QuickTake 15 Performance Prole
Flow range: 5 to 15 L/min
Flow compensation
accuracy: ± 5% of set ow
Compensating ow
back pressure range: 15 L/min at 8 inches water back pressure
10 L/min at 15 inches water back pressure
5 L/min at 20 inches water back pressure
Power: Rechargeable 7.2-volt lithium-ion (Li-Ion) battery (see Li-Ion Battery
Shipment on page 4), or battery with standard AC charger/adapter
(115 V), or battery with deluxe AC charger/adapter (100-240 V)
Battery recharge time: Approximately 5 hours
Typical Run time: VersaTrap Cassette: 4 hours at 15 L/min
Results when tested with a new pump and new fully charged
battery. Pump performance may vary.
The AC charger/adapter can be used to extend baery run
time, but it is not a baery eliminator. Therefore, it will
not provide indenite run times.
Temperature: Operating: 32 - 113 F (0 - 45 C)
Storage: -4 - 113 F (-20 - 45 C)
Charging: 32 - 113 F (0 - 45 C)
Protect sample pump from weather when in use outdoors.
Run time features: User-selectable features, user-adjustable presets (see Advanced
Setup)
Preset timed runs: 1, 2, 5, 10, or 15 min
Manually set continuous
run with manual shut-off: 1 to 999 min (repeats 1 to 999-min
runs indenitely until user stops
pump or power supply is depleted)
Intermittent sampling: See Advanced Setup
Flow fault features: If the pump is unable to compensate due to excessive back pressure,
the pump will go into ow fault mode (see page 12 for details). Flow
fault features are enabled as the default setting. Flow fault may be
disabled at the user’s option. See Advanced Setup.
Flow fault: Flow fault displays immediately.
Pump shuts off after 5 seconds
Fault restart: Pump attempts restart every
10 seconds up to ve times
Compare the information in this table to pump compensation range to determine appropriate applications.
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min) 5.0 8.0 10.0 12.0 15.0
Filter/Pore Size (µm)
37-mm MCE/0.8 11 18 22 28 36
37-mm PVC/5.0 4 7 9 11 15