U1500s
High Voltage Differential Probe
Instructions
Introduction
The Fluke U1500s High Voltage Differential Probe (the Probe or
Product) is a portable voltage sensor with a 1500 V dc/1000 V ac range
and 0.1 % accuracy.
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
• Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-3-6714-3114
• Singapore: +65-6799-5566
• Anywhere in the world: +1-425-446-5500
• Customer Service Center (Fluke China): + 400-810-3435
• Maintenance Service Center (Fluke China): +400-921-0835
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
Go to www.fluke.com to register your product and find more information.
Safety
A Warning identifies conditions and procedures that are dangerous to
the user.
XW Warning
To prevent personal injury and for safe operation of the
Product:
• Read all safety information before you use the Product.
• Use the Product only as specified, or the protection
supplied by the Product can be compromised.
• Comply with local and national safety codes. Use
personal protective equipment (approved rubber gloves,
face protection, and flame-resistant clothes) to prevent
shock and arc blast injury where hazardous live
conductors are exposed.

• Examine the case before you use the Product. Look for
cracks or missing plastic. Carefully look at the insulation
around the terminals.
• Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
• Use Product-approved measurement category (CAT),
voltage, and amperage rated accessories (probes, test
leads, and adapters) for all measurements.
• Do not use test leads if they are damaged. Examine the
test leads for damaged insulation and measure a known
voltage.
• Do not use the Product if it is damaged.
• Do not use the Product if it operates incorrectly.
• The battery door must be closed and locked before you
operate the Product.
• Remove all probes, test leads, and accessories before the
battery door is opened.
• Do not work alone.
• Do not exceed the Measurement Category (CAT) rating of
the lowest rated individual component of a Product,
probe, or accessory.
• Do not touch voltages >30 V ac rms, 42 V ac peak, or
60 V dc.
• Do not apply more than the rated voltage, between the
terminals or between each terminal and earth ground.
• Measure a known voltage first to make sure that the
Product operates correctly.
• Disable the Product if it is damaged.
• Use only specified replacement parts.
• Have an approved technician repair the Product.
Table 1 shows the symbols that can be used on the Product or in this
manual.
Table 1. Symbols
Symbol Definition
WWARNING - RISK OF DANGER.
XWARNING. HAZARDOUS VOLTAGE. Risk of electric shock.
Consult user documentation.
Earth
TDouble Insulated
Alkaline Battery
PConforms to European Union directives.
)Certified by CSA Group to North American safety standards.

Instructions
Table 2 shows the Product. To use the Product:
1. Connect the input and output test leads to the Product (see and
in Table 2).
2. Connect the output test lead (BNC to Banana cable) to the power
analyzer, oscilloscope, or other instruments.
3. Set proper scale for the power analyzer or other instruments, and
adjust the sensitivity based on the tested voltage.
4. Turn on the Product .
The power indicator light (see Table 2) flashes green. If the
battery light flashes red, change the batteries.
5. Connect the input test lead to the DUT (Device Under Test) and
start the test.
Keep the Product away from high-voltage pulse circuits to reduce
interference to the Product while testing.
6. Disconnect the test leads from the DUT, and turn off the Product
after test completion, then unplug the test leads.
Conforms to relevant Australian Safety and EMC standards.
Conforms to relevant South Korean EMC Standards.
Measurement Category III is applicable to test and measuring
circuits connected to the distribution part of the building’s low-
voltage MAINS installation.
Measurement Category IV is applicable to test and measuring
circuits connected at the source of the building’s low-voltage
MAINS installation.
~
This product complies with the WEEE Directive marking
requirements. The affixed label indicates that you must not
discard this electrical/electronic product in domestic household
waste. Product Category: With reference to the equipment types
in the WEEE Directive Annex I, this product is classed as
category 9 "Monitoring and Control Instrumentation" product. Do
not dispose of this product as unsorted municipal waste.
Table 1. Symbols (continued)
Symbol Definition

Battery
XW Warning
To prevent possible electrical shock, fire, or personal injury:
• Disconnect the Product from any measurement device
before you install or replace the batteries.
• Replace the batteries when the low battery indicator
shows to prevent incorrect measurements.
To install the batteries, see Figure 1:
1. Disconnect all Probe connections.
2. Make sure that the power switch is OFF.
3. Use a flat-blade screwdriver to rotate the battery door fastener to
the unlocked position.
4. Remove the battery door.
5. Install the new batteries with the correct polarity.
6. Lock the battery door.
Table 2. The Product
Item Description
Battery indicator (red flash = low battery, green flash =
battery ok)
Power switch
Output terminal
Battery door fastener
Input terminals
Output Test lead
1
2
3
4
U1500S
HIGH VOLTAGEDIFFERENTIALPROBE
6
5

Figure 1. Battery Installation
Maintenance
If the Product fails inspection, tape it shut to prevent unintended
operation.
XWWarning
To prevent electric shock, fire or personal injury:
• Before each use, examine the Product. Do not use the
Product if it is damaged.
• Remove the input signals before you clean the Product.
• Use only specified replacement parts.
• Have an approved technician repair the Product.
If the Product does not perform properly, use these steps to help isolate
the problem:
1. Verify that battery capacity meets the requirements.
2. Verify that the input and output test lead is good and securely
connected.
Clean the Product
Clean with a soft cloth slightly dampened with a mild detergent.
WCaution
To avoid damage to the Product, do not use abrasives or
solvents to clean the Product.
If the Probe is not used for long time periods (>60 days) the battery
should be removed and stored separately.

General Specifications
Maximum between any Terminal
and Earth Ground........................ 1000 V ac rms, continuous, 1500 V dc
(Single-range)
Input Impedance ......................... 10 MΩ / <10 pF
Dimensions ................................. 138 mm x 75 mm x 45 mm
Input terminal .............................. 4 mm banana jacks
Output terminal............................ BNC (female)
Weight ......................................... 295 g, batteries included
Accessory Cable ......................... BNC to banana male to male, 1.6 m
Temperature
Operating ................................. -10 °C to 50 °C
Storage .................................... -30 °C to 60 °C
Operating Humidity ..................... Non-condensing (≤10 °C)
≤90 % RH (at 10 °C to 30 °C)
≤75 % RH (at 30 °C to 40 °C)
≤45 % RH (at 40 °C to 50 °C)
Altitude
Operating ................................. 2000 m
Storage .................................... 12 000 m
Ingress Protection ....................... IEC 60529: IP40 (Terminals Mated)
Battery
Type ......................................... 4 AA, IEC LR6
Battery Life............................... 168 hours (typical)
Safety
General .................................... IEC 61010-1: Pollution Degree 2
Measurement ........................... IEC 61010-2-030: CAT III 1000 V
CAT IV 600 V DC 1500 V
Electromagnetic Compatibility (EMC)
International ............................. IEC 61326-1: Portable
Electromagnetic Environment
IEC 61326-2-2
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or use
conductively coupled radio-frequency energy which is necessary for
the internal functioning of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low voltage power
supply network which supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted and radiated
disturbances.
Caution: This equipment is not intended for use in residential
environments and may not provide adequate protection to radio
reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.

Korea (KCC.............................. Class A Equipment (Industrial
Broadcasting and Communication
Equipment
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user should take
notice of it. This equipment is intended for use in business
environments and not to be used in homes.
USA (FCC ................................ 47 CFR 15 subpart B. This product is
considered an exempt device per
clause 15.103
Electrical Specifications
Measurement Range................... 1000 V ac rms, 1500 V dc
Accuracy ..................................... 0.1 % of reading +0.1 % of Range
Bandwidth ................................... 100 kHz (Max) See Figure 2.
Scale ........................................... 4 mV/V
Common Mode Rejection Ratio .. 60 dB at 50 Hz
Temperature Coefficients............ 0.05 % / °C (<18 °C or >28 °C )
Over-range Protection................. 10 % over range
Figure 2. Voltage versus Frequency Derating Curve
1200
1000
900
707
600
575
424
300
277
212
010k 20k 27k 40k 50k 60k 70k 80k 90k 100k
30k
Frequency (Hz)
U Vrms

LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship
for one year from the date of purchase. This warranty does not cover
fuses, disposable batteries, or damage from accident, neglect, misuse,
alteration, contamination, or abnormal conditions of operation or handling.
Resellers are not authorized to extend any other warranty on Fluke’s
behalf. To obtain service during the warranty period, contact your nearest
Fluke authorized service center to obtain return authorization information,
then send the product to that Service Center with a description of the
problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES,
SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED
OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not
allow the exclusion or limitation of an implied warranty or of incidental or
consequential damages, this limitation of liability may not apply to you.
11/99
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Beijing Fluke
Rm101, 1/F.,Tong Heng Tower
No. 4 Hua Yuan Road Hai Dian
District, Beijing 100088, P.R.C.
Table of contents
Other Fluke Accessories manuals

Fluke
Fluke RUBBER SHOE ACCESSORY PN 3125710 User manual

Fluke
Fluke Thermalert 4.0 Series User manual

Fluke
Fluke PPC4 User manual

Fluke
Fluke 80K-15 User manual

Fluke
Fluke 9040 User manual

Fluke
Fluke FLK-050-CLKT Instruction manual

Fluke
Fluke VIBCODE VIB 8.660 User manual

Fluke
Fluke 80PK-IR User manual

Fluke
Fluke 700P29 User manual

Fluke
Fluke 80PK-3A User manual

Fluke
Fluke 3563 User manual

Fluke
Fluke 80T-150UA User manual

Fluke
Fluke 805ES User manual

Fluke
Fluke Thermalert 4.0 User manual

Fluke
Fluke VIB 6.631 Owner's manual

Fluke
Fluke 80i-110s User manual

Fluke
Fluke 700P01Ex User manual

Fluke
Fluke 3562 User manual

Fluke
Fluke 750P User manual

Fluke
Fluke FLK-075-CLV Instruction manual