YOKOGAWA CW10 User manual

User’s
Manual
保証書付
CW10
クランプ電力チェッカ
Clamp-on Power Meter
IM CW10-JA: Japanese
IM CW10-EN: English
5th Edition: April 2015 (YMI)
この取扱説明書は、いつでも使用できるよう
大切に保管してください。
Store this manual in an easily accessible place
for quick reference.

1
Regarding the Safe Use of This Product
When operating the instrument, be sure to observe the cautionary
notes given below to ensure correct and safe use of the instrument.
If you use the instrument in any way other than as instructed in
this manual, the instrument’s protective measures may be impaired.
YOKOGAWA is by no means liable for any damage resulting from
use of the instrument in contradiction to these cautionary notes.
The following safety symbols are used on the instrument and in
this manual.
Danger! Handle with Care.
This symbol indicates that the operator must refer to an
explanation in the User’s Manual in order to avoid risk of
injury or death of personnel or damage to the instrument.
Hazardous Voltage
This operator must never attempt to touch equipment or
parts marked with this symbol.
This symbol indicates double insulation or reinforced
insulation.
This symbol indicates direct current (DC).
This symbol indicates aiternating current (AC).
This symbol indicates ground (earth).
This symbol indicates a battery.
This symbol indicates that this instrument designed to be
applied around or removed from HAZARDOUS LIVE
conductors provided if the RATED circuit-to-earth voltage
dose not exceed the value indicated in the measurement
category.
This symbol indicates, If the measured voltage is
greater than 30 V DC or AC RMS or Overload
(OL or –OL).
Indicates a hazard that may result in the loss of life
or serious injury of the user unless the described
instruction is abided by.
Indicates a hazard that may result in an injury to
the user and/or physical damage to the product or
other equipment unless the described instruction is
abided by.

2
Always observe the following instructions. Failure to do so may result
in electrical shock or other dangers that may lead to serious
injury or the loss of life.
• The instrument is a power measurement instrument that can measure
parameters such as voltage, current, and power.
Do not use this instrument for any other purpose.
• Do not use the instrument if there is a problem with its physical
appearance.
• Do not use the instrument in an atmosphere where any ammable or
explosive gas is present.
• To avoid a short-circuit or an accident to personnel, use this instrument
within the RATED circuit-to-earth voltage, maximum input voltage and
current.
• Avoid using the instrument if it has been exposed to rain or moisture or
if your hands are wet.
• Barrier is for to avoid touching the HAZARDOUS LIVE conductor.
Be careful not to across the Barrier when using the instrument.
• Do not use the instrument if there is any damage to the casing or
when the casing is removed.
• Use the test leads supplied by Yokogawa with this instrument.
• Do not use test leads that have deteriorated or are defective.
• Check the test leads continuity.
• When you attach or remove the test leads or remove the case
(for example to charge the battery), be sure to remove the test leads
from the circuit under measurement.
• When you remove the case (for example to change the battery),
be sure to remove the test leads from the instrument.
• Safety protectors such as rubber-insulated gloves should be worn to
prevent electrical shock when using the instrument.
• There are caps at the ends of the test leads. To ensure safety
(safety standard EN 61010-031), be sure to put the caps on the leads
when you use them.
• No person, except personnel from Yokogawa, is authorized to
disassemble this instrument.
Do not attempt to repair/modify the instrument yourself,
as doing so is extremely dangerous.
• The product is for domestic use (Class B) and meets the electromagnetic
compatibility requirements.
• Do not use the instrument near noise-emitting equipment or where
there may be sudden changes in temperature. Otherwise,
the instrument may produce an unstable readings or errors.
• Do not wipe the instrument using an organic solvent.
When cleaning the instrument, use a dry cloth.
• Do not leave the tester exposed to direct sunlight or in a hot and humid
location such as the inside of a car, for any prolonged length of time.
• If the instrument will not be used for long periods, remove the battery.

3
Contents
Regarding the Safe Use of This Product.......................1
1. Checking the Contents of the Package ....................4
2. Measurement Category ................................4
3. Components.........................................5
4. Display (LCD)........................................6
5. Power ON/OFF and Auto power off .......................6
6. Cursor/ENTER key....................................6
7. Making Basic Measurements ............................7
7.1 Measuring Voltage ...............................7
7.2 Measuring Current ...............................7
8. Others Function ......................................8
8.1 AUTO SENSE...................................8
8.2 RANGE ........................................8
8.3 Peak Hold (AC mode only) .........................9
8.4 Inrush current (AC mode only) .....................11
8.5 DCA ZERO ....................................12
8.6 Measuring Frequency Hz (AC mode only) ............12
8.7 MAX/MIN......................................13
8.8 Harmonic Measurement (AC mode only) .............13
8.8.1 THD Measurement..........................13
8.8.2 Individual Harmonic Mesurement ..............14
8.9 Low pass filter (AC mode only).....................14
8.10 Measuring Active Power ..........................15
8.10.1 Single phase 2 wire .........................15
8.10.2 3 phase 3 wire (balanced/unbalanced) ..........16
8.10.3 3 phase 4 wire (balanced/unbalanced) ..........16
8.10.4 Phase Rotation.............................17
8.11 Mesuring Resistance.............................19
8.11.1 Resistance (Ω) .............................19
8.11.2 Continuity Check ...........................20
8.11.3 Diode Test.................................20
8.12 HOLD ........................................20
8.13 Voltage Sanse..................................20
8.14 Buzzer........................................21
8.15 Additional Power-on Function......................21
9. Battery State display and Battery Replacement .............21
9.1 Battery State display .............................21
9.2 Bayttery Replacement............................21
10. Specifications .......................................22
10.1 General Specifications ...........................22
10.2 Accuracy ......................................23
11. Maintenance and After-Sales Service ....................26
12. Sales in Each Country or Region ........................26
12.1 Disposing the Product............................26
12.2 How to Replace and Dispose the Batteries............26
12.3 Authorized Representative in the EEA ...............26
12.4 For the Pollution Control of Electronic and
Electrical Products of the People's Republic of China ...27
Disk No. CW10
5th Edition: April 2015
All Rights Reserved, Copyright ©2011,
Yokogawa Meters & Instruments Corporation
Printed in Taiwan

4
1. Checking the Contents of the Package
After opening the package, be sure to check the product as
instructed below before use.
1. Main Unit: CW10 (including battery)
2. Test leads: 1set (Black and Red)
3. User’s Manual
4. Soft Case
2. Measurement Category
When you use the test leads, attach or remove the caps
according to the measurement category.
CW10 main unit:
1000V
CAT III, 600V CAT IV
Test leads:
With Caps: 1000V 10A CAT III, 600V 10A CAT IV
Without Caps: 1000V 10A CAT II
Measurement
Category Description Remarks
O
None
Other
Other circuits that are not
directly connect to MAINS.
Circuits not
connected to a
mains power source.
CAT II For measurement performed
on circuits directly connected
to the low-voltage installation.
Appliances,
portable equipment,
etc.
CAT III For measurement performed
in the building installation. Distribution board,
circuit breaker, etc.
CAT IV For measurement performed
at the source of the low-voltage
installation.
Overhead wire,
cable systems, etc.

5
3. Components
10.Terminals of test leads
5.Hold and Zero key
6.Function switch
9.Cursor/
Enter key
8.LCD display
1.Jaw section
4.Barrier
7.Open/
close lever
3.Voltage sense
LED
2.Light
CW10
Red
Black
Test leads
Caps of Test leads
Main Unit: CW10
1. Is a precision sensor for detecting currents.
2. When the jaws are open, the light will illuminate.
3. In case voltage is detected by the jaw, the red LED lights up.
4. Prevents contact with the wires.
5. Retains the measured data or DCA ZERO: Zero adjustment
6. Use this switch to turn the power on and off and to select
the measurement mode.
7. Opens and closes the jaws.
8. LCD display
9. Cursor and Enter key
10. Input terminals for test leads
[NOTE]
There is the mark on the terminal, but this device does not
have a function for measuring capacitance.

6
4. Display (LCD)
5. Power ON/OFF and Auto power off
Check that the meter operates normally.
Turn the function switch to the measurement (any) position from OFF.
To end measurement, turn the function switch off.
<Auto power off>
Indication: APO
This instrument automatically turns off about 15 minutes after
the last switch operation.
Recovering the Auto power off function:
Turn off the power. Turn on the power.
When the function is not use (cancellation):
Turn off the power. Hold down Downward of cursor key,
turn the function switch to any position.
(The APO display goes off.)
If the instrument is used with the Auto power off function
cancelled, take care not to let the battery run down.
6. Cursor/ENTER key
ENTER
LEFT
UP
DOWN
RIGHT
UP
Cursor key: Moves the Cursor (four directions) to the item (mode)
of display you wish select. Press the convex key
toward selecting mode.
ENTER key: Confirms the item (mode).
[NOTE]
Push whole key toward requisite item.
Push down the center position of key to confirm (ENTER key).

7
7. Making Basic Measurements
7.1 Measuring Voltage
Do not apply more than 1000 VDC or 1000 Vrms (1414.2 Vpk).
Plug the test leads to the input terminals.
[NOTE]
When connecting the test leads to the circuit under measurement,
connect the black (common) before connecting the red (live);
when removing the test leads, remove the red (live) before
removing the black (common).
Turn the function switch to the Voltage position.
Read the voltage value of display.
On Auto Sense mode, the CW10 automatically detects AC or DC.
DC AC
Black Red
Test leads
7.2 Measuring Current
• Do not apply more than 600 ADC or 600 Arms (848.5 Apk).
• When you measure the current, be sure to remove the test leads
from the instrument.
Squeeze the open/close lever to open the jaws. Insert a wire from the
measurable conductors under the test though the jaws, making sure
the tops of the jaws are tightly shut.
Turn the function switch to the Current position.
Read the current value of display.
Using Auto Sense mode, the CW10 automatically detects AC or DC.
[NOTE]
• When performing a measurement, hold the instrument so that
the measured conductor cable runs at the center of the clamp.
• Ensure that the orientation of the clamp to the direction of
the conductor wire is correct as shown right.

8
AC
DC
8. Others Function
8.1 AUTO SENSE
Using Auto Sense mode, the CW10 automatically detects AC or DC.
(The detection is determined by whichever input is greater.)
To select manual mode (Example: DC mode);
Use the cursor key to select under “DC” (flash) and then press the
ENTER key to confirm.
Upper “DC” indication is fixed and AUTO (AUTO RANGE) is displayed.
To return AUTO SENSE, press the ENTER key
for more than 2 seconds.
8.2 RANGE
To select manual range (fixed);
Use the cursor key to select “RANGE” and then press the ENTER key
to confirm.
(To cancel manual range, press the ENTER key
for more than 2 seconds.)
Example: ACV 100 V range ← → 1000 V range
ENTER
On Auto sense mode, the measuring range is automatic.

9
8.3 Peak Hold (AC mode only)
(1) AC Voltage
Turn the function switch to the voltage (V) position.
After using the ENTER key to confirm AC mode, use the cursor key to
select Peak Hold and then press the ENTER key to confirm.
Press the ENTER key to switch between PEAK MAX (polarity+) and
PEAK MIN (polarity-).
(To cancel Peak Hold, press the ENTER key
for more than 2 seconds.)
>2SEC
"ENTER"
[NOTE]
Bar-graph indication is not PEAK value.
Bar-graph indication is affected by the ACV value.
(Example: 500.0 V)
Table of contents
Other YOKOGAWA Multimeter manuals

YOKOGAWA
YOKOGAWA TY520 User manual

YOKOGAWA
YOKOGAWA 90050B User manual

YOKOGAWA
YOKOGAWA 73101 User manual

YOKOGAWA
YOKOGAWA 73201 User manual

YOKOGAWA
YOKOGAWA 732 01 User manual

YOKOGAWA
YOKOGAWA 734 01 User manual

YOKOGAWA
YOKOGAWA 7555 User manual

YOKOGAWA
YOKOGAWA 733 01 User manual

YOKOGAWA
YOKOGAWA 73201 User manual

YOKOGAWA
YOKOGAWA CA450 User manual
Popular Multimeter manuals by other brands

Gossen MetraWatt
Gossen MetraWatt METRAmax 6 operating instructions

PeakTech
PeakTech 4000 Procedure of calibration

Gossen MetraWatt
Gossen MetraWatt METRALINE DMM16 operating instructions

Fluke
Fluke 8846A Programmer's manual

Tempo Communications
Tempo Communications MM200 instruction manual

Elenco Electronics
Elenco Electronics CM-1555 operating instructions