Hach orbisphere C1100 User manual

DOC024.98.93019
ORBISPHERE Model
C1100 Ozone Sensor
10/2016, Edition 5
Basic User Manual
Basishandbuch
Manuel d'utilisation de base
Manual básico del usuario
Manuale di base per l'utente
Manual Básico do Usuário
Alapvető felhasználói kézikönyv
Temel Kullanıcı Kılavuzu
Manual de utilizare de bază

English..............................................................................................................................3
Deutsch.......................................................................................................................... 12
Français......................................................................................................................... 22
Español.......................................................................................................................... 32
Italiano............................................................................................................................ 42
Português...................................................................................................................... 52
Magyar........................................................................................................................... 62
Türkçe.............................................................................................................................72
Română......................................................................................................................... 81
2

Sensor membrane specifications
Specifications are subject to change without notice.
Specification Membrane 2956A Membrane 29552A
Recommended applications Trace measurement High concentration (> 1 mg/l)
Material PFA PTFE
Thickness [μm] 25 50
Calibration gas Span gas or air
Dissolved measurement range 0 ppb to 50 ppm 0 ppb to 200 ppm
Accuracy
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
or ± 0.4 ppb,
or ±1 Pa
The greater of ±1% of reading
(± 5% for sensors calibrated in air)
or ± 20 ppb,
or ± 4 Pa
Expected current in air @ 1 bar 25°C
[μA] 25.3 6.5
Temperature compensation range – 5 to 45° C
Temperature measuring range – 5 to 100° C
Response time125 sec. 6 min.
Recommended minimum liquid flow
rate2 [mL/min] 35031003
Recommended minimum linear flow
rate2 [cm/sec] 30 10
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
1Response time at 25°C for a 90% signal change
2Liquid flow through an ORBISPHERE 32001 flow chamber, with protection cap and no grille
3These flow rates take into account the decomposition of ozone in the tubing between the line
and the flow chamber (theoretical flow rates in the absence of decomposition would be 10 times
less)
English 3

Use of hazard information
D A N G E R
Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death or serious
injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or moderate injury.
N O T I C E
Indicates a situation which, if not avoided, may cause damage to the instrument. Information that requires special
emphasis.
Precautionary labels
Read all labels and tags attached to the product. Personal injury or damage to the product could
occur if not observed.
This symbol, when noted on a product, indicates a potential hazard which could cause serious
personal injury and/or death. The user should reference this instruction manual for operation and/or
safety information.
This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock
and/or electrocution exists and indicates that only individuals qualified to work with hazardous
voltages should open the enclosure or remove the barrier.
This symbol, when noted on the product, indicates that the marked item can be hot and should not
be touched without care.
This symbol, when noted on the product, indicates the presence of devices sensitive to electrostatic
discharge and indicates that care must be taken to prevent damage to them.
This symbol, when noted on the product, identifies the location of the connection for protective earth
(ground).
Electrical equipment marked with this symbol may not be disposed of in European public disposal
systems. In conformity with European local and national regulations, European electrical equipment
users must now return old or end-of-life equipment to the manufacturer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return
end-of-life equipment, producer-supplied electrical accessories, and all auxillary items for proper disposal.
Products marked with this symbol indicates that the product contains toxic or hazardous substances
or elements. The number inside the symbol indicates the environmental protection use period in
years.
What you have received
C1100 electrochemical sensor
The sensor may be delivered separately or as part of an ORBISPHERE system, depending on the
individual order.
4 English

The sensor will be delivered fitted with a plastic screw-on storage cap to protect the sensor head.
This is held in place with a plastic collar for the C1100-S00 sensor, or a stainless steel collar for the
C1100-T00 sensor.
A plastic screw-on base is also provided to protect the connection socket, and which also provides a
suitable stand for the sensor during maintenance procedures, and when not in use.
Protection caps
One protection cap with grille will be delivered as standard with each sensor.
Sensor recharge kit
A recharge kit should have been ordered with the sensor as this will be required to initially make the
sensor operational. It is also required for sensor cleaning and membrane replacement procedures.
Note: The recharge kit for ozone has a green sticker on the front of the box.
Sensor recharge kit
The kit contains:
• four recharge cartridges with pre-mounted membrane and electrolyte. The type of membrane
mounted in the cartridge will be specific to the kit ordered
• anode cleaning tools
• a set of replacement O-rings
• a set of replacement Dacron® mesh patches
The black anode cleaning tool is used to clean the anode of any deposits or residue that may have
formed. It is doubled-ended so it can be used for two membrane replacement processes, each end
being used once.
The Dacron® mesh patches provide protection to the membrane.
Installation
Sensor preparation
Your sensor has been thoroughly cleaned and tested at the factory before shipment. It has been
shipped with a cartridge containing a membrane and electrolyte pre-installed to protect the sensor
head. This cartridge must be removed and replaced with a new one prior to first use to make it fully
operational. The new cartridge is included in the sensor recharge kit. You will also need one of the
mesh patches included with the kit.
The following instructions detail the steps required to make the sensor operational. Should you have
any questions, your Hach Lange representative will be pleased to help.
English 5

CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current material safety data
sheets (MSDS) for safety protocols.
Note: It is advisable to perform this procedure with the plastic sensor base installed so as to avoid any damage to
the connection socket and also to provide a suitable stand for the sensor when required.
1. 2.
3.
4.
1. Hold the main body of the sensor and unscrew the protection cap locking washer by turning
counter-clockwise. Remove it from the sensor and put to one side.
2. Pull/twist off the protection cap and put to one side. Remove the Dacron® mesh from inside the
cap and discard it.
3. Hold the sensor with the membrane facing down to avoid spilling any electrolyte, then carefully
unscrew the shipment cartridge. Drain the old electrolyte into a sink and flush away. Discard the
shipment cartridge and membrane. If present, remove the cotton washer from the top of the
anode and discard.
4. Rinse the sensor head under a tap for 15 seconds, aiming the jet of water directly onto the sensor
head. Do not dry the center electrode area, as the gap between cathode and guard should be left
filled with water.
5. 6. 7.
8.
6 English

5. Place the recharge cartridge container on a flat work surface and, keeping the container upright
to avoid spilling any of the electrolyte inside, carefully unscrew the top. Remove the packing
component from the center of the cartridge, making sure that the O-ring on top of the cartridge
remains in place. If it comes away then replace it before continuing. If there are any visible
bubbles in the electrolyte, remove them using a stirring motion with the packing component.
6. Hold the container steady between thumb and forefinger of one hand. Lower the sensor into the
container until the top of the anode is covered with electrolyte.
7. Gently screw the sensor clockwise into the replacement cartridge, applying minimum pressure to
avoid any damage to the screw threads.
8. Continue turning until the cartridge is attached to the sensor, and the sensor is automatically
released from the container. The empty container, the screw top and packing component can be
discarded.
Note: It is normal that some of the electrolyte will overflow from the replacement cartridge and into the plastic
container.
9.
10.
11.
9. Rinse the sensor under a tap for about 5 seconds to remove any excess electrolyte, then gently
wipe with a soft tissue to ensure all parts are completely dry. Drain the overflow electrolyte from
the container into a sink and flush away. Discard the used container.
10. Take a new Dacron® mesh patch from the box of O-rings in the recharge kit. Place the mesh in
the center of the protection cap. It is very important that the mesh is in the center of the protection
cap and covering the entire grille. Lower the sensor onto the protection cap making sure not to
disturb the mesh.
11. Push the protection cap firmly into place, making sure one of the four slots in the protection cap
fits over the small locking pin (highlighted right). If it is necessary to turn the protection cap to fit
over the locking pin, ensure you only turn it clockwise to avoid unscrewing the cartridge.
12. Finally, screw the protection cap locking washer back into place in a clockwise motion, and
tighten finger tight.
Sensor installation
Sensor positioning
The sensor must be installed in a socket or flow chamber that allows contact with the sample fluid to
be analyzed. The sensor and measuring instrument are connected by a cable and two 10-pin
connectors. The standard sensor cable length is 3 meters though extension cables of up to
1000 meters are available. However, smart sensor technology is only available with distances of up
to a maximum of 750 meters. Ensure that the sensor will be mounted:
• perpendicular to the pipe
• on a horizontal pipe section (or on flow-ascending vertical pipe)
• minimum of 15 meters away from the pump's discharge side
• in a place where the sample flow is stable and rapid, and as far as possible from:
English 7

• valves
• pipe bends
• the suction side of any pumps
• a CO 2 injection system or similar
Note: There may be situations where not all the above conditions can be met. If this is the case, or you have any
concerns, please consult your Hach Lange representative to appraise the situation and define the best applicable
solution.
Sensor insertion
• Insert the sensor straight into the flow chamber or socket. Do not twist the sensor.
• Hand tighten the attaching collar.
• Connect the sensor cable.
• Check for leaks; replace O-rings if product leaks are visible.
Sensor removal
• If not using the ORBISPHERE 32003 insertion/extraction valve you will need to shut off the sample
flow and drain the sampling circuit of liquid.
• Remove the sensor cable connected at the sensor end.
• Hold the sensor body in one hand to avoid rotation, and unscrew the collar with the other hand.
• Pull the sensor straight out of the socket or flow chamber.
• Install the sensor storage cap and sensor base (to protect the connection).
Maintenance
It is recommended to perform standard maintenance on the sensor about once every six months,
though this will vary depending on the application. This involves sensor membrane replacement and
head cleaning as described in Membrane replacement and sensor head cleaning on page 8.
Using the ORBISPHERE 32301 cleaning and regeneration center in addition to the standard
maintenance will allow for a noticeably extended sensor life.
Membrane replacement and sensor head cleaning
A sensor recharge kit (see Sensor recharge kit on page 5) is required as it contains all the
components necessary for this membrane replacement and sensor head cleaning process (i.e. a
cartridge containing the electrolyte and pre-installed membrane, a sensor cleaning tool, replacement
O-rings and Dacron® mesh patches).
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current material safety data
sheets (MSDS) for safety protocols.
Note: It is advisable to perform this procedure with the plastic sensor base installed so as to avoid any damage to
the connection socket and also to provide a suitable stand for the sensor when required.
8 English

1. 2.
3. 4.
1. Hold the main body of the sensor and unscrew the protection cap locking washer by turning
counter-clockwise. Remove it from the sensor and put to one side.
2. Pull/twist off the protection cap and put to one side. Remove the Dacron® mesh from inside the
cap and discard it.
3. Hold the sensor with the membrane facing down to avoid spilling any electrolyte, then carefully
unscrew the old cartridge. Drain the old electrolyte into a sink and flush away. Discard the old
cartridge and membrane. If present, remove the cotton washer from the top of the anode and
discard.
4. Rinse the sensor head under a tap for 15 seconds to remove any remaining electrolyte and shake
dry. With a soft tissue gently clean around the guard area (indicated above) and then wipe off any
excess moisture from the sensor to ensure all parts are completely dry. Repeat this rinse and dry
process with the protection cap.
5.
6. 7.
8.
5. With the help of a pair of tweezers, remove the old O-ring from the sensor body. Replace the O-
ring with a new one from the recharge kit.
6. Clean the anode using the cleaning tool supplied. Place the tool over the sensor head. Clean by
rotating the cleaning tool over the sensor head for a few seconds, in a clockwise direction only.
7. Remove the tool and tap it face down on a flat work surface to remove any powdery deposit.
Check the sensor to ensure that all deposits have been removed from the anode. If not, repeat
step 6. until the anode regains its bright silver appearance.
8. Rinse the sensor head under a tap for 15 seconds, aiming the jet of water directly onto the sensor
head. Do not dry the center electrode area, as the gap between cathode and guard should be left
filled with water.
English 9

N O T I C E
At this stage it is highly recommended to use the ORBISPHERE 323021 electrochemical cleaning and
regeneration center for improved sensor performance and a noticeably extended sensor life. Detailed
information on how to use the cleaning and regeneration center is included in the 32301 Operator Manual.
The ORBISPHERE 32301 is a very efficient cleaning and regeneration tool for electrochemical sensors. This
tool reverses the electrochemical process that is taking place in the sensor cell during normal operation. This
removes oxidation and at the same time regenerates the surface of the electrodes. In addition, the
regeneration center offers a continuity tester for checking the sensor electronics.
9. 10. 11.
12.
9. Place the recharge cartridge container on a flat work surface and, keeping the container upright
to avoid spilling any of the electrolyte inside, carefully unscrew the top. Remove the packing
component from the center of the cartridge, making sure that the O-ring on top of the cartridge
remains in place. If it comes away then replace it before continuing. If there are any visible
bubbles in the electrolyte, remove them using a stirring motion with the packing component.
10. Hold the container steady between thumb and forefinger of one hand. Lower the sensor into the
container until the top of the anode is covered with electrolyte.
11. Gently screw the sensor clockwise into the replacement cartridge, applying minimum pressure to
avoid any damage to the screw threads.
12. Continue turning until the cartridge is attached to the sensor, and the sensor is automatically
released from the container. The empty container, the screw top and packing component can be
discarded.
Note: It is normal that some of the electrolyte will overflow from the replacement cartridge and into the plastic
container.
13.
14.
15.
10 English
Other manuals for orbisphere C1100
3
Table of contents
Languages:
Other Hach Accessories manuals

Hach
Hach AN-ISE sc User manual

Hach
Hach LDO User manual

Hach
Hach Chlorine Sensor User manual

Hach
Hach Flo-Dar User manual

Hach
Hach Polymetron Series User manual

Hach
Hach SOLITAX sc User manual

Hach
Hach 5740sc User manual

Hach
Hach Flo-Dar Quick manual

Hach
Hach POLYMETRON 8398 User manual

Hach
Hach LANGE ULTRATURB sc User manual

Hach
Hach LDO AQS sensors User manual

Hach
Hach ORBISPHERE 31x9 Series Quick manual

Hach
Hach 6120118 User manual

Hach
Hach D3433B8.99 User manual

Hach
Hach 3798-S sc User manual

Hach
Hach ORBISPHERE Series Quick manual

Hach
Hach ORBISPHERE K1200 User manual

Hach
Hach 3700 Series Quick manual

Hach
Hach ULTRATURB sc User manual

Hach
Hach 8362sc User manual