Control Unit Specifications
Rated power source 9v DC
Format Battery
Max current consumption 44mA
Dimensions (L x W x H) 12.55cm x 7cm x 2.4cm
4.94 in x 2.75 in x 0.94 in
Weight (without battery) 116g
Before you start
•When storing or using this device, do not allow it
to encounter liquids or intense heat. This may
damage the unit and its accessories. Smoke and
overheating could result from contact with liquids.
•This control unit and sensor pads are not single
use. The insoles are designed for single use.
This Essential Load Unit is for commercial use
only.
Indications and Intended Use
The Essential Load Unit gives immediate audio visual
feedback on partial weight bearing for clients,
including those with an Acquired Brain Injury or
following Orthopedic Surgery
Essential Load Unit
Includes the following items:
Carry case with foam inserts
Operation Manual
Essential Load Control Unit & Belt Clip
Six sizes of insoles for taking sensors
Two sensor pads. Heel and Forefoot
Headphone
Curly Cable
Double Adaptor.
Warranty
•This unit comes with a 12-month warranty on parts
from date of purchase. See our website for more
details.
•Wherever possible a replacement unit will be
provided while your unit is being repaired
Returns / Repairs Policy
For all returns or repairs you must have a return
number given by nCounters attached to it.
The item should be sent to:
nCounters
PO BOX 2336
KEW VICTORIA 3101
AUSTRALIA
3
3
3
3
3
3
4
4
4
4
5
5
6
6

4
Care of the Control Unit
Installing the battery
•Remove the back panel of the enclosure and
insert the battery with the plus (+) and minus (-)
poles pointing in the designated direction.
•Use one 9 Volt (PP3) battery. A rechargeable 9V
battery maybe used.
•Remove the battery if the control unit will not be
used for a month or longer.
•If the battery is incorrectly replaced, the control
unit may not function.
•In the event of battery leakage, wipe the control
unit completely clean and install a new battery.
Cleaning
•All surfaces of this system except the insoles
are non-porous and should be cleaned between
each client, using a disinfectant wipe
The Control Unit
What’s what?
Figure 1
Control Unit Front Panel
1. ON / OFF switch
2. INPUT socket
For connecting the two sensor pads via curly
cord and adaptor
3. Headphone Socket
4. Volume Control
5. Low Battery LED
Battery needs to be immediately replaced or
recharged when red LED comes on.
6. Speaker
7. 10 LED BAR DISPLAY
8. UPPER LIMIT
Dial to be used by therapist to establish upper
limit of target range -See FAQs to see how this
dial can be used for getting feedback when
target is reached
9.LOWER LIMIT
Dial to be used by therapist to establish lower
limit of target range.
8
9
0
2
3
4
5
6
1
7
3 RED
LED bars
represent
LOWER
values
7 GREEN
LED bars
represent
HIGHER
values

5
Using the Essential Load Unit
Before turning on the control unit:
Choose the correct size insole for the user. Left or
Right. Your unit comes with six sizes of insoles.
These should become the user’s to keep. You can
order more insoles via our website, or by contacting
Replacement or extra insoles are available in pairs
from size 35 –47 (Euro –Sizing). All insoles are
designed for taking the two sensor pads.
Take the heel and forefoot sensors.
.
Attach them on the Velcro on the heel and forefoot
underneath your insole.
Remove existing insole.
Put nCounters insole with sensors inside a runner
(sneaker).
User places foot into shoe. Plug each of the two
cables to the double adaptor, connected to the curly
cable.
Connect the plug to the input socket
of the Control Unit.
Ask user to press up and down on insole for 10
seconds to condition the sensors.
Ask user to place foot on set of scales to get weight
above which the user is to receive no feedback.

6
Turn on control unit using the side toggle switch. One
of the green or red LEDs will come on.
Set Target Range
The therapist needs to set the correct upper and
lower limits. Make sure the two dials are in the
following default positions:
•Upper limit dial fully clockwise (maximum
position)
•Lower limit dial fully anti-clockwise (minimum
position).
Firstly, program the lower limit. Turn this Lower Limit
dial clockwise until you hear a sound from the
speaker simultaneously the target weight is being put
through the insole.
Now get the upper limit by turning this dial
anticlockwise, until you hear a sound from the
speaker while weight is being put through the insole.
Finesse this dial a little (by turning clockwise) to
create your window.
In summary, when the client’s partial weight on the
insole is between the lower and upper limit values,
the client will hear no sound. If the weight shift is
outside these two limits, then the client will receive
feedback. The sound changes and increases in pitch
the further away the client gets from the target range.
Before using the headphones turn the volume control
back to minimum then slowly increase until the sound
is comfortable for the user.
FAQs
Q. Can I just use the heel sensor on its own?
A. Yes, If you are just using one sensor you will not
need the adaptor.
Q. Can the unit be set so user gets feedback only
when they reach the pre-set target?
A. Yes, turn lower limit dial fully anti clockwise and
adjust upper limit dial until a sound is generated
when target weight is being put through the insole.