Troubleshooting
1. The AlcoSense AccessPoint cannot power on
Reason: Power supply is not connected correctly
Solution: Check that the red cable voltage is 12V.
If problem still persists, please contact the manufacturer.
2. Red Status lamp
Reason: Breath sampling is not strong enough for a breath analysis (Flow error)
Solution: Wait until the Status lamp is green, then blow steadily into the sampling hole.
3. All 3 lamps (Status, Pass and Fail lamps) keep blinking
Reason: White cable is connected to the door sensor or door is not closed properly
Solution: Check the door sensor and whether the door is opened or closed
4. Analyzing and showing resul t is very slow, or no result provided after breath sample:
Reason: The AccessPoint sensor module needs calibration or replacing. If the unit is not providing a
result, the flow sensor in the sensor module may be damaged.
Solution: Please contact the manufacturer.
Specifications
Indication of B.A.C.
Test result indicated by LED display
PASS : Indicate Green Lamp
FAIL : Indicate Red Lamp
(Level for PASS can be adjusted)
Accur acy ±10% at 0.500 g/L
Sensor Replaceable fuel cell sensor
Mouthpiece Not required
Power supply DC12V1.5A
Power r equirements Minimum 12V DC 1amp (Amp requirement varies with connected components)
Warm up time 3 seconds ~ 4 minutes
Response time Within 3 seconds at 0.000 g/L
Within 10 seconds at 1.000 g/L
Calibration inter val Recommended 12 months or 1,000 tests
Ambient conditions Storage: 0 to 50/ Operation: 5 to 40
Dimensions 206 (height) × 108 (width) × 37 (thickness) mm
Weight 674g
9