HiWeigh K8 User manual

K8
(S)
K9
WEIGHT INDICATOR


1
Before Use
1.1 Safety precautions
WARNING!
▲Do not use K8 K9 series weighing terminal in hazardous area! Do not
use it within areas classified as hazardous division 1/2 or zone
0/1/2/21/22 because of combustible or explosive atmospheres.
▲ Never immerse it in corrosive chemical liquid.
▲ Static sensitive device, it must be handled only by qualified technicians.
Improper handling may damage the circuit card and the device, which is
not covered by the warranty.
DANGER!
Electric shock hazard!
▲ Make sure the indicator is grounded well.
▲ Always unplug AC cable before performing any service work on the indicator! And wait for
at least 30 seconds before any operation on the indicator.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE), this device may not be disposed of in domestic waste. This also applies to
countries outside the EU as per their specific regulations.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from
which you purchased this indicator.
Should this indicator be passed on to other parties (for private or professional use), the
content of this regulation must also be related.
The indicator has a rechargeable internal battery. The battery contains heavy metals. Please
observe the local regulations on the disposal of environmentally hazardous materials.
All Rights Reserved, any copying, reproducing, republishing, posting, distributing by any means is prohibited without permission
of Moorange Electronics (Shanghai) MFG Co.,Ltd.

2
Introduction
Thank you for selecting and purchasing our K9 weight indicator, this indicator provides a compact and flexible solution for a
variety of weighing needs, especially for the wet working places.
INVENTIONS AND DESIGN PATENTED
ON17-P31351 & ON17-P20743
Double O-ring for the housing, independent sealed PCB case, transparent PC front housing and touch buttons, all give the
highest protection class till IP68 and IP69K, it can be used perfectly in the field of meat, fish process or any other food industry,
chemical materials, pharmaceutics. It can be washed by the high pressure hot water - 100 bar (1450 psi) and 80ºC, suitable for
almost all harsh working environment.
PC alloy front housing and PBT back housing, give the indicator the high strength, no worry any more about the damage to it
caused by falling, crudely use of some careless workers.
It gives three options of checkweigh function, you can select the three small LED lights for checkweigh alarming, also you can set
the different backlight for different range of the weight read, or you can connect with an extra alarming light with three colors.
Various power choices:
100-240V AC Direct
9-24V DC Direct
7.4V6800mA Li-ion Battery
6V1200mA Lead-acid Battery
Various output choices:
Double RS232
4 x OUTPUT
1 x IN*
Optional Bluetooth or WiFi
* for pedal or mechanical switch, functions can be set with zero, tare or print.

3
Index
Technical Specifications
4
Model Identification
4
Packing List
5
Connecting
5
Power board
5
I&O board
5
Load cell
6
RS232
6
Input & Output
6
LCD Display
7
Keypad
8
Parameter setting and calibration
9
Operating method
9
F1 – Scale parameter setting
10
F2 – User functions setting
11
F3 – Display parameters
13
F4 – COM parameters setting
14
F5 – Maintenance
16
Indications & remarks
19
Main functions
19
Communication protocols
21
SM continuous output format
21
MT continuous output format
21
AN continuous output format
23
Demand output
23
Printing format
23

4
1. Technical Specifications
Model
K8 | K8S | K9
Enclosure Type
PBT | SUS304 | PBT+PC
Product Dimension
240x160x110mm (K8, K9) 250x170x65mm (K8S)
Shipping Weight
2.8kg (K8, K9) | 3.3Kg (K8S)
Accuracy
Class III
Display Resolution
1/3,000 – 1/30,000
Power
100-240V AC
9-24V DC
7.4V6800mA Li-ion Battery
6V1200mA Lead-acid battery
Display
FSTN LCD with mutli-color backlight
Load cell No.
6*350ohm or 12*700ohm
Excitation voltage
5 VDC
Units
Kg | lb, g | oz
Operating Temperature
-10℃~40℃
Storage Temperature
-25℃~55℃
Relative humidity
85%Rh non-condensing
Communication
RS232*2, Optional Bluetooth
2. Model Identification
Model: K9 -AC UK 1 0
Corresponding: A B C D E
A = Main model name
B = Power source: -AC: AC power cord + AC/DC power PCB.
-DC: DC power direct.
-Li: Li-ion 7.4V6800mA battery
-La: Lead-acid 6V1200mA battery
C = Plug type, examples:
D = Output: 0 = RS232*2
1 = Bluetooth
E = pending, no function
AU = Australia Type
CN = China Type
EU = EU Type
US = USA Type
SA = South Africa Type
UK = UK Type

5
3. Packing List
After the weighing terminal received, please open the box carefully and check the following items included:
-Indicator x 1
-S.S bracket with screws x 1
-Connectors and screws bag x 1
-Manual x 1
4. Connecting
4.1 POWER BOARD
K8Ac, K8SAc and K9Ac with only AC/DC power PCB.
K8Dc, K8SDc and K9Dc without AC/DC and DC/DC PCB, DC power connect to I/O board.
K8Li, K8SLi and K9Li with both AC/DC and DC/DC PCB, battery connect to DC/DC board.
K8La, K8SLa and K9La with only AC/DC power PCB, battery connect to I/O board.
4.2 I&O BOARD

6
4.3 LOAD CELL
For 6-wires, just connect as the indication on the I/O board:
K8, K9 Load Cell
+
+
For 4-wires, short connect: +EXC and +SEN, –SEN and -EXC.
K8, K9 Load Cell
4.4 RS232
The terminal has two independent RS232 output, TXD1 and RXD1 for the 1st output and TXD2 and RXD2 for
the 2nd output, GND share.
K8, K9 Computer/Printer, etc.
TXD………………. RXD
RXD………………. TXD
GND……………… GND
4.5 INPUT & OUTPUT
4.5.1 Diagram and wirings:
+EXC………………Excitation +
+SEN………………Sense +
+SIG……………… Signal +
SHIELD……….….Shield
-SIG ……………… Signal –
-SEN……………… Sense –
-EXC……………… Excitation –
+EXC………………Excitation +
+SEN……………… Excitation +
+SIG……………… Signal +
SHIELD……….….Shield
-SIG ……………… Signal –
-SEN……………… Excitation –
-EXC……………… Excitation –

7
4.5.2 INPUT
The terminal has one Input interface, which allows the user to connect with foot pedal or other mechanical
switch to control the functions of ZERO, TARE, PRINT, F1 or F2 (refer to F2.3).
Example of connecting to pedal switch
4.5.3 OUTPUT
The terminal has 3 output interface, which allows the user to connect with alarming light or other devices.
Example of connecting to 3-lights tower (charter as above)
5. LCD Display
FSTN LCD with multi-color backlight, clearly read even in the bright sunlight.
Battery
Unit
Mode
Function Weight Read
The backlight color can be set by the user, and also it allows to set different colors for checkweigh function
(refer to F2.8 & F2.9)

8
6. Keypad
K8 and K8S with membrane keypad, K9 with touch keypad which allows the user to operate it in faster way,
even with hand gloves, there are 6 function keys:
(keypad example of K9 model)
On/Off key
Zero key
ESC or return during parameter Setting
Tare key
Enter function during parameter Setting
F1 – User function 1
Move the digits to right during parameter Setting
F2 – User function 2
Increase the digits during parameter Setting
Printing key
This manual suits for next models