MRC MH-4D User manual

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Digital Hotplate
With Stirrer
MH-4D
Operation Manual

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Important Before using this product, read this
entire operation manual carefully. Users should
follow all of the operational guidelines contained in
this manual and take all necessary safety
precautions while using this product. Failure to
follow these guidelines could result in potentially
irreparable bodily harm and/or property damage.
Caution All internal adjustments and maintenance must be
performed by qualified service personnel.
Material in this manual is for information purposes only. MRC is
committed to a continuing program of product development and
improvement, and reserves the right to change information, such as
specifications, appearance, and dimensions, described in this
document without notice. MRC makes no representations or
warranties with respect to this manual. In no event shall MRC be held
liable for any damages, direct or incidental, arising out of or related
to the use of this manual.
This manual contains important safety and operation information.
You must carefully read, understand, and follow all the instructions in
this manual prior to operating this instrument. Keep this manual in a
safe place nearby for reference and make it easily available to all
users.
1) This manual highlight DANGER/WARNING/CAUTION/NOTICE
alerts to prevent injury or property damage and also to achieve
optimum performance of your instrument.
2) These alerts are classified into four types in this manual
depending on the importance and the risk levels as described
below
3) The claim which is out of the quality guarantee published by
the Manufacturer is out of Manufacturer’s responsibility.
4) The damage which is from unexpected fault or damage of user
by Acts of God is out of Manufacturer’s responsibility.

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Table of content
Section 1 Warning and Cautions………….…..……………………………1-1
Section 2 General Description ………….………………..…………………2-1
Features
Safety
Convenience
Construction
Section 3 Unpacking and installation…………………….……….………3-1
Installation environment
Connecting to Power Supply
Location Condition
Connecting to the main power
Section 4 Operation……………………………………….…………………….4-1
Power On
Change the Set Value (SV)
Change the Alarm Value
Autotuning (AT)
Temperature Offsetting (OFFSET)
Stir Control Knob
External Temperature Probe Input (optional)
Section 5 Maintenance………………………..…………….……….….………5-1
Cleaning Product
Relocation
Keeping Product
Section 6 Troubleshooting …………….……………………………….….…6-1
Section 8 Warranty Information………….…………………………..….….7-1

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Section 1 Warnings and Cautions
DANGER
Ignoring the following warnings could cause serious injuries or
even fatal accidents.
Understand all alert labels prior to operating this equipment.
Never remove or damage the alert labels.
Never move this equipment while under operation.
Never install or use this equipment in explosive or flammable
atmospheres.
Never use or store flammable or hazardous substances near to
this equipment.
Never use the equipment to heat flammable or hazardous
substances.
Always wear proper protective gear (such as goggles, gloves,
masks or an apron) when using this equipment.
Never allow water, organic solvents, dust, or corrosive gases to
get inside the equipment.
Never immerse this equipment in water or operate it under
water.
Never reach for this equipment if it has fallen into water.
Check electrical requirements described in this operation
manual or on the ID plate of this equipment before use.
Connect this equipment to a dedicated power outlet nearby.
Connect this equipment only to properly grounded power
outlets to protect you and your instrument. Do not ground to
gas pipes or water pipes.
Never install this equipment closer than 30 cm (12 inches) to a
wall of combustible material. Observe minimum distances (in
general, 30 cm) from other devices.
If you observe strange smoke, odor or noise from the
instrument, unplug the power cord immediately and turn off
the main power switch. After the smoke or odor disappears,
contact your dealer or MRC if any repair is required.

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Never disassemble, repair, or modify this equipment on your
own. Doing so will void your warranty and may result in injuries
or product damages.
Do not heat any substances above temperature which will
cause hazards of explosion, implosion or release of toxic or
flammable gases from the material being heated.
Do not use in the presence of pressurized or sealed containers.
Fire or explosion may result causing injury or death
Do not operate the plate with external temperature probe
in the air or out of the liquid.
The external probe must be properly placed into a liquid
on the top surface during the whole process, to prevent
damage to the plate
The plate might be covered with aluminum foil to protect
it, DO NOT cover any other parts of the device and make
sure the radiation sheets underneath are not covered.
The Aluminum topped plates are not designed to be used with
metal containers. For heating of such containers it's advisable
to use a ceramic top hot plate.
To avoid electric shock, always use a property grounded
electrical outlet of correct voltage, and correct handling
capacity equipment
CAUTION
Ignoring the following cautions could cause injuries or property
damages.
Keep away the power cord from the hot plate, to prevent
melting and damages.
Do not touch the top plate or any object near it even when the
heater is turned off. You may get burned due to residual heat.
After using this equipment, make sure to turn off the main
power switch and also to disconnect the power cord from the
power outlet for the safety of other users.

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Do not place heavy objects, including this equipment, on top
of the power cord and do not strip, scratch, bend, twist, pull, or
heat the power cord. A damaged power cord is a fire and
electrical shock hazard.
Make sure to set up this equipment on a flat, stable, clean,
non-slip, dry, and fireproof surface inside a lab with proper
safety measures.
Do not touch the power outlet, power socket, or power cord
with wet hands. And make sure to connect the power cord
directly and firmly to the power outlet and power socket.
Do not put or insert any objects (especially if conductive or
flammable) inside this equipment.
Do not expose this equipment to any heat sources including
direct sunlight.
Mechanical shock or vibration can damage this equipment. Pay
extra attention while moving it. Damages caused by
mechanical shock or vibration may result in injury or fire.
Do not impact the top plate or heat sink. You can damage the
equipment or get injured.
Do not install this equipment near any device that generates
high frequency noise such as high frequency welding
machines, high frequency sewing machines, or SCR power
controllers.
Before cleaning, make sure to unplug the power cord to avoid
electric shock or fire.
Do not use chlorine bleach, ammonia-based cleaners,
abrasives, ammonia, or metal scouring pads. Wipe with a soft
damped cloth or a sponge soaked in water or diluted neutral
detergent.
The user shall be aware that, if the equipment is used in a
manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
The instrument should be placed so, that handy operation of
the ON-OFF Switch is possible.
Service only by a qualified Service engineer.

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Always disconnect from power supply prior to maintenance
and servicing.
Connect the instrument to a ground mains outlet, after
ensuring that the Voltage is the same as given on the name
plate
The Instrument must be connected to an earthen (grounded)
supply

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Section 2 General Description
Congratulations of your purchase of MRC hotplate which is specially
designed to provide maximum benefit for your investment with
respect to performance, safety, ease of use, and durability.
Item/Model MH-4D
Dimension (W x D x H, mm) 200 X 250 X 130
Top plate dimension (mm) 175 X 175 mm
Maximum heating capacity 800W
Current consumption (110V) 7.3A
Temp. range Up to 300°C
Temp. control Digital
Magnetic stirrer 100 – 1200 rpm
Features Performance
Microprocessor PID Feedback Control
Fast and precise temperature control is provided by the
microprocessor PID controller
Quick Heat Up Time
Ample heating capacity and superb heat transfer rate of the
tightly integrated structure of heater and top plate allow quick
heat up time (up to 300°C).

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Auto-tuning (AT)
Automatic tuning of the PID parameters provides more
accurate temperature control.
Temperature Offsetting
In case your own thermometer is to be used for temperature
control for specific applications, there can be some differences
between the temperature of your thermometer and the
displayed temperature of this unit.
Safety
High and Low Temperature Limits
The high temperature limit of the top plate is set to 300°C to
protect you and your media. However, you can set your own
high and low temperature limits to reduce operation time and
also to avoid inadvertent mistakes.
Convenience
Two large digital displays on the control panel clearly show
both set and actual temperatures of the top plate side by side.
You can easily set your own temperature limit at any value
between 0 and 300°C.
Heating rate can be set from 0 to 100% at 1% interval for
delicate temperature control.
One support rod (optional) can be attached to hold various
kinds of devices such as temperature sensors, thermometers,
and so on.

3, Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529com.mrclab@mrc
Construction
(1) Top Plat
(2) Radiation Sheets
(3) Main Power Switch
(4) FY-400 Temperature Controller
(5) Stirring Control Knob
(6) Stirring Switch
(7) Stirring motor
(8) Socket for Temperature Probe
(9) Identification Label
(10) Threaded Holes for Support Rod (Optional)
(11) Power Cord
Table of contents
Other MRC Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies 5800 ICP-OES user guide

Endress+Hauser
Endress+Hauser Cleanfit CPA875 operating instructions

NI
NI PXI-5422 CALIBRATION PROCEDURE

Collomix
Collomix Aqix operating instructions

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series operating manual

Ocean Insight
Ocean Insight FLAME-NIR+ Installation and operation manual

Parker
Parker ALIGN-MG-NA Installation, operation and maintenance manual

BD
BD 644787 user guide

DENTAURUM
DENTAURUM Compact Megaplus Instructions for use

Biuged Laboratory Instruments
Biuged Laboratory Instruments BGD 626 instruction manual

VWR
VWR SAS Super IAQ instruction manual

illumina
illumina MiSeqDx reference guide