Pyxis LS-200 User manual

Table of Contents
1 Introducon 2
2 Specicaons 2
3 Unpacking Instrument 2
3.1 Standard Provided Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3.2 OponalAccessories ..................................... 3
4 Installaon 3
4.1 Tank Top Installaon and Precauons . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
4.2 Wiring............................................. 4
5 Instrument Overview 5
6 Setup with uPyxis® Mobile App 6
6.1 Download uPyxis® Mobile App . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
6.2 Connecng to uPyxis® Mobile App . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
6.3 OverviewScreen ....................................... 7
6.4 ReadingScreen ........................................ 8
6.5 SengsScreen ........................................ 8
7 Setup with uPyxis® Desktop App 9
7.1 InstalluPyxis®DesktopApp.................................. 9
7.2 Connecng to uPyxis® DesktopApp ............................. 10
7.3 OverviewScreen ....................................... 10
7.4 ReadingScreen ........................................ 11
7.5 SengScreen......................................... 12
8 Outputs 13
8.1 4–20mAOutputSetup .................................... 13
8.2 Communicaon Using Modbus RTU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
9 Sensor Maintenance and Precauon 13
10 Regulatory Approval 14
11 Contact Us 14

Warranty Informaon
Condenality
The informaon contained in this manual may be condenal and proprietary and is the property of Pyxis
Lab, Inc. Informaon disclosed herein shall not be used to manufacture, construct, or otherwise reproduce
the goods described. Informaon disclosed herein shall not be disclosed to others or made public in any
manner without the express wrien consent of Pyxis Lab, Inc.
Standard Limited Warranty
Pyxis Lab warrants its products for defects in materials and workmanship. Pyxis Lab will, at its opon, repair
or replace instrument components that prove to be defecve with new or remanufactured components
(i.e., equivalent to new). The warranty set forth is exclusive and no other warranty, whether wrien or oral,
is expressed or implied.
Warranty Term
The Pyxis warranty term is thirteen (13) months ex-works. In no event shall the standard limited warranty
coverage extend beyond thirteen (13) months from original shipment date.
Warranty Service
Damaged or dysfunconal instruments may be returned to Pyxis for repair or replacement. In some in-
stances, replacement instruments may be available for short duraon loan or lease.
Pyxis warrants that any labor services provided shall conform to the reasonable standards of technical com-
petency and performance eecve at the me of delivery. All service intervenons are to be reviewed and
authorized as correct and complete at the compleon of the service by a customer representave, or des-
ignate. Pyxis warrants these services for 30 days aer the authorizaon and will correct any qualifying
deciency in labor provided that the labor service deciency is exactly related to the originang event. No
other remedy, other than the provision of labor services, may be applicable.
Repair components (parts and materials), but not consumables, provided during a repair, or purchased
individually, are warranted for 90 days ex-works for materials and workmanship. In no event will the in-
corporaon of a warranted repair component into an instrument extend the whole instrument’s warranty
beyond its original term.
Warranty Shipping
A Repair Authorizaon (RA) Number must be obtained from Pyxis Technical Support before any product can
be returned to the factory. Pyxis will pay freight charges to ship replacement or repaired products to the
customer. The customer shall pay freight charges for returning products to Pyxis. Any product returned to
the factory without an RA number will be returned to the customer. To receive an RMA you can generate a
request on our website at hps://pyxis-lab.com/request-tech-support/.
Pyxis Technical Support
Contact Pyxis Technical Support at +1 (866) 203-8397, [email protected], or by lling out a request for
support at hps://pyxis-lab.com/request-tech-support/.

1 Introducon
The Pyxis LS-200 is a general-purpose ultrasonic sensor. It provides connuous level measurement up to 86
inches (7.2 or 2.2 m) with a 4–20mA output, RS-485, and Bluetooth digital outputs. It can be congured
via the uPyxis® App on your mobile phones or computers. The sensor is powered by a 24 VDC external
power supply. This non-contact liquid level sensor is well suited for corrosive liquids and can be used for
industrial and municipal liquid storage and chemical feed applicaons.
2 Specicaons
Table 1. LS-200 Specicaons
Specicaon*LS-200
Part Number (P/N) 54011
Range 4–86 inch (0.01–2.2 m)
Resoluon 0.02 inch (0.5 mm)
Accuracy ±0.15% of the range
Measurement
Interval
Connuous, 10 or 30 seconds, 3 or 30 minutes, 1 or 4 or 6 hours, or stopped
Output Bluetooth 5, 32 (10 m) Line of Sight, 4–20mA Analog Output, RS-485 Digital
Output with Modbus protocol
Installaon 1” male NPT
Cable Length 9.8 (3 m), extension cable available
Power Supply 24 VDC, 2 W
Dead Zone 4 inch (10 cm)
Dimension (H ×
Dia)
4.33 ×2.95 inch (110 ×75 mm)
Weight 0.45 lbs (210 g)
Enclosure Material Polycarbonate (PC)
Transducer Material Polyvinylidene Fluoride (PVDF)
Operaonal
Temperature
5–122 °F (-15–50 °C)
Storage
Temperature
-4–140 °F (-20–60 °C)
Pressure 14–30 psi (0.1–0.2 MPa)
Enclosure Rang IP67
Regulaon CE
*With Pyxis’s connuous improvement policy, these specicaons are subject to change without noce.
3 Unpacking Instrument
Remove the instrument and nd the standard accessories from the shipping container as listed below. In-
spect each item for any damage that may have occurred during shipping. Verify that all accessory items are
included. If any item is missing or damaged, please contact Pyxis Lab Customer Service at service@pyxis-
lab.com or by phone at +1 (866) 203-8397.

3.1 Standard Provided Accessories
• Waterproof Cable Adapter/Flying Leads (10 ) P/N: 50774
• User Manual available online at hps://pyxis-lab.com/product/ls-200-ultrasonic-level-sensor/
3.2 Oponal Accessories
The following oponal accessories can be ordered from Pyxis Customer Service (order@pyxis-lab.com) or
Pyxis eStore at hps://pyxis-lab.com/shop/.
Figure 1.
4 Installaon
4.1 Tank Top Installaon and Precauons
The sensor should be installed to a 1-inch bulkhead ng on the top of the tank. The major dimensions of
the sensor are shown in Figure 2. If a at horizontal surface is not available on the top of the tank, please use
a self-aligning bulkhead ng so that the sensor can be adjusted to be perpendicular to the liquid surface.
• Install and adjust the sensor to be perpendicular to liquid surface
• Installaon locaon shall not be too close to container wall to avoid interference.
• The sensor has a 3.94 inch (10 cm) dead zone (DZ). Raise the probe to avoid the DZ if desired
• Do not install in a locaon which will cause the ultrasonic wave to be obstructed
• Do not install the sensor in a vacuum environment
Figure 2. Installaon illustraon

Figure 3. Dimensions, inch (mm)
4.2 Wiring
If the power ground terminal and the negave 4–20mA terminal in the controller are internally connected
(non-isolated 4–20mA input), it is unnecessary to connect the 4–20mA negave wire (green) to the 4–20mA
negave terminal in the controller. If a separate DC power supply other than that from the controller is used,
make sure that the output from the power supply is rated for 22–26 VDC @ 65 mA.
*NOTE*The negave 24V power terminal (power ground) and the negave 4–20mA ter-
minal on the LS-200 are internally connected.
Follow the wiring table below to connect the LS-200 to a controller:
Table 2.
Wire Color Designaon
Red 24V +
Black 24V Power ground
White 4–20mA +
Green*4–20mA -
Blue RS-485 A
Yellow RS-485 B
Clear Shield, earth ground
*Internally connected to the power
ground

5 Instrument Overview
The indicators on the top of the sensor (Figure 4) are used to indicate power status and Bluetooth connec-
on status.
Table 3.
LED Status On O
Green LED Power Supply is on Power Supply is o
Blue LED Bluetooth is connected Bluetooth is disconnected
*NOTE*The Bluetooth connected LED will illuminate within 30 seconds of powerup of the
device. This feature is designed to minimize controller power draw during startup.
Figure 4. Sensor connecon and indicators

6 Setup with uPyxis® Mobile App
6.1 Download uPyxis® Mobile App
Download uPyxis® Mobile App from Apple App Store or Google Play.
Figure 5.
6.2 Connecng to uPyxis® Mobile App
Connect the LS-200 sensor to a mobile smart phone according to the following steps:
1. Open uPyxis® Mobile App.
2. On uPyxis® Mobile App, pull down to refresh the list of available Pyxis devices.
3. If the connecon is successful, the LS-200 and its Serial Number (SN) will be displayed (Figure 6).
4. Press on the LS-200 image.
Other manuals for LS-200
2
This manual suits for next models
1
Table of contents
Other Pyxis Accessories manuals

Pyxis
Pyxis ST-500 User manual

Pyxis
Pyxis CR Series User manual

Pyxis
Pyxis ST-525 User manual

Pyxis
Pyxis ST-772 User manual

Pyxis
Pyxis LS-202 User manual

Pyxis
Pyxis ST-773 User manual

Pyxis
Pyxis ST-590 Series User manual

Pyxis
Pyxis ST-772 User manual

Pyxis
Pyxis ST-600 User manual

Pyxis
Pyxis LT-73 Series User manual

Pyxis
Pyxis ST-700 Series User manual

Pyxis
Pyxis ST-565 User manual

Pyxis
Pyxis LSP Series User manual

Pyxis
Pyxis ST-601 User manual

Pyxis
Pyxis ST-765SS Series User manual

Pyxis
Pyxis ST-765SS-SO3 User manual

Pyxis
Pyxis ST-500W User manual

Pyxis
Pyxis ST-588 User manual

Pyxis
Pyxis LS-202 User manual

Pyxis
Pyxis ST-565 User manual