4.Place the meter on a flat
surface, like a table.
5. Remove the control solution
bottle cap and wipe the tip of the
bottle with a tissue.
6.Squeeze the bottle until a tiny
drop forms at the tip of the
bottle.
Be aware that there are Normal and High ranges listed on
your test strip bottle. The Normal and High refer to the three
different control solutions that are available. Please be sure
that you are looking at the correct range.
7.Touch the control solution drop
to the tip of the transparent
window of the test strip. Do not
put control solution drop on
top of the strip.
8.The meter starts to count down
from 5 seconds and then
shows the test result.
9. Do not remove the test strip yet.
Check if the reading falls within the
range printed on the test strip bottle.
10.Push the Strip Ejector to remove
the test strip.
11. Throw the test strip away after
you have compared the reading
to the range printed on the test
strip bottle.