DH Instruments SPLT User manual

© 1998-2009 DH Instruments, a Fluke Company
Self Purging Liquid Trap
(SPLT™)
User’s Manual

© 1998-2009 DH Instruments, a Fluke Company
High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be
released unexpectedly and with extreme force. High pressure systems should be assembled and operated
only by personnel who have been instructed in proper safety practices.
© 1998 - 2009 DH Instruments, a Fluke Company All rights reserved
Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any form or
by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments 4765 East Beautiful
Lane Phoenix AZ 85044-5318 USA.
DH Instruments, a Fluke Company makes sincere efforts to ensure accuracy and quality of its published materials; however, no warranty,
expressed or implied, is provided. DH Instruments, a Fluke Company disclaims any responsibility or liability for any direct or indirect
damages resulting from the use of the information in this manual or products described in it. Mention of any product does not constitute an
endorsement by DH Instruments of that product. This manual was originally composed in English and was subsequently translated into
other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the English version and other language
versions, the English version takes precedence.
DH Instruments, a Fluke Company, DH , DHI, PPC, PPC2, PPC3, PPC, and PPC2 AF are trademarks, registered and otherwise.
100521
3152142 REV 103
Printed in the USA.

TABLE OF CONTENTS
Page I © 1998-2009 DH Instruments, a Fluke Company
T
TA
AB
BL
LE
E
O
OF
F
C
CO
ON
NT
TE
EN
NT
TS
S
TABLE OF CONTENTS ..................................................................I
TABLES .....................................................................................II
FIGURES....................................................................................II
ABOUT THIS MANUAL ................................................................III
1. INTRODUCTION..................................................................... 1
1.1 PRODUCT OVERVIEW....................................................................................................................................... 1
1.2 SUBASSEMBLY DESCRIPTION AND LOCATION........................................................................................... 1
1.3 SPECIFICATIONS............................................................................................................................................... 2
2. INSTALLATION ..................................................................... 3
2.1 PNEUMATIC........................................................................................................................................................3
2.2 ELECTRICAL ......................................................................................................................................................4
3. OPERATION.......................................................................... 5
3.1 GENERAL OPERATING INSTRUCTIONS......................................................................................................... 5
4. MAINTENANCE...................................................................... 7
4.1 GENERAL............................................................................................................................................................ 7
4.2 REMOVING AND INSTALLING THE SPLT FILTER.......................................................................................... 7
4.3 SPLT EXHAUST VALVE REMOVAL AND REPLACEMENT............................................................................ 8
4.4 OVERHAUL......................................................................................................................................................... 8
4.5 PARTS................................................................................................................................................................. 9

SPLT™ USER’S MANUAL
© 1998-2009 DH Instruments, a Fluke Company Page II
T
TA
AB
BL
LE
ES
S
Table 1. Parts Listing ...........................................................................................................................................9
F
FI
IG
GU
UR
RE
ES
S
Figure 1. SPLT (Overall View) .............................................................................................................................1
Figure 2. PPC with SPLT and DUT......................................................................................................................3
Figure 3. Front View (Left) and Side View (Right)................................................................................................9

ABOUT THIS MANUAL
Page III © 1998-2009 DH Instruments, a Fluke Company
A
AB
BO
OU
UT
T
T
TH
HI
IS
S
M
MA
AN
NU
UA
AL
L
Manual Conventions
(CAUTION) is used in manual to identify user warnings and cautions.
(NOTE) is used in the manual to identify operating and applications advice and additional explanations.

SPLT™ USER’S MANUAL
© 1998-2009 DH Instruments, a Fluke Company Page IV
N
NO
OT
TE
ES
S

1. INTRODUCTION
Page 1 © 1998-2009 DH Instruments, a Fluke Company
1
1.
.
I
IN
NT
TR
RO
OD
DU
UC
CT
TI
IO
ON
N
1.1 PRODUCT OVERVIEW
The SPLT is a free standing accessory designed to protect the PPC from liquid contamination returning from
the system or device under test.
The SPLT is made up of a stainless steel body with an internal X-Type coalescing filter and a bottom drain port
fitting with an electrically actuated purge valve. The filter and valve assembly is installed in a mounting stand.
1.2 SUBASSEMBLY DESCRIPTION AND LOCATION
1. Purge Tube - Exhaust liquids and contaminants 3. Pressure Connections – 1/4 in. NPT F to test and to
PPC
2. Exhaust Valve - Solenoid valve to release pressure 4. Electrical Connection - Terminal block connections to
PPC valve driver for activation of exhaust valve
Figure 1. SPLT (Overall View)

SPLT™ USER’S MANUAL
© 1998-2009 DH Instruments, a Fluke Company Page 2
1.3 SPECIFICATIONS
Pressure Connections 1/4 in. NPT F
Power Requirements Apply 12 V to actuate valve, 6 watts
Weight 1.7 kg (3.8 lb)
Dimensions 10 cm H x 10 cm W x 17.5 cm D (3.9 in. x 3.9 in. x 6.9 in.)
Maximum Working Pressure 2 000 psi
Maximum Differential Across Filter 80 psi
Operational Capacity of Filter Body 10 cc

2. INSTALLATION
Page 3 © 1998-2009 DH Instruments, a Fluke Company
2
2.
.
I
IN
NS
ST
TA
AL
LL
LA
AT
TI
IO
ON
N
2.1 PNEUMATIC
Figure 2. PPC with SPLT and DUT
Be sure SPLT is at low point of pneumatic system.
The SPLT is intended to collect and exhaust liquid contaminants that may be present in the device or system
under test so that they do NOT return to the PPC.
The SPLT is installed in the test connection line at a low point between PPC and the device or system under
test. The SPLT pressure connections are 1/4 in. NPT female.
Teflon™ tape or another sealing material should be used to assure leak free connection of adaptors
installed in the SPLT pressure connections.
Proper SPLT operation is dependent on the gas flowing through it in the correct direction. Make sure to connect
the SPLT to the device or system under test and the PPC following the connector port labels on the SPLT.
Connect the SPLT electrically to the PPC rear panel DRIVER connection and to the SPLT terminal block (see
Section 2.2).
The SPLT must be mounted vertically with the purge valve at the bottom to perform properly. The SPLT MUST
be at a lower point in the system than the PPC and/or the device under test. In operation, liquid contaminants
collected from the device or system under test will be exhausted through the SPLT purge valve.

SPLT™ USER’S MANUAL
© 1998-2009 DH Instruments, a Fluke Company Page 4
Liquid contaminants will be forcibly ejected from the SPLT purge tube. Provision for collecting the purged
liquids should be considered when installing the SPLT.
Excessively dirty or wet filter elements can adversely affect pressure control.
2.2 ELECTRICAL
In order to take advantage of the automatic purging feature of the PPC, the SPLT valve MUST be connected to
the PPC external driver #8.
Connect like this if drivers connector with attached cable was included with SPLT:
Connect drivers connector to DRIVERS connector on back panel of PPC. Attach the two wires to the
terminal block on the SPLT
Connect like this if you have SPLT without drivers connector:
Solder (2) 24 gauge wires of desired length to the 12-pin connector provided with the PPC. Connect
(1) wire to pin L and the other to pin B. Connect drivers connector to DRIVERS connector on back
panel of PPC. Attach the other end of these wires to the terminal block on the SPLT.
Polarity is NOT significant to the operation of the purge valve.
Table of contents
Other DH Instruments Accessories manuals

DH Instruments
DH Instruments RUBBER SHOE ACCESSORY - SCHEMA P-N 400203 User manual

DH Instruments
DH Instruments 32MM MOLBLOC CRADLE KIT Quick start guide

DH Instruments
DH Instruments RUBBER SHOE ACCESSORY - SCHEMA P-N 400202 User manual

DH Instruments
DH Instruments RPM T-BOX User manual

DH Instruments
DH Instruments EXTERNAL RESERVOIR FOR PPCH User manual

DH Instruments
DH Instruments PG7000 User manual

DH Instruments
DH Instruments CONNECTOR BLOC FOR RPM4 User manual