Tenmars TM-747 User manual

TM-747/747D
4-Channel Thermometer
User’s Manual
HB
2
TM7470000


CONTENTS
1 Introduction.......................................................1
2 Accessories.......................................................1
3 Safety Precaution..............................................1
4 Meter Description..............................................2
5 Operation..........................................................3
5.1 Power on or off.............................................3
5.2 Turn on backlight..........................................3
5.3 Alarm on or off..............................................3
5.4 Auto Recording ............................................4
5.5 Thermocouple Type Selection (TYPE): ........4
5.6 Manual Record.............................................4
5.7 Unit Switch for °C, °F and K .........................5
5.8 Data Hold.....................................................5
5.9 Disable Auto Power Off Function..................5
5.10 Relative Value Measurement........................5
5.11 Offset Value.................................................6
5.12 Change Display of T1-T2 vs. T1 & T2...........7
5.13 Change Display of of T3-T4 vs. T3 & T4.......7
5.14 Time Display................................................8
5.15 Read Record................................................8
5.16 Maximum /Minimum/ Average Value............9
5.17 Setup ...........................................................9
6 Software Installation (TM-747D)......................12
7 General Specifications ....................................14
8 Electrical Specifications:..................................16
9 Maintenance or Repair....................................17
10 Battery Replacement.......................................17
11 End of Life Disposal ........................................17

1
1 Introduction
TM-747X series thermometers are K /J/ T / E/ R / S/ N type
thermocouple thermometers.
2Accessories
1
1
6
1
1
1
1
Meter
User manual
1.5V AAA Carbon zinc battery
Carrying case
9V AC to DC adaptor (TM-747D)
USB cable (TM-747D)
Installation disk
(TM
-
747
D)
3 Safety Precaution
Note! Please refer to this manual.
Improper use may damage the meter and its
components.
Complies with European Directive.
Do not operate in environments with flammable gas or
humid environments.
Operating altitude: up to 2000M.
Operating environment: Indoor use; Pollution degree 2.
Clean with soft cloth when dirty, such as glasses cloth. Do
not clean with chemicals and other solvents.
Class B – Equipment for use in all establishments other
than domestic.
Group 1 – RF energy generated is needed for internal
functioning.

2
4Meter Description
1. Alarm setup button / Auto record button (TM-747D)
2. Power button
3. Backlight button
4. K /J/ T / E/ R / S/ N type switching/manually record
button.
5. Hold button / Auto-shutdown button
6. Interchange button for "T1 - T2" vs. "T1 & T2"
7. Unit ° C / ° F / K switch button
8. Maximum value, Minimum value, Average value,
elapsed time / Setup button
9. Relative value / offset value button
10. Interchange button for "T3 - T4" vs. "T3 & T4"
11. Time button / Read button
12. Thermocouple input terminal T1/T2 T3/T4
13. External power DC 9V input
14. USB data output jack
15. Battery cover and tilt stand
16. LCD display
With DC-IN connected, appears
on the screen.
Without heat-sensing wire connected,
will appear on the screen.

3
5Operation
5.1 Power on or off
Press button,turn on or turn off.
5.2 Turn on backlight
press button to turn on or turn off the
backlight.
*The backlight will automatically turn off after
being lit for 15 seconds.
5.3 Alarm on or off
press to turn “on” or “off” the alarm, followed
by "LIMIT" appears on the LCD.
This LCD shows the status "on" or "off" of the
alarm function. It is possible to set the upper and
lower limit of the temperature range by yourself.
When the temperature reading exceeds the range,
the buzzer will alarm until the temperature back to
the range or the alarm function turned off.
(For the limit range setup, please enter 5.17
Setup Function)

4
5.4 Auto Recording
Press and hold for ≥ 2 seconds to enable or
disable the auto-recording function "LOG"
(TM747D)
The LCD display will first show the number of logs
for REC/MEM START mode and then switch back to
temperature mode after 1 second.
The auto-recording is such if the current number is 3,
it will be increased to X by 1 at each pressing. (the
number of logs in TM-747D is up to 16800)
* Execute "LOG" to disable the auto-shutdown
function. For recording for a long time, please
connect the meter to an external power supply.
5.5 Thermocouple Type Selection (TYPE):
Press to enable selection of different
thermocouple types in a full cycle.
5.6 Manual Record
Press , followed by "REC" appears on the LCD,
the log number of records will increase by +1 at
each time, and one log will be recorded.
The manual records are up to 200 logs.

5
5.7 Unit Switch for °C, °F and K
Press and hold for ≥ 2 seconds to switch the
unit: Celsius (°C), Fahrenheit (°F), or Kelvin (K).
5.8 Data Hold
Press to enable or disable the data hold
function.
5.9 Disable Auto Power Off Function
Press and hold for 2 seconds to enable or
disable the auto power off function.
5.10 Relative Value Measurement
Press to enter the relative:
Just use one thermocouple, the difference between
two temperatures can be measured.
For example, measure the first temperature as 25°C,
and then press to show 0°C on the LCD.
Again, measure the second temperature. If the
second temperature is measured as 30°C, the LCD
will show the difference value 5°C (30-25=5°C).
Again, press to disable the function.
The left shows the relative value, right shows the
temperature value being measured.
press the / and hold to switch the
display between "T1 & T2" and "T3 & T4".

6
5.11 Offset Value
The user can set the offset value to compensate for
the error due to the working thermocouple wire.
Press and hold for ≥ 2 seconds to set the
offset value in OFFSET mode.
b. Press or to increase or decrease the
offset value by the scale of 0.1°C, °F, or K. The
setting range is ±5°C, ±5K, or ±9°F.
*If to set offset value of another channels, press
and to switch and select between "T1
& T2" and "T3 & T4" for adjustment. Press
to save the setting value and exit the setting mode.
*If the value updated to be 0.1, "OFFSET" will
continuously appear on the LCD. "OFFSET" will not
disappear unless it is corrected back to 0.0.

7
5.12 Change Display of T1-T2 vs. T1 & T2
Press to enable or disable T1 – T2 .
The value shown on the left shows the difference of
T1 – T2, and the one shown on the right is the T1
value which is measured presently.
Press and hold for ≥ 2 seconds for the
position change of T1 and T2.
5.13 Change Display of of T3-T4 vs. T3 &
T4
Press to show the value of T3 - T4.
Again, press to disable the function.
The value shown on the left presents the difference
of T3-T4, the one shown on the right is the
temperature measured through T3 channel.
Press and hold for 2 seconds for the
position change of T4 and T3 on the right.