperature alarm settings and MIN/MAX memories will be cleared after resetting the
unit.
DISPLAY MODES
The unit has two display modes: MIN/MAX DISPLAY MODE and ALARM DISPLAY MODE. To change
from one mode to the other press the MODE button once.
MIN/MAX DISPLAY MODE - is indicated by the display of current temperature and the MIN
(minimum) and MAX (maximum) temperatures. These values correspond to the probe channel
selected with the probe switch.
ALARM DISPLAY MODE - is indicated by the display of current temperature and the HI and LO alarm
settings. These values corresponds to the probe channel selected by the probe switch.
RESETTING MIN/MAX MEMORIES
There are two ways to reset the memories:

VWR Collection Manual ver 1, rel 1 20.02.2017
1.To reset both minimum and maximum memories - press MEMORY CLEAR button.
2.To reset individual minimum and maximum memories - press either the MIN or MAX button (on
back of unit) once to reset the respective memory. The other memory will remain unaffected.
NOTE: You must be in MIN/MAX Display Mode to reset the memories.
DISPLAYING °F OR °C
To display the temperature readings in Fahrenheit (°F) or Celsius (°C) slide the switch on the back of
the unit to °F or °C.
SETTING THE HIGH/LOW ALARM
1.Press the MODE button until the unit is in the ALARM DISPLAY MODE indicated by the display of
current temperature and HI and LO alarm settings.
2.To set the HIGH temperature alarm limit press the HI button on the back of unit to increment the
limit in 0.1° intervals; press and hold to increment the limit in 1° intervals.
3.To set the LOW temperature alarm limit press the LO button on the back of unit to increment in 0.1°
intervals; press and hold to increment the limit in 1° intervals.
ALARM
The unit will sound the alarm in both MIN/MAX DISPLAY MODE and ALARM DISPLAY MODE.
Note: The alarm limits cannot be changed while an alarm is active.
Slide the ALARM ON/OFF switch to ON. When the current temperature display rises above the HI set
point, or falls below the LO set point, the alarm will sound for 1 minute and issue a 3 second repeater
beep every minute thereafter. The alarm will continue to sound even if the current temperature display
returns to an “in range” condition and the LCD will flash.
ALARM OPERATION
To clear the active alarm, press the Memory Clear button or slide the Alarm switch to the OFF
position. While in ALARM DISPLAY MODE, pressing either the MIN/LO or MAX/HI button will also
clear the alarm.
Alarm conditions will remain active until cleared, even if the temperature value returns to with the
alarm limits.
The alarm limits cannot be changed while an alarm is active.
VIEWING THE TEMPERATURE ALARM LIMITS
1.While in the ALARM DISPLAY MODE, slide the PROBE switch to the desired position (1 or 2) to
display the temperature sensor readings.
2.Display will show the current temperature and low/high alarm set points for the selected probe.
ENABLE/DISABLE ALARMS
To enable the alarm function, slide the ALARM switch to the ON position.

VWR Collection Manual ver 1, rel 1 20.02.2017
When the alarm is enabled, both Probe 1 and Probe 2 alarm channels will be monitored. If either
sensor enters an alarm condition, the alarm will activate. It is not possible to enable the alarm for
one sensor and disable the alarm for the other sensor.
A disconnected probe does not trigger an alarm state.
To disable the alarm function, slide the ALARM switch to the OFF position.
ALARM SOUNDING
With the alarm enabled an alarm will sound and the LCD will flash when a temperature is measured
outside the alarm limits that have been set. Until it has been acknowledged/silenced (see the “Alarm
operation” section), the alarm will sound and the corresponding LCD set point breached will flash
continuously for the first sixty (60) seconds. After 60 seconds, the alarm will sound for three (3)
seconds every minute.
The unit will continue to alarm and flash the corresponding LCD set point until the alarm has been
acknowledged/silenced.
BENCH STAND
The unit is supplied with a bench stand that is a part of the battery cover. To use the bench stand,
locate the small rectangular opening at the bottom of the unit. Place your fingernail into the opening
and flip the stand up. To close the stand, simply snap it shut.
WALL MOUNTING THE UNIT
Set a screw into the wall at the desired location. Do not set the screw flush to the wall, the head of the
screw will need to slip into the receptacle on the back of the unit. Once the screw has been properly
set, hang the unit in place by sliding the receptacle on the back of the unit over the head of the screw.
ALL OPERATIONAL DIFFICULTIES
If this thermometer does not function properly for any reason, please replace the battery with a new
high-quality battery (see “Battery Replacement” section). Low battery power can occasionally cause
any number of “apparent” operational difficulties. Replacing the battery with a new fresh battery will
solve most difficulties.
BATTERY REPLACEMENT
Low battery symbol, erratic readings, a faint display, or no display are all indications that the battery
must be replaced. Slide the battery cover toward the end of the unit. Remove the exhausted batteries
and replace with 2 AAA batteries. Replace the battery cover.
STATIC SUPPRESSOR INSTALLATION
Static generated, radio frequency can affect any cable through the air or by physical contact. To
protect against radio frequency, install a suppressor onto your thermometers cable to absorb radio

VWR Collection Manual ver 1, rel 1 20.02.2017
frequency as follows:
1. Lay the cable along the center of the suppressor with the connector to your left.
2. Loop the right end of the cable under the suppressor and back up again laying the cable
along the center of the suppressor.
3. Carefully, snap the two halves together with the looped cable routed through the center.
4. This completes the installation of the suppressor.
Equipment disposal
This equipment is marked with the crossed out wheeled bin symbol to indicate that this equipment
must not be disposed of with unsorted waste. Instead it's your responsibility to correctly dispose of
your equipment at lifecycle -end by handling it over to an authorized facility for separate collection
and recycling. It's also your responsibility to decontaminate the equipment in case of biological,
chemical and/or radiological contamination, so as to protect from health hazards the persons involved
in the disposal and recycling of the equipment. For more information about where you can drop off
your waste of equipment, please contact your local dealer from whom you originally purchased this
equipment.
By doing so, you will help to conserve natural and environmental resources and you will ensure that
your equipment is recycled in a manner that protects human health.
Compliance with local laws and regulations
The customer is responsible for applying for and obtaining the necessary regulatory approvals or
other authorizations necessary to run or use the product in its local environment. VWR will not be held
liable for any related omission or for not obtaining the required approval or authorization, unless any
refusal is due to a defect of the product.
Technical service
Web Resources
Visit the VWR website at vwr.com for:
• Complete technical service contact information
• Access to the VWR Online Catalogue, and information about accessories and related products
• Additional product information and special offers
Contact us For information or technical assistance contact your local VWR representative or
visit.vwr.com.

VWR Collection Manual ver 1, rel 1 20.02.2017
Warranty
VWR warrants that this product will be free from defects in material and workmanship for a period of
two (2) years from date of delivery. If a defect is present, VWR will, at its option and cost, repair,
replace, or refund the purchase price of this product to the customer, provided it is returned during the
warranty period. This warranty does not apply if the product has been damaged by accident, abuse,
misuse, or misapplication, or from ordinary wear and tear. If the required maintenance and inspection
services are not performed according to the manuals and any local regulations, such warranty turns
invalid, except to the extent, the defect of the product is not due to such non performance.
Items being returned must be insured by the customer against possible damage or loss. This
warranty shall be limited to the aforementioned remedies. IT IS EXPRESSLY AGREED THAT THIS
WARRANTY WILL BE IN LIEU OF ALL WARRANTIES OF FITNESS AND IN LIEU OF THE
WARRANTY OF MERCHANTABILITY.

VWR Collection Manual ver 1, rel 1 20.02.2017
Detales: Español
Especificaciones del producto
ESPECIFICACIONES
Tasa de muestreo: 30 segundos
Rango: –58,0 hacia 158,0°F / –50,0 to 70,0°C
Resolución: 0,1°
Precisión: ±0,5°C (–30 hacia 50°C), ±1°C en otros casos
Característicos: termómetro con pantalla LED, visualización de la memoria máxima/mínima y alarma
high/low
FUNCIONAMIENTO
Proceso de borrado de la memoria
Tecla: Pulse esta tecla para borrar totalmente la memoria MIN/MAX de la sonda seleccionada
Tecla de la función Mode: Pulse esta tecla para alternar entre “MIN/MAX DISPLAY MODE” y
“ALARM DISPLAY MODE”.
Probe 1/2: Permite seleccionar el canal de la sonda que se quiere visualizar.
() Alarma On/Off: Permite habilitar o inhabilitar la función de la alarma
Tecla Reset : Permite configurar el valor de la alarma alta/baja “HI/LO AL” y las memorias
MIN/MAX para ambas sondas
Tecla MIN/LO: En la sección “MIN/MAX display mode” pulse esta tecla para borrar la lectura mínima
registrada para el canal de sonda seleccionado. En la sección “Alarm displace dome”
pulse esta tecla para incrementar el valor de alarma baja “LO AL” configurado.
Tecla MAX/HI: En la sección “MIN/MAX displace dome” pulse esta tecla para borrar la lectura
máxima registrada para el canal de sonda seleccionado. En la sección “Alarm displace
dome” pulse esta tecla para incrementar el valor de alarma alta “HI AL” configurado.
Interruptor °C/°F: Selecciona la unidad de medida de la temperatura (Celsius o Fahrenheit).
NO SE PUEDE PROGRAMAR LOS VALORES MÍNIMOS Y MÁXIMOS DE TEMPERATURA
INSTALACIÓN DE LA SONDA
Conecte los sensores de la sonda a la toma de alimentación ubicada en la parte lateral del
dispositivo, luego deslice el interruptor de canales de sonda hacia el canal deseado.
Con el sensor de sonda conectado, se visualiza en la pantalla del dispositivo la temperatura actual de
la sonda y las temperaturas MIN/MAX.
Cuando se remueve el sensor del dispositivo, se visualiza en la pantalla : “--.-”.
SONDAS SUMINISTRADAS
Modelo 10050-584 (620-2334)/10050-586 (620-2335)-- Suministrado con una sonda para
frasco(10050-584 (620-2334)) o 2 sondas para frasco(10050-586 (620-2335)), diseñadas para uso en

VWR Collection Manual ver 1, rel 1 20.02.2017
frigoríficos o congeladores. Las sondas para frasco están llenas de solución de glicol patentada y no
toxica clasificada como GRAS (lista de productos generalmente reconocidos como seguros) por la
FDA (Food and Drug Administration) eliminando así preocupaciones relacionadas al contacto
inadvertido con alimentos o agua potable.
El frasco lleno de esta solución simula la temperatura de los otros líquidos almacenados en el
frigorífico o congelador. Se suministra Velcro® y banda magnética para colocar el frasco en el
frigorífico o congelador. Se proporciona igualmente un cable extra fino encima del cual se puede
cerrar la puerta del frigorífico o congelador. (No sumerja las sondas de frasco en líquidos).
Modelo 10050-588 (620-2336) / 10050-590 (620-2337)-- Suministrado con sonda de plástico
estándar dotada de cable. Diseñado por uso en aire y líquidos, el sensor y el cable de esta sonda
pueden quedar totalmente sumergidos.
Modelo 10050-592 (620-2338)– Suministrado con 1 sonda para frasco y 1 sonda de de plástico
estándar dotada de cable.
Modelo 10050-594 (620-2339) / 10050-596 (620-2340)– Suministrado con una sonda para frasco de
vacuna de1,5ml (10050-594 (620-2339)) o 2 sondas de frasco de vacuna de 5ml (10050-596 (620-
2340)). Diseñada para ser compatible con bandejas para vacuna.
Se suministra Velcro® y cinta magnética para el montaje del frasco dentro del frigorífico/congelador
y para el montaje del dispositivo fuera del frigorífico. Se proporciona igualmente un cable extra fino
encima del cual se puede cerrar la puerta del frigorífico.
REINICIALIZACION DEL DISPOSITIVO
Utilice un pequeño objeto de extremidad aguda para pulsar la tecla RESET en la parte trasera del
dispositivo.
NOTA: Las configuraciones de temperaturas de alarma HI/LO y de memoria MIN/MAX serán
borradas cuando se reinicia el dispositivo.
MODOS DE VISUALIZACIÓN
El dispositivo tiene dos modos de visualización: MIN/MAX DISPLAY MODE y ALARM DISPLAY
MODE. Para alternar entre modos, pulse la tecla MODE une vez.
MIN/MAX DISPLAY MODE – está indicado por la visualización de la temperatura actual y de las
temperaturas MIN (mínimas) y MAX (máximas) . Estos valores corresponden a los del canal de
sonda seleccionado con el interruptor de canales.
ALARM DISPLAY MODE - está indicado por la visualización de la temperatura actual y de las
configuraciones de las alarmas HI y LO . Estos valores corresponden a los del canal de sonda
seleccionado con el interruptor de canales.
REINICIALIZACION DE LA MEMORIA MIN/MAX
Existen dos maneras de reiniciar la memoria de lecturas:
1.Para reiniciar las lecturas mínimas y máximas registradas – pulse la tecla MEMORY CLEAR.
2.Para reiniciar sólo las lecturas mínimas o sólo las lecturas máximas registradas en la memoria-

VWR Collection Manual ver 1, rel 1 20.02.2017
pulse la tecla MIN o la tecla MAX (en la parte trasera del dispositivo) una vez para reiniciar la parte
de la memoria correspondiente. La otra parte quedará inafectada.
NOTA: Usted tiene que ser en la sección MIN/MAX Display Mode para reiniciar la memoria.
VISUALIZACIÓN EN °F O °C
Para visualizar las lecturas de temperaturas en Fahrenheit (°F) o Celsius (°C), deslice el interruptor
en la parte trasera del dispositivo hacia °F o hacia °C.
CONFIGURACIÓN DE LA ALARMA HIGH/LOW
1.Pulse la tecla MODE hasta que se visualice la sección ALARM DISPLAY MODE indicada por la
visualización de la temperatura actual y de los valores de alarma HI y LO.
2.Para configurar la alarma de temperatura alta HIGH, pulse la tecla HI en la parte trasera del
dispositivo para incrementar el valor limite de intervalos de 0.1° ; pulse y mantenga pulsada la
tecla para incrementar el valor limite de intervalos de 1°.
3.Para configurar la alarma de temperatura baja LOW, pulse la tecla LO en la parte trasera del
dispositivo para incrementar el valor limite de intervalos de 0.1° ; pulse y mantenga la tecla pulsada
para incrementar el valor limite de intervalos de 1°.
ALARMA
La alarma del dispositivo disparará que se visualice MIN/MAX DISPLAY MODE o ALARM DISPLAY
MODE.
Nota: No se puede modificar los limites de alarma cuando una alarma está activa.
Deslice el interruptor ALARM ON/OFF hacia la posición ON. Cuando la temperatura actual
visualizada supera el punto de ajuste de temperatura alta HI, o se encuentra debajo del punto de
ajuste de temperatura baja LO, la alarma sonará durante 1 minuto y luego emitirá un pitido durante
3 segundos en cada minuto. La alarma seguirá sonando y la pantalla LCD seguirá destellando aun
cuando la temperatura actual revuelva dentro del rango.
FUNCIONAMIENTO DE LA ALARMA
Para apagar la alarma activa, pulse la tecla Memory Clear o deslice el interruptor de alarma hacia la
posición OFF. Cuando se encuentra en la sección ALARM DISPLAY MODE, pulse la tecla
MIN/LO o la tecla MAX/HI para apagar la alarma.
Las condiciones de alarmas quedarán activas hasta que sean borradas aun cuando el valor de la
temperatura vuelva dentro de los limites de alarma.
No se puede modificar los limites de alarma cuando una alarma está activa.
VISUALIZACIÓN DE LOS LIMITES DE ALARMA DE TEMPERATURA
1.Cuando está en la sección ALARM DISPLAY MODE, deslice el interruptor PROBE hacia la posición
deseada (1 o 2) para visualizar las lecturas del sensor de temperatura.
2.Se visualizará en la pantalla la temperatura actual y los puntos de ajuste de alarmas de
This manual suits for next models
15