SCIEX MSR 130 Manual

SCIEX Dry Pump MSR 130
OPERATOR GUIDE
RUO-IDV-05-14943-A Original instructions

Disclaimer
This document is provided to customers who have purchased SCIEX equipment to use in
the operation of such SCIEX equipment. This document is copyright protected and any
reproduction of this document or any part of this document is strictly prohibited, except as
SCIEX may authorize in writing.
Software that may be described in this document is furnished under a license agreement. It
is against the law to copy, modify, or distribute the software on any medium, except as
specifically allowed in the license agreement. Furthermore, the license agreement may
prohibit the software from being disassembled, reverse engineered, or decompiled for any
purpose. Warranties are as stated therein.
Portions of this document may make reference to other manufacturers and/or their products,
which may contain parts whose names are registered as trademarks and/or function as
trademarks of their respective owners. Any such use is intended only to designate those
manufacturers' products as supplied by SCIEX for incorporation into its equipment and does
not imply any right and/or license to use or permit others to use such manufacturers' and/or
their product names as trademarks.
SCIEX warranties are limited to those express warranties provided at the time of sale or
license of its products and are the sole and exclusive representations, warranties, and
obligations of SCIEX. SCIEX makes no other warranty of any kind whatsoever, expressed or
implied, including without limitation, warranties of merchantability or fitness for a particular
purpose, whether arising from a statute or otherwise in law or from a course of dealing or
usage of trade, all of which are expressly disclaimed, and assumes no responsibility or
contingent liability, including indirect or consequential damages, for any use by the
purchaser or for any adverse circumstances arising therefrom.
Trademarks and/or registered trademarks mentioned herein, including associated logos, are
the property of AB Sciex Pte. Ltd., or their respective owners, in the United States and/or
certain other countries (see sciex.com/trademarks).
For Research Use Only. Not for use in Diagnostic Procedures
AB Sciex™ is being used under license.
Made in Czechia
Manufactured for:
AB Sciex LLC
500 Old Connecticut Path
Framingham, Massachusetts 01701
USA
Copyright notice
© 2023 DH Tech. Dev. Pte. Ltd.
Published: 9/4/2023

Contents
1. Safety and compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1. Definition of Warnings and Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2. Safety symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. General description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1. Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2. Interfaces and control panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.1. Auxiliary connector socket. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.2. Logic interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.3. Auto-run. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.2.4. Pump controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3. Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1. Operating and storage conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2. Performance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.3. Materials exposed to gases. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4. Intake gas specification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.5. Electrical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1. Installation safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2. Unpack and inspect. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4.3. Mechanical installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.4. Leak test the system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.5. Electrical installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.5.1. Connect the electrical supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.6. Connect the system. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.6.1. Connection for remote control and monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5. Commission the pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
6. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.1. Operation safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.2. Operational modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.2.1. Manual control mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.3. Logic interface data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.4. Parallel control and monitoring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.5. Analogue speed control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
6.6. Hardware configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.7. Operational modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.8. Start the pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.9. Shut down the pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
6.10. Restart the pump. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.11. Unplanned shutdown. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.12. Gas ballast operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
6.13. Gas ballast connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
SCIEX Dry Pump MSR 130
Page 3 of 40
Operator Guide
RUO-IDV-05-14943-A

7. Maintenance and service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.1. General maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.2. Maintenance plan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.3. Inspect and clean the inlet strainer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.4. Clean the ventilation slots. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.5. Electrical safety check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.6. Return the equipment or components for service (SCIEX). . . . . . . . . . . . . . . . . . . . . . . . . . 35
8. Fault finding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
8.1. Alarm indicator codes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9. Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
10. Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Operator Guide
RUO-IDV-05-14943-A
SCIEX Dry Pump MSR 130
Page 4 of 40

List of Figures
Figure 1: Pump features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Figure 2: Control panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Figure 3: Valve connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Figure 4: Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Figure 5: Performance graph. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Figure 6: Unpacking Instruction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Figure 7: Logic interface connections - parallel control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Figure 8: Logic interface connections - analogue speed control. . . . . . . . . . . . . . . . . . . . . . . . 29
Figure 9: Analogue speed control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Figure 10: Maximum vapour handling capacity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Figure 11: Gas ballast connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
SCIEX Dry Pump MSR 130
Page 5 of 40
Operator Guide
RUO-IDV-05-14943-A

1. Safety and compliance
For safe operation from the start, read these instructions carefully before you install or
commission the equipment and keep them safe for future use. Read all the safety instructions
in this section and the rest of this manual carefully and make sure that you obey these
instructions.
The instruction manual is an important safety document that we often deliver digitally. It is
your responsibility to keep the instruction manual available and visible while working with the
equipment. Please download the digital version of the instruction manual for use on your
device or print it if a device will not be available.
1.1. Definition of Warnings and Cautions
For safe operation from the start, read these instructions carefully before you install or
commission the equipment and keep them safe for future use.
Read all the safety instructions in this section and the rest of this manual carefully and make
sure that you obey these instructions. The equipment must only be operated and maintained
by trained personnel in the proper condition and as described in this instruction manual.
Obey local and state requirements and regulations. If you have any questions about safety,
operation or maintenance of the device, please contact our nearest subsidiary.
Important safety information is highlighted as warning and caution instructions which are
defined as follows. Different symbols are used according to the type of hazard.
WARNING:
If you do not obey a warning, there is a risk of injury or death.
CAUTION:
If you do not obey a caution, there is a risk of damage to equipment, related equipment or
process.
NOTICE:
Information about properties or instructions for an action which, if ignored, will cause
damage to the equipment.
We reserve the right to change the design and the stated data. The illustrations are not
binding.
Operator Guide
RUO-IDV-05-14943-A
SCIEX Dry Pump MSR 130
Page 6 of 40

1.2. Safety symbols
The safety symbols on the products show the areas where care and attention is necessary.
The safety symbols that we use on the product or in the product documentation have the
following meanings:
Warning/Caution
Risk of injury and/or damage to equipment. An appropriate safety instruction
must be followed or a potential hazard exists.
Warning - Heavy object
Risk of injury or damage to equipment. Identifies a possible hazard from a
heavy object.
Warning - Dangerous voltage
Risk of injury. Identifies possible sources of hazardous electrical shock.
Warning - Hot surfaces
Risk of injury. Identifies a surface capable of inflicting burns through contact.
Symbol - Protective earth
Identifies an electrical equipment earth (ground) terminal.
Warning - Use protective equipment
Risk of injury. Use appropriate Personal Protective Equipment (PPE) when
performing the task.
SCIEX Dry Pump MSR 130
Page 7 of 40
Operator Guide
RUO-IDV-05-14943-A

2. General description
2.1. Overview
WARNING: INCORRECT USE OF EQUIPMENT
Risk of injury or damage to equipment. Incorrect use of the equipment can cause injury. The
user is responsible for the safe operation, installation and monitoring of the system.
CAUTION: GAS BALLAST INSTALLATION
Risk of damage to the equipment. If water vapour is present within the system or is pumped
by the product the gas ballast must be used to prevent damage to the product. The gas
ballast should continue to be applied with the pump running, for at least 1 hour after the
water vapour source has been removed to enable the pump to dry.
The MSR 130 is a dry, multi-stage roots vacuum pump that offers high pumping speed in a
compact form. The pump is connected by a power cord and is designed for use on clean duty
applications.
The cord rating is:
Supply voltage (V) Current (A) Frequency (Hz)
200 ‑ 240 16 50 or 60
The pump is not designed for use with flammable, corrosive, toxic or other hazardous gases.
Gas or oxygen can mix in the pump system.
The dry pump operates in steady-state gas load conditions with an inlet pressure from
1 ‑ 15 mbar. If the inlet pressure is more than 15 mbar for an extended period, the pump
controller reduces the motor speed. The motor speed is increased when the pressure is
reduced, or when the transient overload protection has recovered. Refer to Table:
Performance data for information on maximum inlet pressure.
Refer to Figure: Pump features. The system is supplied with NW40 inlet port (item 3) and
NW25 exhaust port (item 2).
The system is air cooled by a fan installed in the pump enclosure. The fan continues to
operate for 5 minutes after the pump is set to off. The pump has a thermal protection device
that will stop the motor if a thermal overload, for example in high ambient temperature,
occurs. You must restart the pump after it has cooled down.
The pump mechanism is operated by an electric motor driven by an internal pump controller.
The rotational elements of the pumping mechanism are simply supported at the end by
lubricated bearings. PFPE lubricant is used and this is contained in both the gearbox and
motor ends of the pump module. These are "sealed for life" and the lubricant does not need
to be replaced until the pump service interval is reached.
The system has a gas ballast connection, if it is necessary to purge the system.
The pump is designed to remove the residual gases used in high vacuum systems. The
gases are:
▪Air
▪Oxygen (O2) < 21% by volume
▪Neon
▪Nitrogen
▪Argon
Operator Guide
RUO-IDV-05-14943-A
SCIEX Dry Pump MSR 130
Page 8 of 40

▪Helium
You can use the pump to pump water vapour. You must use the gas ballast when water
vapour is pumped. The water vapour must not condense in the pump.
If you wish to pump a gas not contained within the list above, or if you are unsure about safe
gas mixtures or concentration levels, contact your supplier for advice. If you do not contact
your supplier, you may invalidate the warranty on the pump. The pump is not suitable for
pumping aggressive, corrosive or mixture of flammable gases.
Figure 1 Pump features
1. Dashboard or interface 2. Exhaust port
3. Inlet port 4. Castor x 4
5. Leveling feet x 4 6. Gas ballast connection
7. Lifting eye positions 8. Mains circuit breaker
9. Mains connector port 10. Protective earth stud
11. 15-way D-type connector 12. Auxiliary connector
1. Dashboard or interface 2. Exhaust port
3. Inlet port 4. Castor x 4
5. Leveling feet x 4 6. Gas ballast connection
7. Lifting eye positions 8. Mains circuit breaker
9. Mains connector port 10. Protective earth stud
11. 15-way D-type connector 12. Auxiliary connector
SCIEX Dry Pump MSR 130
Page 9 of 40
Operator Guide
RUO-IDV-05-14943-A

2.2. Interfaces and control panel
Figure 2 Control panel
1. Start button 2. Standby button
3. Standby increase button 4. Standby decrease button
5. Service indicator LED 6. Alarm indicator LED
7. Auto‑run indicator LED 8. Standby decrease indicator LED
9. Standby increase indicator LED 10. Standby indicator LED
11. Stop button 12. Run indicator LED
13. Power status LED
1. Start button 2. Standby button
3. Standby increase button 4. Standby decrease button
5. Service indicator LED 6. Alarm indicator LED
7. Auto‑run indicator LED 8. Standby decrease indicator LED
9. Standby increase indicator LED 10. Standby indicator LED
11. Stop button 12. Run indicator LED
13. Power status LED
2.2.1. Auxiliary connector socket
An auxiliary control connection on the rear panel controls an optional inlet valve. This inlet
valve can be operated in parallel with the normal pump output signal. Refer to Figure: Pump
features item 6. The valve is usually closed and:
▪will open when the normal signal is active (pump at speed)
▪will close when you select stop, or if there is a fault condition.
The reaction time will be in line with the valve selection. The output signal is 24 V d.c. Refer
to Figure: Valve connector for the polarity of the connector pins when the connector is
energized.
The auxiliary connector is regulated to 24 V d.c. to control the accessories. Refer to Table:
Auxiliary load currents, the output will shut down to protect the pump controller.
Refer to Table: Recommended mating plugs for the recommended mating plugs.
Operator Guide
RUO-IDV-05-14943-A
SCIEX Dry Pump MSR 130
Page 10 of 40
Table of contents
Other SCIEX Water Pump manuals
Popular Water Pump manuals by other brands

Sykes AmeriPumps
Sykes AmeriPumps GP100M Operation and maintenance instructions

DUROMAX
DUROMAX XP WX Series user manual

BRINKMANN PUMPS
BRINKMANN PUMPS SBF550 operating instructions

Franklin Electric
Franklin Electric IPS Installation & operation manual

Xylem
Xylem e-1532 Series instruction manual

Milton Roy
Milton Roy PRIMEROYAL instruction manual