Reed Instruments R2810 User manual

Thermocouple
Calibrator
Instruction
Manual
R2810
www. .com information@itm.com1.800.561.8187

2
Table of Contents
Introduction ................................................................................................ 2
Product Quality........................................................................................... 3
Safety ......................................................................................................... 3
Features...................................................................................................... 4
Included...................................................................................................... 4
Specications..........................................................................................4-5
Output Functions ................................................................................... 4
Measure Functions................................................................................. 5
General Specifications ........................................................................... 5
Instrument Description ............................................................................... 6
Display Description .................................................................................... 7
Operating Instructions...........................................................................8-11
Power ON/OFF ...................................................................................... 8
Auto Power-Off ...................................................................................... 8
Selecting the Tempreture Unit of Measure (°C/°F)................................. 8
Output Function..................................................................................... 9
Simulating DC Voltage Output ............................................................... 9
Simulating Thermocouple (TC) Output .............................................9-10
Measure Function ...........................................................................10-11
Warning............................................................................................ 10
Caution........................................................................................10-11
Measuring DC Voltage ......................................................................... 11
Measuring Thermocouples .................................................................. 11
Battery Replacement................................................................................ 12
Applications.............................................................................................. 12
Accessories and Replacement Parts ....................................................... 13
Product Care ............................................................................................ 13
Product Warranty ..................................................................................... 13
Product Disposal and Recycling .............................................................. 14
Product Support....................................................................................... 14
www. .com information@itm.com1.800.561.8187

3
Introduction
Thank you for purchasing your REED R2810 Thermocouple Calibrator.
Please read the following instructions carefully before using your
instrument. By following the steps outlined in this manual your meter will
provide years of reliable service.
Product Quality
This product has been manufactured to meet the stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center. To
avoid injury to the user or damage to the instrument, please read the
safety information below before initial use:
• Donotoperatetheinstrumentaroundammableorexplosivegas,
vapor or dust.
• Neverapplymorethan30Vbetweenanytwoterminals,orbetweenany
terminal and ground terminal.
Note: For optimal accuracy, allow the instrument to warm up 5 minutes
before operating. If the automatic reference-junction temperature
compensation of the instrument deviates from its designed accuracy,
contact an Authorized REED Service Center.
www. .com information@itm.com1.800.561.8187

4
Features
• Sourceandmeasure8thermocoupletypes,includingR,S,B,E,K,
J, T, N; and volts
• Basicaccuracyof0.05%
• Internalcoldjunctioncompensation
• Easytoread6-digitLCDdisplay
• Zeroadjustmentbutton
• Userselectableunitofmeasure(°F/°C)
• Lowbatteryindicatorandautoshutoff
Included
• TestLeads
• AlligatorClips
• ThermocoupleAdapter
• ProtectiveHolster
• Batteries
Specifications
Output Functions
Accuracyspeciedat23°C±5°C&75%RHforaperiodofoneyear
after calibration.
Output Range Output Range Resolution Accuracy
DCV 100mV -10.00mVto110.00mV 0.01mV 0.05%rdg.+30µV
1V -0.1000Vto1.1000V 0.1mV 0.05%rdg.+0.3mV
TC
R-40to1760°C 1°C ±0.05%rdg.+3°C
(≤100°C)±0.05%rdg.
+2°C(>100°C)
S-20to1760°C 1°C
B400to1800°C 1°C
±0.05%rdg.+3°C
(≤600°C);±0.05%rdg.
+2°C(>600°C)
E-200.0to1000.0°C 0.1°C
±0.05%rdg.+2°C
(≤-100°C)±0.05%rdg.
+1°C(>-100°C)
K -200.0to1370.0°C 0.1°C
J-200.0to1200.0°C 0.1°C
T-200.0to400.0°C 0.1°C
N-200.0to1300.0°C 0.1°C
continued...
www. .com information@itm.com1.800.561.8187

5
Measure Functions
Output Range Output Range Resolution Accuracy
DCV 100mV -10.00mVto110.00mV 0.01mV 0.05%rdg.+30µV
TC
R-40to1760°C 1°C ±0.05%rdg.+3°C
(≤100°C)±0.05%rdg.
+2°C(>100°C)
S-20to1760°C 1°C
B400to1800°C 1°C
±0.05%rdg.+3°C
(≤600°C);±0.05%rdg.
+2°C(>600°C)
E-200.0to1000.0°C 0.1°C
±0.05%rdg.+2°C
(≤-100°C)±0.05%rdg.
+1°C(>-100°C)
K -200.0to1370.0°C 0.1°C
J-200.0to1200.0°C 0.1°C
T-200.0to400.0°C 0.1°C
N-200.0to1300.0°C 0.1°C
General Specifications
Display: 6-DigitLCD
ZeroAdjustmentButton: Yes
KickStand: Yes
Power Supply: 2 AA Batteries
Cold Junction
Compensation: Yes
Autoshutoff: Yes(after15minutes/off)
LowBatteryIndicator: Yes
ReplaceableTestLeads: Yes
Product Certications: CE
OperatingTemperature: 32to122°F(0to50°C)
OperatingHumidityRange: 0to85%
StorageTemperature: 14to122°F(-10to50°C)
Dimensions: 7.1x3.5x1.9"(180x90x47mm)
Weight: 8.2oz(500g)
www. .com information@itm.com1.800.561.8187

6
Instrument Description
R2810 Thermocouple Calibrator
Ω°C°F
mV
Pt1000 Pt200 Pt500 Cu50 Cu10
OUTPUT
RJ-ON
1
2
3
4
5
6
7
8
9
10
OUTPUT
INPUT
1. LCDDisplay
2. POWER Button
3. Input/OutputButton
4. ZeroResetButton
5. Function Selection Output Button
6. Range Selection Button
7. RJ-ONButton(T/CColdJunctionCompensationButton)
8. OutputValueSettingButton
9. Output Digit Selection Button
10.Input/Outputterminals
www. .com information@itm.com1.800.561.8187

7
Display Description
C°F
mV
OUTPUT INPUT
RJ-ON
OUTPUT Indicates that the instrument is in output mode.
INPUT
Indicates that the instrument is in input mode.
Indicates that the instrument is performing its reference cold
junction compensation.
Indicates battery power is low and batteries need to
be replaced.
Indicates which output digit is being set.
Indicates unit of measure of current output value.
Indicates thermocouple type.
www. .com information@itm.com1.800.561.8187

8
Operating Instructions
Power ON/OFF
Press the POWER button to turn the instrument on. To turn the instrument
off, press and hold the POWER button.
Auto Power-Off
As a default, the instrument will automatically turn off after 15 minutes of
inactivity. To turn off this feature:
1. Turn the meter off.
2. PressthePOWERbutton(todisplaythefullscreen).
3. Quickly press the RANG button when the instrument is in the
maintenancestateasindicatedby"1.8.8.888".
4. "AP-XX"willnowappearonthedisplay.
5. Press the buttontotogglebetween"AP-ON"&"AP-OF"."AP-OF"
indicates that the no automatic power-off function is disabled, while
"AP-ON"indicatestheautomaticpower-offfunctionisenabled.
6. Press the RJ-ON button to save the required setting.
7. PressandholdthePOWERbuttontoexitthemaintenancestateand
turn the instrument off.
Selecting the Temperature Unit of Measure (°C/°F)
1. Turn the meter off.
2. PressthePOWERbutton(todisplaythefullscreen).
3. Quickly press and hold the RANG button when the instrument is in
themaintenancestateasindicatedby"1.8.8.888".
4. "AP-XX"willnowappearonthedisplay.
5. Press the FUNCbuttononceandyouwillsee"tC-0C"conrming
that the instrument is measuring in Celsius.
6. Press the and buttonstotogglebetween"tC-0F"forFahrenheit
and"tC-0C"forCelsius.
7. Press the RJ-ON button to save the required setting.
8. PressandholdthePOWERbuttontoexitthemaintenancestateand
turn the instrument off. continued...
www. .com information@itm.com1.800.561.8187

9
Output Function
• TheoutputterminaloftheinstrumentcanproduceDCvoltagesorcan
simulate thermocouple temperature set by the user.
Do not apply any voltage to the output terminal during the operation.
If any improper voltage is applied to the output terminal, it will cause
damage to the internal circuit.
Simulating Output from DC Voltage
1. Insert one end of the test lead into
theoutput(OUTPUT)terminalofthe
meter and connect the other end to
the input terminal of the instrument
under test. See diagram on right.
2. Press the FUNC button to select the
VDCfunctionand'V'willappearon
the display.
3. Press the RANG button to select the
outputrangeof0.0000Vor00.00mV.
4. Press the button to select the set digits for output.
5. Press the button to change the numerical value of the set digits.
(Numericalvaluecannotbechangedbeyondthemaximumrange
ofthemeter).
6. Press the ZERO button and the output will be set to
00.00mVor0.0000V.
Simulating Output from Thermocouple (TC)
1. InsertoneendofthetestleadintotheOUTPUTterminalofthe
calibrator and connect the other end to the input terminal of the
instrument under test. See diagram above.
2. Press the INPUT/OUTPUT button to select the Output function.
3. Press the FUNC button to select Thermocouple output.
'R'and'°C'willappearonthedisplay.
4. Press the RANG button to select the type of thermocouple.
5. Press the button to move the on-screen cursor in order to
select the desired digit on the display.
continued...
www. .com information@itm.com1.800.561.8187

10
6. Press the button to change the numerical value of each digit.
(Numericalvaluecannotbechangedbeyondrangemaximum.)
7. Automatic compensation for cold junction temperature.
Note: While calibrating an instrument with cold junction temperature
compensation, press the RJ-ON button so that the meter can start the
function of automatic cold junction compensation. The meter will output
the necessary temperature thermoelectric force, and display 'RJ-ON':
Output thermoelectric force = corresponding thermoelectric force
of set temperature – thermoelectric force of room temperature
• Ittakestwosecondsfortheinstrumenttostartitsinternal
reference-junction temperature. After this, each automatic
compensation occurs at 10 second intervals.
• Ifthereisachangeintheoperatingambienttemperature,donotstart
the operation until the built-in compensating sensor has become stable
(approx.10minutes).
• Ifthereisnoneedforthecalibratortoperformautomatic
reference-junction compensation, press the RJ-ON button and
the symbol RJ-ON will no longer appear in the display.
8. Press the ZERO button and the output will be directly set to
0000°C(RorStype),400°C(Btype)or0000.0°C(othertypes).
Measure Function
Warning
Themaximumvoltageallowedbetweentheterminalsandbetweenthe
terminalsandthegroundis30V,allvoltagesexceedingthespecied
voltage may cause damage to the meter and injury to the user.
Caution
• Donotapplyanyvoltageexceedingthemaximumallowedtotheinput
terminals, as this may cause damage to the meter.
• Pleasepoweroffinstrumentundertestwhenconnectingtotheinput
terminal. Connection to the instrument under test with power may
cause damage to the meter.
continued...
www. .com information@itm.com1.800.561.8187
Table of contents
Other Reed Instruments Test Equipment manuals

Reed Instruments
Reed Instruments R5800 User manual

Reed Instruments
Reed Instruments BX-500 User manual

Reed Instruments
Reed Instruments R9030 User manual

Reed Instruments
Reed Instruments R5320 User manual

Reed Instruments
Reed Instruments R5710 User manual

Reed Instruments
Reed Instruments R5820 User manual

Reed Instruments
Reed Instruments R7200 User manual

Reed Instruments
Reed Instruments TM-8811 User manual

Reed Instruments
Reed Instruments R5700 User manual

Reed Instruments
Reed Instruments R5600 User manual