XRB80N100 MONOBLOCK® MANUAL 3 118090-001 REV D
1.2 Operating Features
The MONOBLOCKSeries incorporates several
standard features designed to optimize user satisfaction
and safety.
INTERNAL FAULT PROTECTION: The
MONOBLOCKSeries continually monitors
internal circuits critical to the proper operation of the
power supply. In the event that one of these circuits
does not function correctly, the fault detection circuit
latches the appropriate fault on the front cover display
and turns off the output.
LED FAULT/STATUS INDICATORS:
Indicators are located on the cover of
Inverter/Control Assembly and are provided to give
the user complete indication of system operation and
fault conditions.
POWER OVERVOLTAGE (OV)
ARC FLT UNDERVOLTAGE (UV)
OVERTEMP (OT) OVER CURRENT (OC)
UNDER CURRENT (UC) X-RAY ON
WATCHDOG OPERATION: The HVPS employs
an internal watchdog timer to discontinue X-ray
operation in instances where the host computer has
lost communication and control of the HVPS system
for a period greater than ten second. This feature is
enabled via RS-232 host command.
LOCAL PROGRAMMING: Allows Local
adjustment of the output voltage and current via
analog voltage inputs on J2 on the front cover. To
operate in Local mode, position jumpers for JP4 and
JP5 in the 2-3 position as per table on the front cover.
LOCAL MONITORS: provides Local analog
monitors for the output voltage and current via analog
voltage outputs on J2 on the front cover.
REMOTE PROGRAMMING: Allows remote
adjustment of the output voltage and current via RS-
232 digital interface port at J3 on the front cover. To
operate in Remote mode, position jumpers for JP4
and JP5 in the 1-2 position as per table on the front
cover. All program values default to zero upon
power up.
REMOTE MONITORS: Provides remote monitors
of the output voltage, output current and Filament
current via RS-232 digital interface port at J3 on the
front cover.
X-RAY ENABLE CONTROL: Provides control of
X-ray ON and X-ray OFF either via a dry contact
connection from J2-15 to J2-13 when operating in
Local Mode (analog control) or Remote Mode RS-
232 communication when operating in remote mode.
Upon power up the unit will default to Local Mode
and will require re-assertion of J2-15 to enable unit.
After an X-ray Enable command is received via RS-
232 the unit will switch to Remote Mode.
EXTERNAL INTERLOCK: The X-ray cannot be
enabled unless the external interlock is closed by
connecting J2-11 to J2-12. During high voltage
operation, opening the interlock circuit will cause the
High Voltage to be disabled. Interlock status
indication via RS-232 interface at J3.
IMPORTANT
This control signal is not a safety
interlock and should not be used for
protection from X-ray generation for
safety purposes.
HV ON LAMP RELAY: When X-ray ON command
is enabled a NO and a NC relay contact closure will
occur at connector J2. Refer to pin assignments
datasheet SHV#128065-001 and specification for
contact ratings.
OIL TEMPERATURE MONITOR: Provides
remote oil temperature monitor via RS-232 digital
interface port at J3 on front cover.
PS Fault Indication: PS Faults an open collector
output on J2-1, indicates that a faults has occurred.
High = no faults.
X-ray On Indication: X-ray On Signal is an open
collector output on J2-14, indicates that a faults has
occurred. High = X-ray off