Uniks RAPID PRO User manual

USER MANUAL
POCKET MULTIFUNCTION TESTER
© Copyright UNIKS Srl Ver EN 3.0 - 04/07/2022

2

3
PRODUCTS REGISTRATION
REGISTER YOUR PRODUCT NOW ON :
hps://my.uniks.it
Registering the RAPID PRO will allow you to
access your RESERVED AREA to download
the measurements and create REPORTS.
pag.4
pag.4
pag.5
pag.6
pag.7
pag.8 - 9
pag.10
pag.11
pag.12
pag.13
pag.14 -15
pag.16
pag.17
pag.18
SUMMARY
INTRODUCTION
GENERAL DESCRIPTION
SAFETY and WARNINGS
PRODUCT DESCRIPTION
VOLT e SOCKET TEST
RCD (RESIDUAL CURRENT DEVICE)
LOOP
LOOP TN - TT
HARMONICS
MEM MEMORY / REPORT
TECHNICAL DATA
ADAPTER
WARRANTY
ASSISTANCE

4
INTRODUCTION
Maximum Voltage input is 265 V .Do not connect the instrument to higher voltage .Permanent
damage to the instruments and electrical shock can occur
if you don’t respect this guidelines
Congratulaon for purchase Rapid Pro, the rst mulfuncon tester in the shape of
a socket tester.It allow the test of electrical system according the IEC standard as the
common installaon tester do.In addion to the TRMS Voltage measurement and
the Socket Tester features allow Installaon test to be performed from anyone ( RCD,
LOOP L-PE ) in according to IEC61557-3 and IEC61557-6.
Super bright color Display TFT 1,77” With internal memory for store results.Super-
Cap technology let the instruments turn on even when disconnected from the socket
This products is manufactured following IEC/EN61010-1, guidelines for
safety installaon testers.Follow this user manual for avoid damage the
instruments :
• Do not measure in high humidity or wet enviroment.
• Avoid measurement in highly ammable place.
• Avoid contact with the circuit unders test if you are not measuring
• Avoid contact with metallic part that can be under voltage
• If the instruments look broken or damaged please avoid the use for measure.
SAFETY and WARNINGS

5
GENERAL DESCRIPTION
Rapid Pro execute the following test:
1. VOLT and Socket Test (Correct wiring of Plug)
2. Voltage measurement L-N , TRMS
3. Residual Current (RCD) type A, AC and F also in AUTO mode with
CONTACT VOLTAGE measurement
4. Z LOOP (Global earth resistance in TT system and impedance L-Pe
in TN with short circuit current )
5. Short circuit current
6. Voltage harmonics up to 49 th
7. THD% (Total harmonics Distoron)
8. Frequency of the fundamental Harmonics and up to 49 th
Funcon can be selected thorugh the dedicated buon .Sub funcon fea-
tures can be selected through the arrow buon for example for selecng
the RCD type or evaluate dierent harmonics. The instruments is powered
from the line and due to the SUPERCAP technology can keep the power
even when disconnected from the line .Fully charged supercap allow 45
second aer switch o from the line.This allow the user to read the value
aer the RCD is tripped out(with tripping me) or when is dicult to read
the display value
Outcome are always GREEN when the measured value is correct in accor-
ding the local regulamentaon or RED when the test is not passed.
In the SOCKET TEST uncorrect value the instruments generate also a
sound allarm.

6
PRODUCT DESCRIPTION
Descrizione dei tasti
Funcon buon allow to select the measure. Display shows the following features:
Socket Test / VOLT - RCD - LOOP - Harmonics / Memory.
Arrow buon ,allow to select subparameters for test and read the memory or to
analyze the harmonics value up to 49th. Red buon can start the RCD test or LOOP
Resistance ( no trip RCD).
1
2
3
4
5
6
7
Legend:
1. Anscratch glass
2. Display TFT 1,77”
3. Funcon
4. TEST
5. subfeatures selecon
6. subfeatures selecon
7. SCHUKO plug

7
VOLT e SOCKET TEST
Fig. 1
Fig. 2 Fig. 3
Legend:
1. Anscratch glass
2. Display TFT 1,77”
3. Funcon
4. TEST
5. subfeatures selecon
6. subfeatures selecon
7. SCHUKO plug
This feature measure Voltage between phase and neutral and test the socket wiring
in according to local requirement.
Operaon
Plug the instruments in the socket and select Voltage
funcon with buon.Voltage between Phase and
Neutral will be shown at display if socket is correctly
wired and voltage contact is <50V (Fig 1) error due
to the uncorrect wiring is shown(Fig 2 e 3). Green bar
indicate every parameter is correct. Frequency can be
shown through buon .
Wiring Error ( bar is Red and
sound allarm)
• Phase - Neutral Reverse
• Phase - Ground Reverse
• Open Neutral
• No Ground
• Voltage contact >50V
Soluon
Adjust the socket wiring
Adjust the socket wiring
Connect Neutral wire
Check Earth connecon
Dangerous Voltage on mass

8
RCD (RESIDUAL CURRENT DEVICE)
Fig. 5
Fig. 4
This feature allow test of RCD according EC/EN 61557-6 with tripping me and
Voltage contact. Test Current of 30 mA is injected through the ground for type A ,
AC and F.
Test Mode:
• x1 I∆n –Test with 30mA
• x5 I∆n –Test with 150mA
• AUTO –Test sequence x1/2 - x1/2 180° - x1 - x1 180° - x5 - x5 180°
RCD x1 , x5
Connect RapidPro to the socket and press FUNC in order to select RCD funcon.
Press the buon for select the type of RCD and buon for select the current
(x1,x5) Press RED buon for 3 second for start the measure.
Tripping Time is shown on the display and Gre-
en bar means the test is performed correctly
Vc is the contact Voltage.
If Contact Voltage >50V the RCD tripping test
will be stopped and the bar became RED
Display shown tripping me is higher than the
limit standard and Red Bar means outcome
NOT OK (Fig 5).
Vc is the Contact Voltage.

9
RCD (RESIDUAL CURRENT DEVICE)
STEP 1
STEP 2
STEP 3
STEP 4
STEP 5
STEP 6
I∆n x1 0°
I∆n x1 180°
I∆n x5 0°
I∆n x5 180°
I∆n x1/2 0°
I∆n x1/2 180°
OK IF <300 ms
OK IF <300 ms
OK IF <40 ms
OK IF <40 ms
OK IF >1000 ms
OK IF >1000 ms
Fig. 7
Fig. 6
AUTO funcon
Plug the instrument in the socket and press FUNC for select RCD funcon. Press
arrow buon for select type of RCD and buon for select funcon AUTO .
Start measuring by pressing the Red buon,The instruments will execute all test in
sequence when you re-arm the RCD and voltage is on.
AUTO mode consist of 6 test in the folowing sequence:
Display Show RCD tripping me in ms. For every
STEPtripping me in ms will be shown. GREEN (Fig
6) if the test and tripping me are conform to local
tripping me regulamentaon.
Value in RED (Fig 7) means the test is not conform
to the local guideline and the test will be stopped.

10
LOOP
Fig. 8 Fig. 9
The measurament is performed according IEC/EN61557-3 it measure impedance
between Phase and Ground withouth RCD trip. Test Mode:
• TT Measure Earth Resistance (Ω) and Contact Voltage (V) Fig.8
• TN Measure Impedance Z L-Pe (Ω) and Ipsc prospecve short circuit current
(A) Fig.9
• Std Test with 15mA nominal current
• Low Test with 6mA. Use this setup if during the test the RCD trip due to the
presence of leackage current in the electric system under test.
LOOP FUNCTION
Plug Rapid PRO in the socket and press FUNC buon for select LOOP. Press buon
for select TT or TN and press buon for select TEST STD or LOW.
Press the RED buon for 2 second to start the measurement
During the LOOP measurement, if the SUPER CAP is not
fully charged, the Display may switch o or decrease the
intensity, however the measurement will be performed and
shown at the end.
Other manuals for RAPID PRO
1
Table of contents
Other Uniks Test Equipment manuals