PENKO 960 User manual

1. Indication of LEDs
2. Push button functions
3. Power and load cell connection
4. First use of Indicator
5. Digital inputs and outputs
6. RS232 comport
7. Standard Factory Settings
For more information visit
www.penko.com - technical support - literature library
960
PENKO Engineering BV
The Leading Experts In Weighing & Dosing
Indicator
Quick Start

960 Indicator
2
1. Indications of LEDs
1. Weigher stable
2. Zero active
3. Net weight active
4. Tare active
5. Code active
6. Piece counting active
7. Setpoint 1 (e1) active
8. Both setpoint active (e3)
9. Setpoint 2 (e2) active
10. Battery charge level
1
2
4
3
5
7
8
9
6
10

960 Indicator
3
3. Load cell / power connection
2. Push button functions
Tare/Up
Preset Tare/Down
Zero
Print/Left
Enter, press >2s =>Menu
On-Off/Esc
Power cable needs to be filtered
with ferrite bead for CE approval.
Make sure the load cell is connected before the power is switched on!

4. First use of indicator
960 Indicator
4
To make the indicator ready for first use set the maximum net weight value, display
step size, decimal point position and calibration.
The start
Turn on the indicator using the ON-OFF button.
Set up menu
To enter the Set up menu the service code method is used. Press the keys in the
correct order, all within 7 seconds.
Enter the keys in the following order:
Use UP and DOWN to select a menu or to edit a value. Use LEFT to select a digit
within a value.
Use ENTER to enter a menu or to confirm a setting. Use ESC to leave a setting or
a menu.
4
63
6
52
Key number

960 Indicator
5
First use of indicator -continue-
Indicator Settings
Indicator settings are used to set the maximum net weight value (Ind 1), display step
size (Ind 5), and decimal point position (Ind 6).
First, enter the set up menu by the given code on page 4.
The following screen is visible:
Maximum net weight value
To enter the Indicator settings, press ENTER.
The following screen is visible:
Use Ind 1 to set the maximum net weight value. Set maximum load to prevent
overload by the user. The indicator will not show any weight above this value.
Range: 0 – full display.
To change the value, enter Ind 1.
Use the UP, DOWN and LEFT keys to enter the value. The UP and DOWN keys are
used for changing the number (1-9), the LEFT key is used for changing the position
of the cursor. Confirm the value with ENTER.

960 Indicator
6
First use of indicator -continue-
Indicator Settings -continue-
When the maximum net weight value is set succesfully, the following screen is
visible:
Display step size
Use the UP key to go to Ind 5 and press ENTER
Use Ind 5 to set the display step size. The step size defines the scaled parts of the
weight value. The display value will be rounded off to the nearest value with a valid
step size.
Use the UP and DOWN keys to select the correct step size.
Choose between 1, 2, 5, 10, 20, 50 and confirm with ENTER
Done succesfully, the following screen is visible:

960 Indicator
7
First use of indicator -continue-
Indicator Settings -continue-
Decimal point
To set the Decimal point position, enter Ind 6.
The following screen is visible:
Press the UP and DOWN keys to define the point position and confirm by ENTER
Done succesfully, the following screen is visible:
Press ESC to go back to the Ind menu

960 Indicator
8
First use of indicator -continue-
Calibration Settings
Calibration settings are used to check, delete and set calibration points for the 960
indicator.
First, enter the set up menu by the given code on page 4 and press the UP key to go
to the CAL menu. If you are already in the set up menu, use the UP key to go to the
CAL menu.
Check and delete calibration point
To enter the Calibration settings, press ENTER
The following screen is visible:
Use the UP key to go to CAL 3 and press with ENTER
Use CAL 3 to check and delete all existing calibration points. Step through the
calibration points with the UP and DOWN key. Delete a calibration point by pressing
the ESC key > 3 seconds.
>3 sec
During deletions, the following screen is visible:

960 Indicator
9
First use of indicator -continue-
Calibration Settings -continue-
When a number is shown, the deletion of one calibration point is completed and
more points need to be deleted. Press ESC >3 sec to do so.
> 3 sec
When all calibration points are deleted, the following screen is visible:
Entering new calibration points.
Use the DOWN key to go to CAL 1 and press ENTER.
After entering, the following screen is visible:
and will automatically jump to
First calibrate the zero point (CP1). Make sure the weigher is unloaded. Press the
ENTER key.

960 Indicator
10
First use of indicator -continue-
Calibration Settings -continue-
The indicator now shows CP2 to calibrate the gain point (CP2).
and will automatically jump to
Use the UP, DOWN and LEFT keys to enter the reference value. The UP and
DOWN keys are used for changing the number (1-9), the LEFT key is used for chan-
ging the position of the cursor. Load the weigher with the reference value and press
the ENTER key.
Done succesfully, the following screen is visible:
Press ESC to go back to the Cal menu
Press ESC to go back to the weighing screen