LMS MCF-1350 User manual

OPERATIONS MANUAL
Mini Centrifuge
Model MCF-1350
Certified
DOMINIQUE DUTSCHER SAS

Contents
1. Safety 3
2. Introduction 7
3. Package Contents 7
4. Specifications 8
5. Features 9
6. Parts of the Mini Centrifuge 10
7. Control elements and Function 11
8. Installation and Operation 13
9. Maintenance 15
10. Trouble shooting 16
11. Technical data 17
12. Warranty statement 18
2
DOMINIQUE DUTSCHER SAS

Safety
Read the manual before using the Mini Centrifuge.
Warning Improper handling may cause a serious hazard,
resulting in personnel injury or death.
Caution
Improper handling may cause a serious hazard
resulting in personnel injury or damage to equipment
or buildings.
Meanings of symbols.
Caution / Warning. Refrain from doing this.
Follow all Instructions.
Do Not Touch.
Warning
Do not use a power source other than which the unit is designed to operate
on. Use of an inappropriate power supply may result in damage to the
equipment. Adhere to all electrical codes.
This device must be connected to an earthed power supply.
Insert the plug securely into a wall socket.
Loose connections result in excessive heat generation to the device and may
cause arcing to occur at the connection.
Before operation, ensure that the power cord is fully extended. Keep the cord
away from heat sources and do not place any objects on any part of the cord.
Avoid using multi-point extension power cords.
Disconnect the power supply plug if the unit is not going to be used for an
extended period of time.
3
DOMINIQUE DUTSCHER SAS

Installing the device
Warning
Place the centrifuge on a level, horizontal surface.
Ensure that the device sits completely on a worktop.
Do not install the device where it can get damp or wet.
Do not use the device in a dusty environment.
Do not use the device near a heat source or where temperatures are elevated.
Do not operate the device when temperatures fall below 4oC and rise above
45oC.
Do not use this device where there are flammable or corrosive vapors.
Install the device in a well-ventilated place. Insufficient ventilation may result
in breakdown should there be overheating.
Do not use the device outdoors.
Avoid prolonged exposure to direct sunlight.
4
DOMINIQUE DUTSCHER SAS

Operating the device
Warning
Do not operate the device before reading and fully understanding all
instructions in this manual.
Before using the device for the first time, make sure the rotor is securely
fastened to the spindle.
The rotor must be loaded symmetrically. Asymmetrical loading causes
excessive vibrations and may result in accidents.
Close all microcentrifuge tube lids before operation. Open lids can be
ripped off during operation and damage the centrifuge.
When moving the device from a cold to warm location, operate the device for
30 minutes prior to the move to avoid excessive condensation. Alternatively,
allow it to warm to room temperature for at least three hours before plugging
the socket into a power source.
Do not attempt to stop the rotor when it is moving. Always operate the
device in an upright position, and avoid moving it during operation.
This device is not designed to be user serviceable. Do not attempt to open
or modify the device. Should be device be opened by unauthorized
personnel, all warranties are void.
Should there be any abnormal smell, or noise during operation, switch off the
device immediately and contact your dealer for service.
5
DOMINIQUE DUTSCHER SAS

Caution
Do not operate with uneven loads. Uneven loads causes abnormal wear to the
bearings and damages the motor.
When the device is used to process biohazards, proper disinfection
procedures must be carried out after each use.
Use caps on your microcentrifuge tubes. Tubes without caps will cause
samples to leak during operation.
Do not use the device to process samples with flammable contents. This unit
is not designed to be explosion-proof.
Do not expose the device to Ultra Violet (UV) light. The lid may craze and
crack under such exposure.
After operation, clean the device and unplug the power if it will not be used for
extended periods.
Do not operate the device if you wear excessively loose clothing, wrist chains,
long necklaces or have uncovered long hair that can get caught in the device.
6
DOMINIQUE DUTSCHER SAS

2. Introduction
Thank you very much for purchasing the HARMONY mini centrifuge.
Your Mini Centrifuge is CE Certified and designed with function, reliability, and safety in
mind. For safe operation, please pay attention to the alert symbols throughout the
manual.
3. Package Contents
Mini Centrifuge 1 unit
Angle rotor for 1.5/2.0ml micro tubes 1 ea
Lid for Angle rotor 1 ea
Nut for Rotor Fixing 1 ea
Power cable 1 ea
Operations Manual 1 ea
Optional Items
Rotor for 8 x 0.2ml PCR tubes strip
Adapter for 0.5ml Micro tube
Adapter for 0.2ml Micro Tube
7
DOMINIQUE DUTSCHER SAS

4. Specifications
Model No. MCF-1350
Max. Speed (rpm/g) 13,500/12300
Capacity 1.5/2.0ml x 12 Tubes
0.4/0.5ml x 12 tubes
0.2ml 8 Strip PCR tubes x 4
Max. Temperature in
Chamber (after 20 min at
max. speed)
35at room temp. 24
Angle RotorAnodized AluminumAutoclavable
Lid for rotorStainless Steel Autoclavable
Material
Nut for Rotor Stainless Steel Autoclavable
Timer 130 min.
Noise Level Less than 60dB
Acceleration Time 12 sec.
Braking Time 16sec.
Dimensionmm 208(W)x(245(D)x145(H)
Weight 4.4 Kg
Power VAC100, 50/60Hz VAC220, 50/60Hz
OPTIONAL ACCESSORIES
MCF1350A05 Adapter for 0.5ml Micro tube
MCF1350A02 Adapter for 0.2mlMicro tube
MCF1350STR Rotor for 8-strip PCR tubes (8 Strip PCR tubes x 4
8
DOMINIQUE DUTSCHER SAS

5. Features
- Exceptionally small footprint
- Low noise level (less than 60dB)
- Capacity: 1.5ml/2.0ml x 12 micro tubes
- Acceleration time is less than 12 sec and
braking time is less than 16 sec. (at 13,500 rpm)
- Complies to IEC1010-2-020.
- Max. Centrifugal force is 12,300 x g
- Easy to follow digital display
- Separate short-spin button
- Max. temp. in chamber at Max. rpm: 35 (at room temp. 24)
- Automatic lid release
- Maintenance-free drive
- Autoclavable rotor and rotor lid (121, 20 min.)
- Timer can be set up to 30 min.
- rpm/rcf setting as required
- Variable accessories: rotor for PCR tubes
(0.2ml 8-Strip PCR tubes x 4)
Adapter for 0.5 ml and 0.2 ml PCR tubes
9
DOMINIQUE DUTSCHER SAS

6. Parts of the Mini Centrifuge
(Front View)
(Rear View)
10
DOMINIQUE DUTSCHER SAS
Table of contents
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies 5800 ICP-OES user guide

Endress+Hauser
Endress+Hauser Cleanfit CPA875 operating instructions

NI
NI PXI-5422 CALIBRATION PROCEDURE

Collomix
Collomix Aqix operating instructions

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series operating manual

Ocean Insight
Ocean Insight FLAME-NIR+ Installation and operation manual

Parker
Parker ALIGN-MG-NA Installation, operation and maintenance manual

BD
BD 644787 user guide

DENTAURUM
DENTAURUM Compact Megaplus Instructions for use

Biuged Laboratory Instruments
Biuged Laboratory Instruments BGD 626 instruction manual

VWR
VWR SAS Super IAQ instruction manual

illumina
illumina MiSeqDx reference guide