1. Congratulations!
In the seca 644 scale with standing aid,
you have purchased a high-precision yet
sturdy piece of equipment.
seca has been putting its experience at
the service of health for over 150 years
now, and as market leader in many coun-
tries of the world, is always setting new
standards with its innovative develop-
ments for weighing and measuring.
In accordance with national regulations, the
seca 644 scale with standing aid is used
primarily in hospitals, doctors’ surgeries
and in-patient nursing facilities.
The weight display can be switched be-
tween kilogrammes (kg), pounds (lbs) and
stones (sts). Weight is determined within a
few seconds. Weight is determined within
a few seconds.
In addition to the conventional determina-
tion of weight, the seca 644 also has a
function for determining body mass index.
For this purpose, height is entered on the
keypad and the body mass index associ-
ated with the weight value is output.
You can use the pre-Tare function to
record the weights of, e.g. three types of
chair. This weight is automatically deduct-
ed in the weighing operation.
The large, secure standing aid also ena-
bles persons who are unsteady on their
feet to stand safely on the scale.
The scale is of a very sturdy construction
and will give you long and loyal service. It
is easy to operate and the large display is
easy to read.
The seca 644 can be moved on castors
and its low power consumption combined
with its rechargeable batteries mean it will
provide long, mobile service.
2. Safety
Before using the new scale, please take a little time to read the following safety instruc-
tions.
• Follow the instructions in the instruc-
tions for use.
• Keep the operating manual and the
declaration of conformity contained
therein in a safe place.
•Take measures to ensure that persons
who have difficulty walking do not fall
over when using the scale.
•The scale must be used with care. The
upper surface may be slippery if wet!
• After moving the scale, please lower
the scale slowly to the operating height.
• Do not drop the scale or subject it to
violent shocks.
•Be sure to use only the seca mains unit
supplied. Before plugging the mains
unit into a socket, check that the mains
voltage indicated on the mains unit
matches local mains voltage.
•When using the scale with a mains unit,
ensure that the supply cable is routed in
such a way as to exclude any type of
tripping hazard.
•Have repairs carried out only by author-
ised persons.

Model 644 21
GB
3. Before you start…
Unpacking
–Remove the packaging, although leave the display head attached to the column. Place
the scale platform on a solid, flat surface.
The scope of supply includes:
•Side section with two columns and the
display head attached for transport
• Side section with one column
• Scale platform
• Railing
• Cover for the wheel casing
• Cover for the side casing
•2triangular covers right/left
• Display mount
• Set of assembly bolts
• Allan key, size 6
• Phillips screwdriver size 1 and size 3
•1screwdriver
•1seca mains unit
•1set of operating instructions

22
Assembling the scale
Step 1: Assemble the first side section
You will need the following bolts:
–Position the side section on the right-
hand side of the scale platform.
–Bolt the side section to the scale plat-
form using two Allan bolts.
–Use an Allan bolt to attach the triangular
reinforcement to the scale platform.
Step 2: Assemble the side section with the wheel casing
You will need the following bolts:
–Position the side section with the wheel
casing on the rear side of the scale plat-
form.
–Bolt the side section to the scale plat-
form using two Allan bolts.
3 x Allan bolts M5
3 x Allan bolts M5 1 x Sheet metal bolt, ø 6,3 mm

Model 644 23
GB
–Use an Allan bolt to attach the triangular
reinforcement to the scale platform.
–Use the sheet metal bolt to attach the
two side sections to the central column.
Step 3: Assemble the railing
You will need the following bolts:
–Position the railing on the three col-
umns. The large bore holes must point
downwards, and must be inserted into
the heads of the columns.
–Use an oval head bolt to attach the rail-
ing to the two outer columns.
1 x Allan bolts M6 2 x Oval head bolt M6

24
–Place the display mount with the peg-
shaped end over the central column on
the railing.
–Turn the display mount until the peg
clicks into place.
–Screw the display mount tightly into
place using the Allan bolt.
Step 4: Assemble the display head
You will need the following bolts:
–Remove the transport protection from
the display head.
–Position the display head over the dis-
play mount in such a way that the posi-
tion markers are aligned.
–Use the three Allan bolts to attach the
display head to the display mount.
3 x special Allan bolt, ø 2,5 mm
Markers

Model 644 25
GB
Step 5: Assemble the side section covers
You will need the following bolts:
–Place the scale on its side, with the rear
side section facing upwards.
–Position the cover with the wheels over
the side section, and attach this secure-
ly using 5 Phillips bolts:
3 x from below, and
2 x from above.
–Position the cover on the reinforcing tri-
angle, and bolt this firmly in place using
two bolts.
–Now turn the scale over so that the oth-
er side faces upwards.
–Position the cover with the recess
pointing inwards over the side section,
and attach this securely using 5 Phillips
bolts:
3 x from below, and
2 x from above.
–Position the cover on the reinforcing tri-
angle, and bolt this firmly in place using
3 bolts.
Step 6: Assemble the electrical connections
–Connect the cable which comes out of
the central column to the electronic box
socket. In order to prevent measure-
ment errors, lay the cable as shown in
the diagram.
–Unscrew the bolts which secure the lid of
the battery compartment, and remove.
–Connect the plug with the rechargeable
battery block located in the battery
compartment.
–Replace the lid of the battery compart-
ment, and bolt this firmly in place.
– Turn the scale to the upright position.
16 x Phillips bolts, ø 3,5 mm
Recess
Plug

26
Power supply
The scale is powered by a mains unit and by rechargeable batteries. The scale works in-
dependently of the mains, which is very handy for mobile use.
The LC display and the evaluation electronics have a low power consumption – 2,000
weighing operations can be performed with a fully charged battery. Once battery voltage
is no longer adequate, the battery symbol first appears in the display to indicate
that battery power is dropping. If the batteries are not charged, battery voltage will ulti-
mately not be adequate for error-free measuring. In this case, bAtt appears in the display
and weighing is no longer possible.
The rechargeable batteries are charged by
the mains unit supplied. For this purpose,
the mains unit needs to be connected to
the mains for at least 24 hours.
4. Setting up and aligning the scale
– Place the scale on a level surface.
–Undo the locknuts using the four foot
screws (1).
–Align the scale by undoing the foot
screws. The air bubble in the spirit level
must be right in the centre of the
circle (2).
– Then re-tighten the locknuts (3).
–Ensure that only the feet of the scale are
in contact with the floor. The scale may
not be in contact at any other point.
Important!
The alignment of the base of the scale must be checked and corrected if neces-
sary every time the location of the scale is changed.
1.
2.
3.
Libelle
spirit level

Model 644 27
GB
5. Operation
Controls and displays
The controls and the display are on the display head on the column.
Switch scale on and off
press briefly sswitch between kg and lbs
hold down sswitch between kg and sts
in function menu sincrease value, scroll
press briefly sactivate hold function
hold down sset display to zero, activate/deactivate tare function
in function menu sdecrease value, scroll
Activate BMI or pre-Tare function, and confirm key for entries
In order to ensure that the display can be
read from different positions, the display
head can be rotated 180° in both
directions.

28
Correct weighing
– Press the Start key.
SE(a, 8.8.8.8.8. and 0.00 appear consecu-
tively in the display. The scale is then auto-
matically set to zero and ready to use.
–If necessary, switch the weight display
(see page 28).
–Ask the patient to mount the scale.
Once on the scale, the patient should
stand still.
–Read the weighing result off the digital
display.
–If a load greater than 300 kg or 660 lbs
is placed on the scale, St0P appears in
the display.
–To switch off the scale, press the Start
key again.
Switching the weight display
The weight display can be switched between kilogrammes (kg) and pounds (lbs) or be-
tween kilogrammes (kg) and stones (sts).
–Switch on the scale using the green
Start key. Weight is initially displayed in
the last setting selected.
–To switch the weight display between
kilogrammes and pounds, press the kg/
lbs/sts key.
– The kg or lbs display lights up.
–To switch weight display between
pounds and stones, hold down the
kg/lbs/sts key for approx. 2 seconds.
The lbs or sts display lights up.
Automatic switch-off/economy mode
The scale automatically detects whether it is being operated on rechargeable batteries
alone or whether it is connected to the mains.
– With rechargeable battery operation: automatic shut-down after 5 minutes.
– With mains operation: shut-down only via keypad.
Other manuals for 644 Series