4- DESCRIPTION
The D-25Vplus peristaltic pumps in this manual mount the 153-3r head that allows easy access to the tube for
removal when it must be replaced.
On the 153-3r easy-load head, simply pull up and to the left of the lever (1) located on the top of the head to raise
the top and open the head for tube replacement.
When removing the tube, slide the buttons (3) up to move the tube retainers.
The head admits various tube sizes which, combined with the speed regulation, gives a great variety of flows, as
can be seen in the flow chart .
In the description on the front panel, the FULL button, nº 5, obtains the maximum speed of the motor, in the
loading, purging and cleaning operations and the control 2 allows to choose the direction of rotation of the motor
for the inversion of the flow and the unemployment. The cyclical timer nr. 1 has section 7 of this manual with the
instructions for its programming and the speed regulator nr. 3 provides control of the flow with a repeatability of
100%.
In the description on the back there is the connection for the network cable with integrated fuse holder, connection
for the pedal and the timing mode selector no. 12.
Consult the indicative table of dosages and install the appropriate tube.
NEVER CARRY OUT THIS OPERATION WITH THE EQUIPMENT RUNNING.
4.1- HEAD 153-3r :
1
3 3 3
4
2
2 4 4 4
Closed head Open head
1 - Lever to open the head
2 - Tractor axle output guard
3 - Sliding buttons for fixing the peristaltic tube
4 - Complementary head fixing screw holes
The 153-3r head admits another equal head on its same axis of rotation. Remove the protector 2 to face the
complementary head with the axis and insert the two assembly screws in the holes 4.
With 2 heads it is possible to cancel the typical cadence of the peristaltic flow. To do this, when locating the
complementary head, the rotor rollers must be oriented in opposition to the main head rollers.
When the heads are used as independent channels, each one will provide the flow corresponding to the tube
installed in the heads.
If it is intended to double the flow, the two suction and discharge tubes can be installed directly in the feeding and
receiving containers.
The option to use a Y-connection to join the tubes of both heads will allow the cancellation of the peristaltic pulse.
It must be remembered that the final discharge tubes and especially the suction ones must be of a larger diameter
than the head tube. If this is not possible, the total flow will be somewhat less than the theoretical expected.