Xsensor ForeSite SS User manual

User Guide

1
ForeSite SS User Guide Revision
All trademarks appearing in this document belong to XSENSOR
Technology Corporaon.
© XSENSOR Technology Corporaon. All rights reserved.
The XSENSOR Pressure Imaging System is protected by copyright
law. Unauthorized reproducon or redistribuon of XSENSOR
soware is prohibited. No part of this document may be
reproduced or transmied in any form without the express
wrien permission of XSENSOR Technology Corporaon.
ASY-05-00005-02

1
TABLE OF CONTENTS
INTRODUCTION 1
The ForeSite SS System 3
Overview 3
Intended Use 3
Product Compliance 4
Safety Standards 4
Emission & Immunity Standards 5
Warnings&Cauons 11
FORESITE SS HARDWARE 15
ForeSite SS Tablet Wireless System 17
ForeSite SS Tablet System 21
DisplayTabletSpecicaons 27
Maintenance 30
Storage & Transport 30
Cleaning 32
Cleanable Sensor Instrucons 33
FORESITE SS SOFTWARE 37
Introducon 39
QuickStart 39
User Roles & Security 41
AdministraveSengs 41
StandardSengs 45
Client Management 48
PressureImaging 49
Gallery Management 51
APPENDIXES 55
Appendix1-Troubleshoong 56
Appendix 2 - The Desktop System 58
Appendix3-ContactInformaon 64


Introducon
The ForeSite SS System 3
Overview 3
Intended Use 3
Product Compliance 4
Safety Standards 4
Emission & Immunity Standards 5
Warnings&Cauons 11


Introducon 3
The ForeSite SS System: Overview
Thank you for purchasing the ForeSite SS Pressure Imaging System!
The ForeSite SS sensor is embedded with thousands of sensor cells that
connuously measure the level of pressure between two solid objects.
For clinical purposes, this may be the level of pressure between a
wheelchair and a client’s body surface. The sensor connects to a touch
screen tablet that displays live pressure informaon and stascs on
a colour-coded scale. The resulng pressure images may be rotated,
captured as snapshots, recorded as a series of frames, and even
displayed in 3D.
ForeSite SS is designed to be intuive and user-friendly in an eort
to help wheelchair seang professionals make more informed clinical
decisions
Wheelchair users are at an increased risk of developing pressure ulcers
which can form when muscles and so ssue press against surfaces
such as a chair or a bed. As such, preventave measures must be taken
in order to reduce these risks. ForeSite SS provides clinicians with
the tools they need to help their clients reduce the risk of developing
pressure ulcers. By ulizing ForeSite SS, clinicians are able to create an
image of the relave pressure distribuon over a surface. The dynamic
pressure images then allow them to evaluate the pressure distribuon,
and, if necessary, take appropriate measures to enhance the pressure
distribuon for that parcular surface.
ForeSite SS is calibrated at the factory to improve the overall accuracy
of the pressure distribuon images. However, the system is not
intended to replace current best pracces for prevenon of pressure
ulcers. It is intended to provide addional monitoring and data
collecon while standard monitoring techniques and turn schedules are
maintained.
The ForeSite SS System: Intended Use
!

Introducon4
Product Compliance: Safety Standards
The XSENSOR Pressure Imaging System complies with the following
standards:
√IEC 60601-1:2005, CSA C22.2 No. 60601-1:2008, ANSI/AAMI ES 60601-1/
A2:2010, UL 60601-1:2003
√Mode of Operaon
√Class I Medical Equipment
√Type B Applied Part. Only the sensor(s) touch the paent in normal use.
The ForeSite SS is intended for use by healthcare professionals only.
ForeSite SS must be installed and used as described in this user manual.
This system is designed in compliance with medical safety standards and is
intended for the use of clinicians only. It is not recommended for private use
or client use without the assistance of a trained medical professional. There
are no known contraindicaons to its use. For any quesons or concerns in
regards to the safety of this product, do not hesitate to contact XSENSOR
(see Appendix 4).

Introducon 5
Product Compliance:
Emission & Immunity Standards
The ForeSite SS is a class A digital device suitable for business, industrial and
commercial environments. It is not intended for residenal use. The ForeSite
SS may cause radio interference or may disrupt the operaon of nearby
equipment. Portable and mobile RF communicaons equipment can also
aect ForeSite SS. It may be necessary to take migaon measures, such as
reorienng or relocang the equipment or shielding the locaon.
The ForeSite SS should not be used adjacent to or stacked with other
equipment. If adjacent or stacked operaon is necessary, the Pressure
Imaging System should be observed to verify normal operaon in the
conguraon in which it will be used.
Use of accessories and cables other than those provided by XSENSOR may
result in increased emissions or decreased immunity.
ForeSite SS complies with Part 15 of the FCC Rules. Operaon is subject to the
following condions: (1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference
that may cause undesired operaon.
Standards Compliance:
√EN 60601-1-2:2007 (Ed3.0) Medical Electrical Equipment – Part 1:
General Requirements for Safety 2. Collateral Standard: Electromagnec
Compability (Non-life-supporng, Group 1, Class A Equipment)
√FCC 15.107/CISPR 11 (Class A) Conducted Emissions
√FCC 15.109/CISPR 11 (Class A) Radiated Emissions
√Industry Canada ICES-003(A)/NMB-3(A) Compliant
√IEC 61000-3-2:2005 Harmonics Emissions
√IEC 61000-3-3:2008 Flicker Emissions
√IEC 61000-4-2:2008 Electrostac Discharge
√IEC 61000-4-3:2008 Radiated Suscepbility
√IEC 61000-4-4:2007 Electrical Fast Transients
√IEC 61000-4-5:2005 Surge Suscepbility
√IEC 61000-4-6:2006 Conducted Suscepbility
√IEC 61000-4-8:2001 Magnec Suscepbility
√IEC 61000-4-11:2007 Dips & Interrupts

Introducon6
Guidanceandmanufacturer’sdeclaraon–
electromagnecemissions
The ForeSite SS system is intended for use in the electromagnec
environments specied below. ForeSite SS system users should ensure
that it is used in such an environment.
Emissions Test Compliance Electromagnec
Environment Guidance
RF emissions CISPR 11 Group 1 The ForeSite SS system
uses RF energy only for
its internal funcon.
Therefore, its RF
emissions are very low
and are not likely to
cause any interference
in the nearby electronic
equipment.
RF emissions CISPR 11 Class A The ForeSite SS system
is suitable for use in all
establishments other
than domesc and those
directly connected to the
public lowvoltage power
supply network that
supplies buildings used
for domesc purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuaons/
icker emissions
IEC 61000-3-3
Complies
Recommendedseparaondistancesbetween
portableandmobileRFcommunicaonsequipmentandtheForeSiteSS
The ForeSite SS is intended for use in an electromagnec environment in which
radiated RF disturbances are controlled. The customer or the user of the ForeSite SS
can help prevent electromagnec interference by maintaining a minimum distance
between portable and mobile RF communicaons equipment (transmiers) and the
ForeSite SS as recommended below, according to the maximum output power of the
communicaons equipment.
Rated maximum
output power
of transmier
W
Separaon distance according to frequency of transmier
M
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.24
0.1 0.37 0.37 0.74
11.17 1.17 2.34
10 3.69 3.69 7.38
Table of contents