Kambic OBM User manual

User Manual
Calibration Bath
OBM

User Manual
OBM
2
WARRANTY
Kambic d.o.o. guaranties that the product
delivered has been thoroughly tested to ensure
that it meets its published specifications. The
warranty is valid only if the product has been
installed and used in accordance to the
instructions supplied by Kambic d.o.o.
Kambic d.o.o. shall in no event be liable for
incidental or consequential damages, including
without limitations, lost profits, loss of income, loss
of business opportunities, loss of use, and other
related exposures, caused by incorrect use of
product.
Warranty is valid for 13 months from date of
shipping and includes spare parts and labor.
Type: OBM
Serial number:
Date of shipping:
Signature/stamp:
Manufacturer:
Kambic d.o.o.
Metliska cesta 16
8333 Semic
Slovenia
Tel: +386 7 35 65 220
Fax: +386 7 35 65 232
Email: info@kambicmetrology.com
Web: www.kambicmetrology.com
Thank you for your trust and for buying this
device. We hope it will successfully serve the
purpose for many years.
CAREFULLY READ THESE INSTRUCTIONS
BEFORE USING THE DEVICE!

User Manual
OBM
3
CONTENT
1. WARNINGS AND SAFETY PRECAUTIONS........... 4
2. INTRODUCTIONS ................................................ 6
3. TECHNICAL DATA ................................................ 7
3.1 Front view............................................................. 8
3.2 Rear view ............................................................. 8
3.3 Top view............................................................... 9
4. INSTRUCTIONS FOR USE.................................. 10
4.1 Basic Setup ......................................................... 10
4.2 Control Panel with OLED display and keyboard...... 10
4.2.1 Main Screen (running) ...................................... 11
4.2.2 Main Screen (not running)................................. 11
4.2.3 Program selection and start............................... 12
4.2.4 Menu and navigation ........................................ 15
4.3 Program Setup ................................................... 16
4.4 Instrument Setup ............................................... 18
4.4.1 Display Parameters ........................................... 18
4.4.2 Time and Date ................................................. 19
4.4.3 System Parameters........................................... 19
4.4.4 Bath Calibration ............................................... 20
4.4.5 PID parameters ............................................... 20
4.4.6 Change Password ............................................. 21
4.5 Service .............................................................. 22
5. SERVICE AND MAINTENANCE .......................... 23
5.1 Draining the Bath ................................................ 24
6. REQUIRES SPACE AND CONECTIONS .............. 25
7. SAFE USE AND SAFETY ELEMENTS ................... 26
8. DISMANTLING AND DISPOSAL ........................ 27
9. ABBRIVATIONS ................................................. 28
10. TROUBLESHOOTING ....................................... 29
11. ATTACHMENTS ................................................ 30
11.1 Wiring diagram
11.2 Declarations
11.3 RS232 Communication protocol

User Manual
OBM
4
1. WARNINGS
AND SAFETY
PRECAUTIONS
Always keep this manual in close reach for reference. This product should only be used
and operated by skilled and trained personnel familiar with technical capabilities and
potential safety hazards.
Always follow and obey the national regulations related to technology and labor laws.
To avoid injuries and equipment damage study and follow the safety instructions and
warnings.
Legal:
The content of the manual is subject to change without notice.
Understanding and following this operating manual is first step to hazard-free and safe
use of the equipment.
This user manual can not reach out to all application that might even develop in the
future. If you find information unclear or you are not sure about safe use of the product in
further information or opinions.
In no event shall Kambic d.o.o. be held liable for any damages, direct or incidental related
to the use of this product and manual.
Product can be equipped with variety of accessories and options. This warning list might
include elements that are applicable only with certain models and accessories.
READ ALL INSTRUCTIONS BEFORE USE!
Reduce the possibility of injury and failure.
WARNING
Before you perform
any kind of maintenance, disconnect all electrical power
to the unit! DO NOT operate this unite without a properly grounded,
properly polarized power cord. DO NOT connect this unite to a non-
grounded, non-polarized outlet.
DANGER
Appliance must not be installed in a room where the possibility of explosive
atmosphere is present!
DANGER OF BREAKDOWN
Service only allowed to authorized personnel!

User Manual
OBM
5
WARNING
Never use with toxic, corrosive, flammable or organic materials unless special
precautions are in place to prevent injury to personnel or damage to equipment.
WARNING
Never operate the appliance with the housing removed!
RISK OF FAILURE
Due to exceptional volumes of fluid expansion at elevated temperatures,
regularly check the fluid level in the bath!
DANGER
Temperature fluid expansion may cause overspill!
Adjust fluid level!
RISK OF FAILURE
Do not use rubber cap at elevated temperatures
DANGER OF BURNS
Use safety gloves when operating with high/low temperatures!
ELECTRIC HAZARD
Do not operate the equipment in wet and condensing environment!
STABILITY HAZARD
Do not load or affect product stability with heavy loads or climbing the product!
RISK OF FAILURE
Ensure correct flow. Insuficient flow will resault poor stability and uniformity.
Check stir magnet-bar position and rotation to ensure stirring / flow.
✓

User Manual
OBM
6
2. INTRODUCTIONS
Calibration bath OBM was developed and manufactured specifically for the calibration of
temperature instruments, probes and liquid thermometers at temperatures from 35°C to
220°C. Fluid calibration baths ensure high temperature stability and uniformity across
working volume.
To ensure such specifications all calibration baths of our brand are designed with a special
double cylinder construction with uniform vertical flow in the working volume.
Of course, the construction is not enough that is why we are using high capacity
microprocessor controllers and different levels of heating.
Control panel with display is used to set the individual temperatures. All controllers are
also equipped with communication modules, so the appliance can be operated and
controlled via a PC.
Bath top, bath bottom and working chamber is completely made of stainless steel AISI 304,
which guarantees easy maintenance and long life time. Exterior housing is made of
aluminum sheets coated with high quality varnish.
WARNING
Temperature range must be limited / adjusted to reflect the properties of
the selected fluid!

User Manual
OBM
7
3. TECHNICAL DATA
OBM
Width
225 mm
Height
638mm
Depth
330 mm
Working chamber dimensions
50 x 210 mm
Volume
~ 1 L
Temperature range
+ 35 °C …+ 220 °C
Temperature Stability
from ± 0.005 °C to ± 0.05 °C
Temperature homogeneity
Better than ± 0.05 °C
Display resolution
0.001 °C
Set-up resolution
0,01 °C
Control
MPC –PID
Power supply (±10%)
230 V / 50 Hz
Wattage
300 W
Communication port RS 232
Built-in
Temperature probe
PT 100
Construction materials
SS vessel, exterior housing & cover
Width
225 mm

User Manual
OBM
8
3.1 Front view
3.2 Rear view
Control panel
Bath working chamber
Rubber cap
Refrigeration system
compartment –air intake
Perforated back
Main switch
Power cord connection
RS 232 PC port
USB port
Rubber legs
Front housing
Back housing
Carrying handle
Tank drain plug

User Manual
OBM
9
3.2 Rear view
Bath working chamber
Removable inner tank wall
Openings on inner tank wall

User Manual
OBM
10
4. INSTRUCTIONS
FOR USE
4.1 Basic setup
Step 1: Remove rubber cap and fill the bath with appropriate fluid (up to 80%)
Step 2: Connect the power cord into the power supply 230V±10%; 50Hz
Step 3: Put the main switch in position 1 (on the back of the device)
Step 4: Press button and in Program Start Menu select the desired program. After selection press
button to start cycle
Step 5: Check the fluid level and add if necessary
Step 6: Press button to stop the cycle
4.2 Control panel with OLED display and keyboard
START button
ESC button
STOP button
OLED display
Increase value button
Decrease value button
Navigate right button
Navigate left button
Navigate down button
Navigate up button
ENTER button
MENU button
Table of contents
Other Kambic Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies 5800 ICP-OES user guide

Endress+Hauser
Endress+Hauser Cleanfit CPA875 operating instructions

NI
NI PXI-5422 CALIBRATION PROCEDURE

Collomix
Collomix Aqix operating instructions

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series operating manual

Ocean Insight
Ocean Insight FLAME-NIR+ Installation and operation manual

Parker
Parker ALIGN-MG-NA Installation, operation and maintenance manual

BD
BD 644787 user guide

DENTAURUM
DENTAURUM Compact Megaplus Instructions for use

Biuged Laboratory Instruments
Biuged Laboratory Instruments BGD 626 instruction manual

VWR
VWR SAS Super IAQ instruction manual

illumina
illumina MiSeqDx reference guide