Reed Instruments R7800 User manual

Coating
Thickness
Gauge
Instruction
Manual
R7800
REED Instruments

2
Table of Contents
Introduction ................................................................................................... 3
Product Quality.............................................................................................. 3
Safety ............................................................................................................ 4
Features......................................................................................................... 4
Included......................................................................................................... 4
Specications................................................................................................ 5
Instrument Description.................................................................................. 6
Display Description ....................................................................................... 6
Operating Instructions................................................................................... 7
Power ON/OFF......................................................................................... 7
Measuring Procedure............................................................................... 7
Setup Mode..............................................................................................8-12
Enabling/Disabling Audible Beep............................................................. 9
Setting the LCD Brightness...................................................................... 9
Selecting the Thickness Unit of Measure ................................................. 9
Selecting Calibration Mode...................................................................... 9
Setting Higher Limit Value ...................................................................... 10
Setting Lower Limit Value....................................................................... 10
Enabling/Disabling LED Indicators ......................................................... 11
Factory Reset ......................................................................................... 11
Selecting Continuous Measurement Mode............................................ 11
Delete Recorded Data............................................................................ 12
Minimum, Maximum, and Average Values .................................................. 12
Quick Decision Mode .............................................................................12-13
Single-Point Decision Configuration ...................................................... 13
Multi-Point Decision Configuration ........................................................ 13
continued...
REED Instruments

3
Zero Calibration......................................................................................14-16
Zero Calibration...................................................................................... 14
Two-Point Calibration ............................................................................. 15
Calibration Verification............................................................................ 16
Locking Screen Auto-Rotation.................................................................... 16
Data Upload ................................................................................................ 16
Software Installation.................................................................................... 16
Battery Replacement................................................................................... 17
Applications................................................................................................. 17
Accessories and Replacement Parts .......................................................... 17
Product Care ............................................................................................... 17
Product Warranty ........................................................................................ 18
Product Disposal and Recycling ................................................................. 18
Product Support.......................................................................................... 18
Introduction
Thank you for purchasing your REED R7800 Coating Thickness Gauge.
Please read the following instructions carefully before using your instrument.
By following the steps outlined in this manual your meter will provide years
of reliable service. This coating thickness gauge is designed for non-invasive
coating thickness measurements. The R7800 features zero or two point
calibration, user adjustable high/low alarms and a USB interface to further
analyze results in the software.
Product Quality
This product has been manufactured in an ISO 9001 facility and has
been calibrated during the manufacturing process to meet stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
REED Instruments

4
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
Features
• Automaticallydetectsferrousandnon-ferroussubstrates
• Easy-to-readcolorLCDdisplay
• 360°screenrotationallowsuserstoviewmeasuredreadingsfrom
any angle
• Durablegaugeandprobeconstruction
• Zeroandtwo-pointcalibrationensuresmeasurementaccuracy
• Singleormulti-pointmodeallowsausertosettolerancesandquickly
identify whether a coating passes or fails
• Tri-colorLEDsquicklyindicatewhenmeasurementsarebelow(red),
above(yellow)orwithinsetparameters(green)
• Internalmemorystoresupto500groupsofdata
•
View stored data, charts or real-time measurement with software
• Lowbatteryindicationandautoshutoff
Included
• FerrousSubstrate
• Non-FerrousSubstrate
• 5ReferenceFilms
• USBCable
• WristStrap
• SensorCover
• Batteries
• HardCarryingCase
REED Instruments

5
Specifications
Measuring Range: 0 to 1250µm
0.1 to 49.2mils
Accuracy: 0 to 1250µm: ±3% +1µm
0 to 49.21mils: ±3% +0.04mils
Resolution: 0 to 99.9µm: 0.1µm
100 to 1250µm: 1µm
0 to 4.99mils: 0.01mils
5.0 to 49.2mils: 0.1mils
Measurement Modes: Single and Continuous
Display: 4-digit color LCD display
Resolution: 320 x 240 pixels
BacklitDisplay: Yes(5levels)
Min/Max/Average Readings: Yes
MinimumCurvatureRadius: 5mm(196.85mils)
MinimumAreaDiameter: 20mm(787.40mils)
MinimumSubstrateThickness: 0.5mm(19.69mils)
AutoShut-Off: Yes(After5minutes)
Low Battery Indicator: Yes
Power Supply: 2 x AA Batteries
InternalMemory: Yes(Upto500groups)
PC Connectivity: USB Cable
Software: Yes
Product Certications: CE
RefreshRate: 0.5seconds(continuousmode)
OperatingTemperature: 32to104°F(0to40°C)
StorageTemperature: -4to140°F(-20to60°C)
OperatingHumidityRange: ≤80%RH
Dimensions: 5.9x2.6x1.38"(152x65x35mm)
Weight: 6.35oz(180g)
REED Instruments

6
Instrument Description
1. Wrist Strap Anchoring Point
2. LED Status Indicator
3. LCD Display
4. Power Button
5. Menu/Conrm/Calibration
Button
6. Increase Value/Up/Lock
Screen Button
7. USB Port
8. Cancel/Zero/Return Button
9. Decrease Value/Down/Quick
Decision Mode Button
10.BatteryCompartment(Back)
11. Sensor
Display Description
1. Low Limit Indicator
2. Measured Value
3. Minimum Value Indicator
4. Stored Measurements
Indicator
5. Average Measurement Value
6. Maximum Value Indicator
7. Battery Level Indicator
8. Locked Screen Indicator
9. Substrate Type
10. Unit of Measure
11. Calibration Mode Indicator
12.HighLimitIndicator
OK
MENU
CLEAR
R7800
Coating Thickness Gauge
2
1
5
3
48
7
9
11
6
10
100
Num:
Lo: 90.0
99.4 100
99.84
101Hi:
um/mil
NFe/Fe
Avg:
Max:Min:
1
4
2
75 6
3
8
9
11
12
10
REED Instruments

7
Operating Instructions
Power ON/OFF
To turn the meter ON or OFF press and hold the power button for 2 seconds.
Note: If the buzzer is enabled, there will be an audible beep when the
instrument is powered ON or OFF.
Measuring Procedure
Before each test session perform a Zero Calibration. See the Zero Calibration
section for details.
1. During start-up the meter begins an initialization process and enters
normal operation upon completion.
2. Select the closest reference lm to the estimated coating thickness
of the test area and perform a 2-point calibration on the appropriate
substrate plate.
3. Proceed with the 2-point calibration. See the Two-Point Calibration
section for details.
4.
Once the 2-point calibration has been completed, proceed with testing.
5. Select 3-5 measuring points on the test surface.
6. For each measuring point, it is recommended to take 5 measurements
and record the average value as indicated on the LCD display.
7. After each measuring point has been tested, the overall average value
should be noted as the coating thickness value of the test piece.
Note: Lightly press the sensor against the substrate being measured.
Keep the sensor in close contact with the surface of the substrate without
the use of excessive force as it can cause measurement errors. If the
measured value is greater than 1250μm and less than 1500μm, the LCD
displays "OL" indicating it is over the measurement range. If the measured
value is greater than 1500μm, the gauge will not respond.
REED Instruments

8
Setup Mode
1. Press the
OK
MENU
CLEAR
button to enter Setup Mode.
OK
MENU
ZERO
R7800
Coating Thickness Gauge
2/2
OK
MENU
ZERO
R7800
Coating Thickness Gauge
1/2
OK
MENU
ZERO
R7800
Coating Thickness Gauge
1/2
um
Fe
509
1/2
um
Fe
500
um
Fe
Calibration completed!
0.0
2/2
um
Fe
Calibration completed!
0.0
MENU
um/mil
2. Use the
OK
MENU
CLEAR
and
OK
MENU
CLEAR
arrows to scroll through the following parameters.
OK
MENU
CLEAR
Enabling/Disabling Audible Beep
OK
MENU
CLEAR
Setting the Brightness
um/mil Selecting the thickness Unit of Measure
OK
MENU
CLEAR
Selecting Calibration Mode
OK
MENU
CLEAR
SettingHigherLimitValue
OK
MENU
CLEAR
Setting Lower Limit Value
OK
MENU
CLEAR
Turning LED Indicators ON/OFF
OK
MENU
CLEAR
Factory Reset
OK
MENU
CLEAR
Selecting Continuous Measurement Mode
OK
MENU
CLEAR
Delete Recorded Data
3.
Follow the instructions below to adjust each setting
.
continued...
REED Instruments

9
Enabling/Disabling Audible Beep
1. Press the OK button when the audible beep icon
OK
MENU
CLEAR
is selected in
Setup Mode.
2. Use the
OK
MENU
CLEAR
and
OK
MENU
CLEAR
arrows to select between ON
OK
MENU
CLEAR
or OFF
OK
MENU
CLEAR
X.
3. Press the OK button to conrm selection.
4. Press the
OK
MENU
CLEAR
button to exit Setup Mode and resume normal operation.
Setting the LCD Brightness
1. Press the OK button when the brightness icon
OK
MENU
CLEAR
is selected in
Setup Mode.
2.
Use the
OK
MENU
CLEAR
and
OK
MENU
CLEAR
arrows to increase or decrease the brightness level.
3. Press the OK button to conrm selection.
4. Press the
OK
MENU
CLEAR
button to exit Setup Mode and resume normal operation.
Selecting the Thickness Unit of Measure
1. Press the OK button when the unit of measure icon um/mil is selected
in Setup Mode.
2. Use the
OK
MENU
CLEAR
and
OK
MENU
CLEAR
arrows to select between μm and mil.
3. Press the OK button to conrm selection.
4. Press the
OK
MENU
CLEAR
button to exit Setup Mode and resume normal operation.
Selecting Calibration Mode
1. Press the OK button when the calibration mode icon
OK
MENU
CLEAR
is selected
in Setup Mode.
2. Use the
OK
MENU
CLEAR
and
OK
MENU
CLEAR
arrows to select between zero and
2-point calibration.
3. Press the OK button to conrm selection.
4. Press the
OK
MENU
CLEAR
button to exit Setup mode and resume normal operation.
continued...
REED Instruments

10
Setting Higher Limit Value
1. Press the OK button when the higher limit value icon
OK
MENU
CLEAR
is selected
in Setup Mode.
2. Use the
OK
MENU
CLEAR
and
OK
MENU
CLEAR
arrows to increase or decrease the higher
limit value.
3. Press the OK button to conrm selection.
4. Press the
OK
MENU
CLEAR
button to exit Setup Mode and resume normal
operation.
Note: If the higher limit value is exceeded while in normal measuring
modeandLEDindicatorsareenabled(SeeEnabling/Disabling LED
Indicatorsfordetails),theLEDwarninglightwillashyellow.When
measurements are within set parameters, the LED warning light
ashesgreen.
Setting Lower Limit Value
1. Press the OK button when the lower limit value icon
OK
MENU
CLEAR
is selected in
Setup Mode.
2. Use the
OK
MENU
CLEAR
and
OK
MENU
CLEAR
arrows to increase or decrease the lower
limit value.
3. Press the OK button to conrm selection.
4.
Press the
OK
MENU
CLEAR
button to exit Setup Mode and resume normal operation.
Note: If the lower limit value is exceeded while in normal measuring mode
andLEDindicatorsareenabled(SeeEnabling/Disabling LED Indicators for
details),theLEDwillashred.Whenmeasurementsarewithinset
parameters,theLEDwarninglightashesgreen.
continued...
REED Instruments
Table of contents
Other Reed Instruments Test Equipment manuals

Reed Instruments
Reed Instruments R5710 User manual

Reed Instruments
Reed Instruments TM-8811 User manual

Reed Instruments
Reed Instruments R5820 User manual

Reed Instruments
Reed Instruments R2810 User manual

Reed Instruments
Reed Instruments R5700 User manual

Reed Instruments
Reed Instruments R9030 User manual

Reed Instruments
Reed Instruments R5500 User manual

Reed Instruments
Reed Instruments R5320 User manual

Reed Instruments
Reed Instruments R7200 User manual

Reed Instruments
Reed Instruments R5800 User manual