pixlTM Real-Time PCR Platform
Instructions for Use Manual
Page 6 of 37
Use of power supply
Before the instrument adapter is connected to the power cord, it must be ensured that
the AC power supply voltage (100 to 240 VAC) and frequency (50/60Hz) are consistent
with those required by the instrument adapter. When making the power cord connection,
make sure that the instrument power switch is off. Do not touch the power switch and
power cord with wet hands. Do not disconnect the power cord when the instrument is not
powered. Keep power cord away from hot surfaces. Do not clean the instrument when it is
not disconnected. Please turn off the power when the instrument is no longer in use.
Power cord
The instrument should normally use the power cord supplied with it. If the power cord is
broken, it must be replaced without repair. When replacing the power cord, it must be
replaced with the same type of power cord of the same specifications. When this
instrument is in use, do not place anything on the power cord, and do not place the power
cord in a place where people move around.
Connect the power cord and ensure that the plug is secured into the socket. Disconnect
power cord at the adapter connection point.
Placement of the Instrument
1) The instrument should not be placed in a location where it is difficult to disconnect the
power supply.
2) The instrument uses semiconductor cooling and fan-assisted heat dissipation, so when
placing the instrument, ensure that there is no obstacle within 15 cm around the
instrument, and when multiple instruments are used at the same time, the distance
between each instrument should be not less than 30 cm.
3) The instrument should be placed in a location with the following parameters: low
humidity, dust-free, no running water, well ventilated, no corrosive gases, no magnetic
field interference, away from direct sunlight or strong light sources. Place the
instrument on a horizontal and stable surface.
4) High ambient temperature will affect the testing performance of the instrument or
cause malfunction. Do not use this instrument in the place of direct sunlight and strong
light source to avoid affecting the fluorescence detection of the instrument, and it
should be far away from heating, stove, and all other heat sources.
5) Turn off the power when the instrument is not in use. When the instrument is not used
for an extended period, unplug and store to prevent dust accumulation.