3
1. First, check to make sure that the voltage selector voltage indicator is
matched to the AC power supply voltage.
Reset the voltage selector as described below if it is not set to the
correct voltage.
(1) Use a regular screwdriver to rotate the voltage selector so that the
groove in the center points to the correct AC power supply voltage.
(2) Check to make sure that there is a fuse installed in the fuse holder.
2. Connect the Power Supply Cord and Sensor Cable to the main unit. Connect a ground wire to the Ground Terminal and
be sure to ground properly. Depending on environmental conditions, it may not be possible to obtain accurate measure-
ment results if the unit is not grounded properly.
Note:
Make the RS-232C, buzzer and AC Output Connector connections as necessary. Please refer to section 4., “The Serial
Interface” regarding the RS-232C output format.
Note:
When the Upper Limit Setting described on page 5 is performed, an AC current is output from the AC Output Connector
in accordance with the settings made. Be very careful to avoid electric shock.
Voltage Indicator Window
Voltage Selector
2. Operating the HG-770