Clements Orbital 360 User manual

User Manual
CLEMENTS
Orbital 360
Benchtop Centrifuge
CEN 91705
0- 40V 50/60Hz
Manual No. CEN 91705 017
Issue 1

CLEMENTS
User Manual Orbital 360 Centrifuge
Manual Number CEN 91705 017 Issue 1
Copyrig t © 2014 BMDi TUTA Healt care Pty Ltd
T e information in t is manual was originated by, and is t e exclusive
property of BMDi TUTA Healt care Pty Ltd. It is furnis ed for customer
information only, and is not an aut orisation or licence to make t is
product or to furnis t is information to ot ers.
BMDi TUTA Healthcare Pty Ltd
Unit 4B, 1 8-130 Frances St
Lidcombe NSW 141
Australia
Phone: +61 9466 5300
Website: www.medaust.com
Safety
T ank you for purc asing t is Clements Orbital 360
Centrifuge For your safety it is imperative t at t is unit
only be operated by aut orised personnel in accordance
wit t e instructions as described in t is manual. Operated
in t is way, t e Orbital 360 Centrifuge will provide t e
standard of service specified.
Due to continual improvements in product design, t e
Orbital 360 Centrifuge may vary in detail from t e
descriptions in t is manual. In t e event of furt er
questions please contact your local distributor or BMDi
TUTA Healt care direct.

3
CLEMENTS
Contents
Contents
Description and Features.................................................... 4
Specifications ..................................................................... 5
Warnings and Cautions ...................................................... 6
Installation .......................................................................... 7
Controls.............................................................................. 8
Operation ..........................................................................10
Parameter Setting Menu....................................................12
Operation Summary ..........................................................14
User Programs ..................................................................16
Rotors................................................................................17
Accessories and Spares....................................................19
Cleaning and Maintenance ................................................20
Fuse Replacement and Error Codes..................................21
Troubleshooting.................................................................22
RCF Graphs ......................................................................23
Warranty............................................................................26

4
CLEMENTS
Description and Features
Description
T e Orbital 360 Centrifuge is a versatile, compact and reliable
benc top centrifuge. It is designed for clinical and small
laboratory applications w ere precise digital control of speed and
time is required.
Features
• Brus less DC motor tec nology for low maintenance
operation
• Accurate digital control of speed and time for repeatable
processing
• Hig visibility backlit LCD display wit touc control panel
• Electronic lid interlock to ensure t at centrifuge can be
started and run only wit lid locked
• Stainless steel bowl
• Adjustable acceleration and deceleration for sensitive
samples
• Concurrent display of rpm and RCF
• 25 user programs
• Simultaneous display of all parameters
• Previous parameters maintained t roug power off/on

5
CLEMENTS
Specifications
Specifications
Speed Range 100 to 5000 rpm
Maximum RCF 4100 x g
Speed Accuracy ±50 rpm
Time Range 1 to 99 minutes
Angle Rotors 24 x 10 ml
Swing-out Rotors 4 x 50 ml
12 x 15 ml
24 x 5 ml
Power Requirement 220-240V 50/60Hz 5A
Electrical Protection Class I
Noise < 65dB(A)
Dimensions 370 (W) x 250 (H) x 432 (D) mm
Weight 15kg
Environment Indoor
10 to 35°C, RH < 85%
Package Contents
Item
Description
Qty
CEN 91705 Cell Cycle 360 Centrifuge 240V 50Hz 1
CEN 91700 036 Tool, Rotor Screwdriver, Philips Head,
110 x 5.5mm
1
CEN 91700 037 Tool, Lid Release Screwdriver, Philips
Head, 120 x 2.9mm
1
CEN 91700 038 Lead, Mains Power, IEC Socket –
Earthed Australian Plug
1
CEN 91700 040 Kit, Fuse, 5 x 20mm, T5.0A 250V
(Pack 4)
1

6
CLEMENTS
Warnings and Cautions
For safe operation ensure t at users read, understand and
observe t e instructions and recommendations in t is user
manual.
Power
Ensure t at t e mains power supply matc es t at s own on t e
rating label of t e centrifuge.
Flammable Samples
T e centrifuge is not flameproof. Do not use flammable samples in
t e centrifuge. Do not store flammable or volatile material wit in
300mm of t e centrifuge.
Biohazardous Samples
Local protocols must be observed w en andling potentially
bio azardous materials. Carriers wit aerosol protection are to be
used.
Fluid Spill
Do not allow fluids to enter t e centrifuge. If a spill occurs wipe
clean, paying attention to t e nature of t e material.
Balancing
T e centrifuge must always be run wit a full complement of
buckets or carriers. Balance samples eit er by eye or by weig ing.
Fill unused tubes wit an equivalent amount of fluid as in sample
tube.
Operating
Do not stand wit in 300mm of centrifuge w ile it is running.
Do not move centrifuge w ile t e rotor is spinning.
Do not open lid w ile rotor is spinning.
Do not attempt to stop rotor by and.
Warnings and Cautions

7
CLEMENTS
Installation
Unpacking
Keep centrifuge orizontal w ile removing it from carton. Retain
packaging for transporting centrifuge. Wit power off, release lid
lock by inserting manual release tool in small ole on top rig t
and side of centrifuge. Keep tool parallel to front of panel and
press release plate. Remove tool.
Remove packing, carriers etc from centrifuge bowl. Ensure t at
bowl is clean.
Location
T e centrifuge s ould be placed on a solid level surface in a well
ventilated area. T e back of t e centrifuge must be at least 10cm
away from any wall or vertical surface.
Power Requirements
Before connecting t e centrifuge to t e mains power supply,
ensure t at t e electricity supply conforms to t e requirements for
voltage and frequency s own on t e rating label on t e back of t e
unit.
T e power supply s ould be stable, clean and free of surges and
brownouts.
Connect t e centrifuge to t e mains power supply using t e
supplied eart ed power cable.
For extended periods of non-use, switc off main power switc on
rear panel of centrifuge.
Installation

8
CLEMENTS
Controls
Controls
Manual
Lid release
access hole
Control panel
Front View Rear View
On/Off Switch IEC power inlet
Fuse holder
Condition LCD Display Shows
In Spin Cycle Actual and Set Values of Speed / RCF and Time
Remaining
Setting Mode Set Values of Parameter
Error Error code n of form En
360

9
CLEMENTS
Controls
Controls
Item Function
On/Off Switch Switches mains power to centrifuge.
IEC Inlet Mains power inlet. Also contains fuse holder.
Lid release access hole Access for manual lid release tool. Centrifuge
must be disconnected from power.
S
TOP
/ O
PEN
Button Stops spin cycle. When rotor has stopped acts
as lid release button. Cancels error condition.
S
TART
Button Starts spin cycle.
P
ULSE
Button Spins rotor when held down.
E
NTER
Button Enters and saves changed parameters.
Exits parameter setting mode.
Selects user program. Exits program mode.
I
NC
Button Increases value of set parameters.
D
EC
Button Decreases value of set parameters.
P
ROG
Button Enters user program mode.
Times out after 5 seconds.
M
ENU
Button Steps through parameter change menu in
circular fashion. Times out after 5 seconds.
S
PIN
LED (Green) Indicates that spin cycle is in progress.
S
TOP
LED (Red) Flashing - indicates that spin cycle is stopping.
Continuous - indicates that rotor has stopped.

10
CLEMENTS
Operation
Rotor Fitting
1. Open lid and switc off centrifuge.
2. Remove carriers from rotor.
3. W ile supporting rotor wit one and, unscrew (CCW) rotor-
retaining cap screw.
4. Lift off rotor being careful not to bend or apply side load to
motor s aft.
5. Select a compatible rotor, align t e slot in t e rotor s aft
wit t e drive pin in t e motor s aft and fit toget er.
6. Fit cap screw and was er to rotor.
7. W ile supporting rotor wit one and, and tig ten (CW)
rotor-retaining cap screw - again being careful not to bend
or apply side load to motor s aft..
8. Confirm t at rotor rotates smoot ly and evenly by and.
Carrier Fitting (
swing-out style rotors)
1. Fit a full complement of compatible carriers to t e rotor.
2. Ensure t at carriers are fully seated and can swing freely.
3. Switc centrifuge power on.
Tube Fitting
Select a full complement of compatible tubes wit an RCF
rating in excess of t e maximum RCF of t e centrifuge.
Balancing of Samples
It is important t at t e samples are correctly balanced wit in t e
centrifuge prior to starting t e unit. For correct balance t ere are
two requirements:
1. Balance t e weig t of t e samples
2. Balance t e location of t e samples
Operation
Table of contents
Popular Laboratory Equipment manuals by other brands

Agilent Technologies
Agilent Technologies 5800 ICP-OES user guide

Endress+Hauser
Endress+Hauser Cleanfit CPA875 operating instructions

NI
NI PXI-5422 CALIBRATION PROCEDURE

Collomix
Collomix Aqix operating instructions

SPEX SamplePrep
SPEX SamplePrep 6875 Freezer/Mill Series operating manual

Ocean Insight
Ocean Insight FLAME-NIR+ Installation and operation manual

Parker
Parker ALIGN-MG-NA Installation, operation and maintenance manual

BD
BD 644787 user guide

DENTAURUM
DENTAURUM Compact Megaplus Instructions for use

Biuged Laboratory Instruments
Biuged Laboratory Instruments BGD 626 instruction manual

VWR
VWR SAS Super IAQ instruction manual

illumina
illumina MiSeqDx reference guide