English 73 Series III Instruction Sheet.
Panel
10
General Specifications
Maximum Voltage Between
any Terminal and Earth
Ground
600 V
Display Digital: 3,200 counts, updates 2.5/sec
Analog: 31 segments, updates 25/sec
Response Time of Digital
Display V ac < 2 s
V dc < 1 s
Ω< 1 s to 320 kΩ, < 2s to 3.2 MΩ,
< 10 s to 32 MΩ
Operating Temperature 0°C to 50°C
Storage Temperature -40°C to 60°C
Temperature Coefficient 0.1 x (specified accuracy) / °C
(<18°C or >28°C)
Electromagnetic Compatibility
in RF field of 3 V/m on all
functions except MΩ
Total accuracy = Specified accuracy
plus +2.0 % of range for all functions
except:
320.0 mV range: total accuracy =
specified accuracy + 1.0 % of range
320.0Ωrange: total accuracy =
specified accuracy + 8 % of range
Relative Humidity
except 32 MΩRange:
32 MΩRange only:
0 % to 90 % (0 °C to 35 °C)
0 % to 70 % (35 °C to 50 °C)
0 % to 80 % (0 °C to 35 °C)
0 % to 70 % (35 °C to 50 °C)
Altitude Operating: 2000 meters
Storage: 12,000 meters
Battery Type 9 V, NEDA 1604 or 6F22 or 006P,
or NEDA 1604A or 6LR61
Battery Life 2000 hrs typical with alkaline
1600 hrs typical with carbon zinc
Continuity Beeper 4096 Hz
Shock, Vibration per MIL-T-PRF 28800F Class III,
Sinusoidal, Non Operating
Size (H x W x L) 2.8 cm x 7.5 cm x 16.6 cm
(1.12 in x 2.95 in x 6.55 in)
Weight 340 g (12.0 oz)
Safety 600 V CAT II per ANSI/ISA S82.01-
1994, EN 61010-1: 1993, CSA C22.2
No 1010.1-92, UL 3111-1.
EMC Regulations EN 61326-1: 1997
Certifications/Listings