Lotus DT-8806H User manual

Non-Contact Forehead
InfraRedThermometer
UserManual
DT-8806H

Non-Contact Forehead IR Thermometer User Manual

3
Non-Contact Forehead IR Thermometer User Manual
Contents
1.General Description...................................................................
2.Safety Information....................................................................
3.Features....................................................................................
4.Intended Use.............................................................................
5.Configuration............................................................................
6.Indicator....................................................................................
7.Description of Symbols..............................................................
8.Technical Specifications.............................................................
9.Use..........................................................................................
10.Measuring Operation...............................................................
11.Advice.....................................................................................
12.Maintenance and Cleaning......................................................
13.Troubleshooting......................................................................
14.This appliance conforms to the following standards...............
15.EMC Statement.......................................................................
16.Classification...........................................................................
17.Disposal..................................................................................
4
4
5
6
7
7
8
9
10
11
13
13
14
16
16
21
21

4
Non-Contact Forehead IR Thermometer User Manual
1.General Description
2.Safety Information
The device is intended for the intermittent determination of
human body temperature in a no touch mode using the center of
the forehead as the measurement site on people of all ages. The
measured temperature correlates to axillary temperature. The
device can be used in clinical and home environments.
Depending on various skin types and thickness, there may be
temperature difference.
This device must only be used for the purposes described in this
instruction manual
This device must only be used in an ambient temperature range
between +10 and +40°C
Do not expose this thermometer to electric shocks.
Do not expose this thermometer to extreme temperature conditions
of >+50°C or <-20°C
Do not use the device in relative humidity higher than 85%.
Do not use the device near large electromagnetic fields such as found
with cordless or cell phones.
Keep the device away from water and heat, including direct sunlight.
Do not drop or knock the device, and do not use if damaged.
It may affect the accuracy of measurements when the forehead is
covered by hair, perspiration, cap or scarf.(See Part 10-4)
Keep the Measuring distance as 5cm 15cm (2in-5.9in).(See Part 10-4)
When the body infrared thermometer should be left in that room
during 15 to 20 minutes before using.
It may affect the accuracy of measurements when the forehead is
covered by perspiration or other factors.
•Clean the glass with a cotton bud lightly moistened with 70% isopropyl
alcohol.
•
•
•
•
•
•
•
•
•
•-
•
•
•It is recommended to calibrate this device every 2 year.
•At least 11 years old (5 years intensive reading experience).
The device can be used in clinical and home environments.
•

5
Non-Contact Forehead IR Thermometer User Manual
Importance:
Before taking of the temperature make sure to remove hair and
perspiration from the forehead.
Selecting “Body” mode to measure the body temperature;
Use of this thermometer is not intended as a substitute for consultation
with your physician.
Should a problem occur with your device,please contact your retailer.
Do not attempt to repair the device yourself.
According to EMC standard, the medical electronic products should
be maintained specially.
•
•
•
•
•
•Precise non-contact measurements
•User selectable ºC or ºF
•Selectable Body and Surface temp
•Set Alarm value
•Memorization of the last 32 measurements
•Automatic Data Hold & Auto power off
•Automatic selection range and Display Resolution 0.1ºC(0.1ºF)
•Backlight LCD display
3.Features

Measurement Method
Rectal
Oral
Axillary
Ear
Normal Temp
36.6 to 38
35.5 37.5
34.7 37.3
35.8 38
°C
to
to
to
Normal Temp °F
97.8 to 100.4
95.9 to 99.5
94.4 to 99.1
96.4 to 100.4
6
Age
0-2 years
3-10 years
11-65 years
>65 years
Temp °C
36.4 to 38.0
36.1 37.8
35.9 37.6
35.8 37.5
to
to
to
Temp °F
97.5 to 100.4
97.0 100.0
96.6 99.7
96.4 99.5
to
to
to
Non-Contact Forehead IR Thermometer User Manual
4.Intended Use
The device is intended for the intermittent determination of human
body temperature in a no touch mode using the center of the
forehead as the measurement site on people of all ages. The
measured temperature correlates to axillary temperature.
The temperature of the human body varies throughout the day. It can
also be influenced by numerous external factors: age, sex, type and
thickness of skin…
Normal Temperatures According To Measurement Method
Normal Temperatures According To Age
Intend operator:
At least 11 years old (5 years intensive reading experience),
no maximum.
The device can be used in clinical and home environments.

7
5.Configuration
1-IR Sensor
2-LCD Display
3-Mode Selection
4-Down Button
5-Up Button
6-Mode Button
7-Measurement Trigger
8-Battery Cover
1-Surface mode Symbol
2-Body mode Symbol
3-Digital readout
4-Battery Symbol
5-The order number
6-Save data readout
7-Temperature °C(Celsius)/
°F(Fahrenheit) Scale
8-Buzzer symbol
6.Indicator
Non-Contact Forehead IR Thermometer User Manual
1
2
7
3
6
5
4
8
1
2
3
5
8
7
4
6

8
7.Description of Symbols
Non-Contact ForeheadIRThermometerUserManual
Body
Surface
The device is in accordance with Medical Device Directive
93/42/EEC
The device is in accordance with FCC Part 15 Subpart
B:2007/Radio Frequency Devices IC Regulation ICEC-003:
2004 Interference-causing Equipment Standard-Digitial
Apparatus
3V DC power supply, [2x1.5V AA (R6P/LR6)]
Application part of Type B
In order to protect the environment, please recycle the
battery according to the local regulations
Indication of Mode
Attention,consult accompanying documents
The name and the address of the manufacturer
Production Date
Serial Number
SN

9
Non-Contact Forehead IR Thermometer User Manual
8.Technical Specifications
Normal Conditions of Use
Operating Temperature
Storage
Humidity Rate
Power
Size
Weight
Measuring Range
In Body Mode
In Surface Temp Mod
Measuring Distance
Automatic Stop
Display Resolution
Temperature
0.1°C (0.1°F)
10 to 40°C (50 to 104°F)
-20 to 50°C (-4 to 122°F)
<85%
DC 3V [2x1.5V AA (R6P/LR6)]
149x77x43mm/5.9x3x1.7in(LxWxH)
Gross 400g / Net 172g
32.0 to 42.5°C (90 to 108°F) ±0.3°C(0.54°F)
0 to 60°C (32 to 140°F) ±1°C(1.8°F)
5 cm – 15 cm (2 in – 5.9 in)
7sec.

10
Non-Contact Forehead IR Thermometer User Manual
9.Use
9-1.Install battery
9-2.For the first use or when inserting new batteries, wait for
10 minutes for the warm-up of the apparatus.
9-3.If the device is not used for a long time, once you turn i
t on again,
the device will test the room temperature first and will delay
turning on for one or two seconds.
9-4. Aim towards the forehead (see the diagram below for the
positioning), from a distance of 5cm(2in), press the measuring
key, the temperature is displayed immediately. Making sure there
is no hair, perspiration, cosmetic or cap covered on the forehead.
5 to 15cm
2 to 5.9in
Selecting "Body" mode to measure the body temperature
This mode displays the armpit-equivalent value of the measured
forehead temperature.
Intend operator:
At least 11 years old (5 years intensive reading experience),
no maximum.
The device can be used in clinical and home environments.