DP-16V HIGH VOLTAGE DIFFERENTIAL PROBE
1. FEATURES
The DP-16V differential voltage probe provides a safety means of
measuring floating potentials for all models of oscilloscopes
incomplete safety.
It converts the high differential voltage (≦16KV peak) into a low
voltage (≦4.0V) with reference to the earth for display on the
oscilloscopes. It is designed specially for high voltage using.
Frequency is up to 100MHz.
The BNC output is designed to operate on an input with an
impedance of 1 MΩ. It is 2 times of the 50Ω.
DP-16V can observe more exact measured testing voltage from
DMM. (Oscilloscope accuracy is 3%, and DMM is 10 times).
2. SPECIFICATIONS
(1) Bandwidth:
DC to 100 MHz (-3 dB)
(2) Attenuation: x 10, x 100, x1000
(3) Accuracy: ± 2%
(4) Voltage Input Ranges (DC + AC peak to peak)
≦ 16KV for x 2000, (i.e about 5.6KV RMS or ±8KV DC)
≦ 1.6KV for x 200, (i.e about 560V RMS or ±800V DC)
≦ 160V for x 20, (i.e about 56V RMS or ±80V DC)
(5) Permitted Max Input Voltage
Max differential voltage: 16KV (DC + AC peak to peak)
Max voltage between each input terminal and ground:
8KV RMS
(6) Input Impedance:
Differential: 30MΩ // 1.5 PF
Between terminals and ground: 15MΩ // 3 PF
2