Connect the cylinder to the controller with supplied tube and fasten
with band. Using another tube causes breakage and escape of gas.
If the tube is not fastened well, there is a risk that it comes off.
Do not bend the silicon tubes. If they are plugged, gas cannot be
adequately supplied to the chamber.
Do not use any gas other than specified. Gas leaks, dissolution of
tubes and equipment, and malfunction may be occurred.
Connect gas cylinder to the controller independently. If gas is
blanched off and used with other devices, it may be the cause of
trouble for both the equipment and other devices.
Use only the specified electricity and cables. If other cables are used,
this might result in electric shocks. Using a non-specified voltage
might result in equipment damage.
When using the equipment
Please use this equipment in a large room. If you use the equipment
in a small tightly isolated room, CO2 concentration in air might rise
and adversely affect human bodies.
Do not handle the equipment with wet hands. It might cause
damage to electric equipment.
In the case of glass breakage, do not touch the glass. It might result
in injury and electric shock.
When revolving the objectives, be careful not to hit and damage
both microscope and lens heater by the wire.
Take the following points into consideration in order to avoid focus
drift during long-term microscope observations.
・Keep the room temperature stable at 20℃-28℃.
・Do not let air flow directly over the chamber.
With the exception of bath unit, wipe off media or water quickly. It
can be the cause of damage.
Do not immerse the gas spouting nozzle into water. It may cause
vibration and affect images under observation.
5