U-Line H-1096 User manual

PAGE 1 OF 5 0412 IH-1096
DELUXE
COUNTING SCALE
1-800-295-5510
uline.com
1-800-295-5510
uline.com
πH-1096, H-1097
H-1098, H-1099
INSTALLATION
UNPACKING
1. Unpack and verify the following components have
been included:
• Scale unit
• Instruction manual
• Weighing pan
• Power adapter
• Allen wrench
• Four head cap screws
2. Save the packaging material. This packaging ensures
the ideal protection for the storage or transport of the
product.
LEVELING
3. Retract the right rear foot (A) completely into the
scale. (See Figure 1)
4. Adjust the two front feet (A) until the leveling bubble (B)
is centered in the circle. (See Figure 1)
5. Extend the right rear foot until it contacts the mounting
surface. Make certain that all four feet are contacting
the mounting surface.
6. Make certain the scale is stable by placing both
hands on the scale platform and attempt to rock
the scale. If it moves, adjust the leveling feet until the
scale stands firmly on the mounting surface.
ASSEMBLY
1. Using the Allen wrench and head cap screws
provided, secure the pan support (C) to the four posts
(D) over the sensor (E). Next, place the weighing pan
(F) over the pan support. (See Figure 2)
A
B
Figure 1
C
D
E
F
Figure 2

PAGE 2 OF 5 0412 IH-1096
CONNECTING TO POWER
NOTE: When your scale is first powered on, it
automatically runs through a 6-second display
test. The scale next displays “C nnn g” (where
nnn equals the full scale capacity in grams),
performs self-diagnostics to ensure proper
operation and then sets itself to zero. A wake-up
message is always transmitted upon power
up. The scale is then ready for use. For best
accuracy, allow the scale to warm up for 5
minutes.
1. Insert the AC adaptor into an electrical outlet.
Connect the other end of the adapter into the back
of the scale (G). (See Figure 3) 2. Once plugged in, the scale automatically turns on.
INSTALLATION CONTINUED
KEYBOARD FUNCTIONS
G
Figure 3
zero Assigns the pan and whatever is currently being
weighed a value of zero. While the scale is determining
a stable weight reading, the display will indicate “BUSY.”
sample
size
Displays preprogrammed sample sizes (10, 25, 50, or
100 pieces).
count Instructs the scale to begin or resume counting. This
key may also be used to “update” the average piece
weight for increased counting accuracy.
setup/cal Instructs the scale to enter a setup or perform a
calibration.
a.p.w. Instructs the scale to accept an average piece weight
entry.
clear Clears the last number entered or operation requested,
and displays weight.
tare Instructs the scale to accept a numerical tare weight
entry. If a container is placed on the pan, pressing tare
assigns its weight as the “tare” (i.e. container weight)
and saves it in memory. The display will read net
weight (i.e. contents weight).
gross Toggles between the gross weight and the net weight.
(gross weight - tare weight = net weight)
print Scale sends displayed and stored or programmed
information to a printer, computer (or other RS-232
peripheral device).
units Converts the weighing units of measure (e.g. changes
from grams to ounces).
accum Sums and stores the results of multiple weighing or
counting operations.
i.d. Instructs the scale to store an identification number of
up to 25 digits (e.g. a part number or account number)
to be printed or transmitted to a peripheral device.
display Instructs the scale to recall the value stored in any of
the various registers (e.g. tare, accumulate, etc).
reset Sets all counting and weighing registers to zero. Clears
all volatile ID registers. Assigns the pan and whatever
is currently being weighed a value of zero. Displays full
scale capacity.

PAGE 3 OF 5 0412 IH-1096
KEYBOARD FUNCTIONS CONTINUED
OPERATION
•Instructs scale to enter a decimal point during numeric
entry.
- Instructs scale to change the sign of a number.
Interpreted as a hyphen when entered as part of an
ID number.
When preceding any register that stores a value,
instructs the scale to clear the stored value from that
specific register. (e.g. Pressing the - tare keys will clear
the tare value in memory.)
When preceding the units key, instructs the scale
to convert the weighing units to the factory default
(grams).
0-9 Instructs the scale to enter numerical values given for
registers such as ID number, count, tare or average
piece weight.
BASIC WEIGHING
1. Press the zero key and place the object(s) to be
weighed on the pan. (See Figure 4)
2. Wait for the “OK” symbol, then read the weight from the
display.
CONVERTING THE WEIGHING UNITS
1. Press the units key repeatedly. (See Figure 5)
Continue pressing the key until the units you wish to
use are displayed.
NOTE: The order of units is as follows:
GRAMS - OUNCES - POUNDS - TROY OUNCES -
PENNYWEIGHTS - CARATS - “x” UNITS (user defined)
- GRAMS.
2. To return to the factory default (grams) press - units.
WEIGHING WITH A CONTAINER
1. Press the zero key. Place the empty container on the
pan.
2. Press the tare or zero key. (If the tare weight is known, a
full container may be placed on the pan. In this case,
enter the tare weight through the keyboard and then
press the tare key.)
3. Wait for the “OK” symbol. Place or pour objects or
liquids into the container.
4. Wait for the “OK” symbol, then read the net weight from
the display.
NOTE: When you press the tare key, the
container weight is stored in memory and
the display will read zero. You must use the
tare key, not the zero key, in step #2 to store
the container weight in memory. To recall the
stored tare value, press display tare. To clear
the stored tare value, press - tare.
ZERO SAMPLE
SIZE COUNT CLEAR RESET
PRINT
Figure 4
display int
tare gross a.p.w.
units accum i.d.
low
bat
set up
cal 789
456
123
-0•
remote
local
bat
charge
Figure 5

PAGE 4 OF 5 0412 IH-1096
DISPLAYING THE GROSS WEIGHT
1. Press display followed by the gross key.
2. To toggle between the gross weight and the net
weight, press the gross key repeatedly. The gross
weight will equal the net weight unless a container
weight has been entered using the tare key (see
section, “Weighing with a Container”).
OPERATION CONTINUED
BASIC COUNTING
0.00 g
clearprintzero sample
size count
1. Place an empty container on the scale. Press the zero key.
...........................................................................................................................................................................
Add 10 Pc
clearprintzero sample
size count
clearprintzero sample
size count
clearprintzero sample
size count
clearprintzero sample
size count
2. Press the sample sizekey. Each time the key is pressed the requested sample size will increase.
(i.e., Add 10, 25, 50, 100 PCs).
...............................................................................................................................
...............................................................................................................................
............................................
10 Pc
3. Place the requested number of pieces in the container. Press the count key.
202 Pc
4. Fill the container to the desired number of pieces.
...........................................................................................................................................................................
0.00 g
5. Remove the pieces from the container and press the clear key to return the scale to the weight display.
Figure 6
NOTE: The scale will never request a sample size less than the preset minimum sample size. The scale may
request that more pieces be added after the initial sampling if a minimum accuracy requirement has
been preset. (See the “Using the Minimum Sample Size or Minimum Accuracy Feature.”)

PAGE 5 OF 5 0412 IH-1096
OPERATION CONTINUED
USING THE FULL COUNTING FUNCTION
1. Press the zero key.
2. Place a known number of pieces on the pan.
3. Enter the number of pieces through the scale
keyboard.
4. Press the count key.
5. Wait for the “OK” symbol.
6. Add pieces until you reach the desired piece count.
7. Wait for the “OK” symbol, then read the display.
NOTE: The scale may request that more
pieces be added after the initial sampling if a
minimum accuracy and/or minimum sample
size requirements have been preset.
COUNTING WITH A CONTAINER
1. Press the zero key.
2. Place the container on the pan.
3. Press the tare key or zero key. (If the tare weight is
known, a full container may be placed on the pan
in step #2. In this case, enter the tare weight through
the keyboard and then press the tare key.)
4. Wait for the “OK” symbol. Then press the sample size
key.
5. The scale display will say “Add 10. PC”. (The sample
size requirement can be increased by pressing the
sample size key repeatedly. The predetermined
sample sizes are: 10, 25, 50 and 100 pcs.
This sequence of predetermined sample sizes will
repeat itself as the sample size key is pressed.)
6. When you reach the desired sample size, place the
requested number of pieces in the container.
7. Press the count key.
8. Wait for the “OK” symbol. Then add pieces until you
reach the desired piece count.
9. Wait for the “OK” symbol. Then read the display. The
scale will read the total number of pieces you have
placed in the container.
NOTE: You may recall the tare weight by
pressing display, tare, but only if it was stored
in step #3 by using the tare key and not the
zero key.
DISPLAYING GROSS WEIGHT
DURING COUNTING
1. Press the gross key to display the gross weight.
2. Press count to return to the counting mode.
COUNTING WITH A KNOWN
AVERAGE PIECE-WEIGHT
1. Press the zero key.
2. Place the pieces on the pan.
3. Enter the known average piece weight through the
keyboard.
4. Press the a.p.w. key.
5. Wait for the “OK” symbol. Then read the display. The
scale now reads the number of pieces on the pan.
NOTE: Before you enter an average piece
weight, make sure your scale is reading in the
units you wish to enter.
πCHICAGO•ATLANTA•DALLAS•LOSANGELES•MINNEAPOLIS•NYC/PHILA•SEATTLE•MEXICO•CANADA
1-800-295-5510
uline.com
CALIBRATION
WARNING! Scale can only be accurately
calibrated using standard calibrated scale
weights. Using non-calibrated weights may result
in inaccurate scale calibration. Nothing should
be on the scale. The scale should be level.
1. Turn on the scale. Press the reset key.
2. Press the units key until the scale displays the units of
the known weight (i.e., grams, oz., lbs., etc.)
3. Place the known weight (standard scale weight) on
the weighing pan.
NOTE: For maximum accuracy, the weight used
should be as close as possible to the scale
capacity.
4. Use the numeric keypad to enter the value of the
known weight.
5. Press the setup/cal key.
Other manuals for H-1096
1
This manual suits for next models
3
Table of contents
Other U-Line Scale manuals

U-Line
U-Line OHAUS VALOR 3000 User manual

U-Line
U-Line H-2289 User manual

U-Line
U-Line H-1121 User manual

U-Line
U-Line H-1114 User manual

U-Line
U-Line H-103 User manual

U-Line
U-Line H-670 User manual

U-Line
U-Line H-2382 User manual

U-Line
U-Line H-3677 User manual

U-Line
U-Line H-5819 User manual

U-Line
U-Line H-5836 User manual

U-Line
U-Line H-2714 User manual

U-Line
U-Line H-2382 User manual

U-Line
U-Line H-8109 Manual instruction

U-Line
U-Line H-1101 User manual

U-Line
U-Line H-1679-MNBRD User manual

U-Line
U-Line H-10903 User manual

U-Line
U-Line H-9672 User manual

U-Line
U-Line H-1607 User manual

U-Line
U-Line H-1653 User manual

U-Line
U-Line H-1671 User manual