Tanita DC-360 User manual

BODY COMPOSITION ANALYZER
DC-360
Instruction Manual
Please read this Instruction Manual carefully and keep it forfuture reference.
REMOTE DISPLAY VERSION COLUMN MOUNTED VERSION

3
Contents
Before Use
For Your Safety .........................................................4
Part names & Connection Procedure.......................8
Setting the Printer Paper / Setting an SD Card .....10
Power Supply..........................................................12
Settings...................................................................13
How to Use
Taking a Measurement...........................................15
Body Composition Mode............................................... 15
Measurement Results .................................................. 17
Scale Mode.................................................................. 19
Various Criteria.......................................................20
What is body fat percentage?....................................... 20
What is total body water percentage? .......................... 21
What is visceral fat rating?........................................... 21
What is basal metabolic rate (BMR)?............................ 22
What is metabolic age?................................................ 22
What is muscle mass? ................................................. 23
What is physique rating?.............................................. 23
What is bone mass?..................................................... 24
Output and Storage of Measurement Results .......25
If Necessary
Troubleshooting ......................................................31
Technical Notes ......................................................32
The Regression Formula for
Basal Metabolic Rate (BMR) ..................................36
Specifications .........................................................38

Before Use
44
For Your Safety
This section explains precautionary measures to be taken to avoid injury to the users of this device
and others, and to prevent damage to property. Please familiarize yourself with this information to
ensure safe operation of this equipment.
Warning Failure to follow instructions highlighted with this mark could result in
death or severe injury.
Caution Failure to follow instructions highlighted with this mark could result in
injury or damage to property.
This mark indicates actions that are prohibited.
This mark indicates instructions that must always be followed.
Warning
This equipment must not be used on subjects with pacemakers or other
mechanical implants.
This equipment passes a weak electrical current through the body which
could interfere with and cause the malfunction of electrical medical implants,
resulting in serious harm.
Do not handle the plug with wet hands.
This may result in electric shock, fire, or leakage.
Do not modify this equipment in any way.
This can cause electric shock or injury, or affect the accuracy of analysis.
Do not use a multi-plug adapter.
This may cause fire.
Keep this equipment away from flammable gas and oxygen-rich
environments.
Prohibited

Before Use
55
Caution
Prohibited
Always…
Keep away from water.
Avoid using on subjects with metal allergies.
Allergic reactions may be caused by the stainless steel used in the electrodes of this device.
Do not jump on the equipment.
Do not lean on the equipment.
Do not use this equipment near other products that emit electromagnetic waves.
Do not insert fingers into gaps or holes.
Do not apply force to the display.
The screen panel may break and cause injury.
Assist persons with disabilities or children under 15 years old.
A person who is familiar with the device should assist persons with disabilities or children under
15 years old who may not be able to take measurements alone.
Clean the scale platform with appropriate disinfectant after each use.
Stand clear of the subject during measurement to ensure accuracy.
Continually monitor both the subject and the equipment for anomalies.
If an anomaly in the subject or equipment is discovered,take appropriate action, such as stopping
the equipment, while ensuring the safety of the subject.
Use the included AC adapter.
Do not lean against the equipment.
Unplug the AC cord from the equipment when moving it.
Tighten the adjustable feet when moving the equipment.
Interpretation of analysis results (e.g. evaluation of measurements and formulation of
exercise programs based on results) must be performed by a professional.
Weight loss measures and exercise based on self-analysis could be harmful to your health.
Always follow the advice of a qualified professional.

Before Use
66
For Accurate Measurements
Avoid measuring after strenuous exercise.
This may cause inaccurate measurements. Please take measurements after sufficient rest.
Avoid measuring after excessive food or fluid intake or when dehydrated.
This may cause inaccurate measurements. For greater accuracy, avoid using directly after
waking up. Use at the same time on each occasion, at least three hours after a meal.
Ensure inner thighs are not touching each other during measurement.
If necessary, place a dry towel between thighs.
Do not take measurements while using transmitting devices, such as mobile phones, which may affect
readings.
Use the equipment under the same conditions and in the same position as much
as possible for accurate tracking of changes.
Readings are affected by the level of hydration and position of the body. Please use at the
same time of day each time, under the same conditions and in the same body position.
Avoid measuring in multiple locations with greatly differing temperatures.
This may cause inaccurate measurements. Allow the equipment to stand for at least
2 hours before using if it is moved to a new location with a temperature difference of
20℃or more.
Make sure the soles of feet are free of excess dirt, as this may block the mild electric current.
Bare feet should be placed correctly on the electrode platform.
Place arms straight down during measurement.
Use in a stable location.
Errors in measurements may occur if the device is used in an unstable location.
Do not wipe the equipment with corrosive chemicals (gasoline, cleaner, etc.). Please use a neutral
detergent to clean the equipment.

Before Use
77
Scheduled Maintenance
TANITA recommends that each facility conduct periodic checks of each unit.
1. Check the following at least daily:
• The unit is on a stable and level surface ie on a firm flooring, not on a thick carpet
• Date and time settings
2. Visually inspect the following at least weekly:
• The display for any damage or contamination
• All cables, cords, and connector ends for damage or contamination
• All safety-related labeling for legibility
• All accessories (electrodes, etc.) for wear or damage
3. Visually inspect the following at least monthly:
• Mounting screws on stand
Update settings, replace items, or call for service as necessary according to the results of the visual
inspections. Do not use the unit if you see any signs of damage. Equipment that has been damaged
must be checked for proper operation by qualified personnel before using again.

Before Use
88
Part names & Connection Procedure
Symbols and their Meanings
Positive polarity Serial number
Alternating current Direct current
Input, Output
WEEE - Waste Electrical
and Electronic
Equipment Directives
Caution
Refer to the attached
notes.
For indoor use only
SD card See the instructions
Serial interface Manufacturer
*The SD logo is a registered trademark of the SD Association.
Remote display version Column mounted version
Control Box
Platform
Platform
Control box
Column
Accessories
This manual
AC adapter
AC cord
Printer paper
✓
Printer cover
Display
Operating key
DC jack for AC adapter
*RS-232C and USB-B are mutually exclusive.
RS-232C
USB-B
SD card slot
Adjustable feet x 4

Before Use
99
DC-360
Display and Keys
Meanings of the LED Indicators and Keys
Turn ON / OFF the power
Feeds the printer paper
Set various functions
Reset zero point
Set preset value
(Clothes weight)
Select measurement mode
Display Body Fat (percentage and mass)
*Not measured value but calculated value
Display Body Muscle (percentage and mass)
*Not measured value but calculated value
Display Body Water (percentage and mass)
*Not measured value but calculated value
Display Visceral
Fat Rating
Display Basal
Metabolic Rate
Indicates whether “Standard mode or Athletic
mode” is selected as the body type.
Indicates whether “Male or Female” is selected as
the gender
Enter the age between “5 to 99 years”
Enter the height between “90.0cm to 249.9cm”
Confirms the entered numerical value.

Before Use
1010
Setting the Printer Paper /
Setting an SD Card
1
2
3
Setting the Printer Paper
(1) Turn the power OFF and push the
handle on the left hand side of the
controller.
(2) Open the printer cover.
(3) Set the printer paper in place.
Peel off the adhesive and pull out approximately
10cm of paper from the printer paper.
(4) Replace the printer cover to its
original place.
Press to turn on the device.
After all the indicators are displayed, is
displayed.
If the device is turned on with the printer cover open,
is displayed.
Press . The printer paper is cut
automatically, and the settings are
completed.
Paper is not automatically cut when ‘Auto Cut’ is set
to ‘Off’.
②
①
③
④
Cutting automatically setting →See P.13 setting4
Note

Before Use
1111
1
2
3
Setting an SD Card
When the power is turned OFF, insert
an SD card into the slot with the logo
side facing upwards.
Make sure the card is facing in the correct direction
when inserting it.
Press to turn on the device.
When the device detects the SD card,
the mark on the right appears in the
upper left corner of the screen.
Be sure to insert and remove the SD card when the
device power is off to avoid damaging the SD card.
Compatible with SD and SDHC memory cards
Not compatible with SDXC memory cards
Note
Table of contents
Other Tanita Scale manuals

Tanita
Tanita WB-380 User manual

Tanita
Tanita BC-545F User manual

Tanita
Tanita BF-681 User manual

Tanita
Tanita UM-041F User manual

Tanita
Tanita SC-330ST User manual

Tanita
Tanita BF-679 User manual

Tanita
Tanita HD-351BT User manual

Tanita
Tanita UM-030 User manual

Tanita
Tanita innerscan User manual

Tanita
Tanita BD-585 User manual

Tanita
Tanita HD-361 User manual

Tanita
Tanita HA 502 User manual

Tanita
Tanita WB-100MA User manual

Tanita
Tanita WB-100ST Instruction Manual

Tanita
Tanita UM-040 User manual

Tanita
Tanita Ironman InnerScan BC 548 User manual

Tanita
Tanita BF-522 User manual

Tanita
Tanita HS-302 User manual

Tanita
Tanita KD-200 User manual

Tanita
Tanita MC-780MA-N User manual