Medicatech MasterRad30 User manual

Page 1of 15
Rev. 01
Installation Manual
MasterRad30
Universal Radiographic System

Page 2of 15
Rev. 01
Introduction:
Important! .. X-ray Protection
X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS
MEASURES OF PROTECTION ARE STRICTLY OBSERVED
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
should be thorougly read and understood before attempting to place this equipment in
operation. We will be glad to assist and cooperate in placing this equipment
Although this apparatus I built to the highest safety and incorporated a high degree of
protection against x-radiation other than the useful beam, no practical precautions to prevent
the possibility of any persons carelessly, unwisely, or unknowingly exposing themselves or
other to x-radiation.
It is important that everyone working with x-radiation be properly trained and take adequate
steps to insure protection against injury. The manufacturer assumes that all operator and
service personnel authorized to use, install, calibrate an maintain this equipment is cognizant of
the danger of excessive exposure tp x-radation, is sufficiently trained and has the required
knowledges for it. The equipment herein described which may result from exposure to x-
radiation.
Various protective material and devices are availablt. It is recommended that such materials
and devices be used

Page 3of 15
Rev. 01
SECTION 1: INSTALLATION
1.1 TOOLS
The following hand tools are required for the installation:
•Standard service engineers tool kit.
•Electric drill motor and assorted bits.
1.2 Pre-installation checks
Prior to beginning installation, it is recommended to inspect the sire and verify that
the x-ray room complies with requirement such as:
•Space requirements to allow installation and system movements must
consider the maximum dimensions and travels of the equipment.
Maximum Height 96 in (8 ft)
Maximum Width 132 in (11 ft)
Maximum Length 180 in (15 ft)
•Conduits and walls are ready to install the system.
•Electricity installation:
oMain supply: Single phase, 50/60 Hz
115/208/230/240 V ~
•Minimum Power input required: 400 VA

Page 4of 15
Rev. 01

Page 5of 15
Rev. 01
1. Hardware Installation:
1.1. Refer to the layout of the room
1.2. Locate and unscrew the 10 x allen flat head screws #10-32 screws
Figure 1:take the bare strips and the floor mount cover
1.3. Take off the 2 brake strips and the floor mount cover
1.4. Place the floor mount 6” from the back wall.
1.5. Mark the place of the anchor on the floor.
5x Allen flat head #10-32 screws
5x Allen flat head #10-32 screws
Brake strips
Floor
mount
cover

Page 6of 15
Rev. 01
1.6. Using a 3/8” cement drill bit, Drill 7 mounting holes (refer to figure 2)
1.7. Place the washer and the nut on the anchor, as it will be hard to place them after
hammering the anchors in the floor.
1.8. Use the hammer to fix the anchor into the ground. (Make sure that the anchors are
deep enough in the floor, so it will clear the cover on top. )
1.9. Place the top cover and the 2 brake strips using the provided screws. (refer to figure 1)
1.10. Place the tube stand on the floor mount. (figure 3) (requires 2 people minimum)
Figure 2: floor mount mounting holes

Page 7of 15
Rev. 01
Figure 3:place the tube stand on the floor mount.
1.11. Align the tube mount base with the 4 carts on the floor mount.
1.12. Use the 16 xM6 screws to connect the tube stand to the floor mount. (figure 4)
Figure 4: 16x M6 screws

Page 8of 15
Rev. 01
1.13. Place the 4 side covers and the 4 corners simply by connecting them and place
them around the floor mount (refer to figure 5)
1.14. Make sure that the side cover with the cable tray attached is at the back side.
1.15. Connect grey draw wire sensor at the back right of the floor mount at its
designated place. (use the three screws provided)
1.16. Place the sensor’s cover by inserting the cover’s lip into the floor mount cover’s
grove.
1.17. Connect the wire to the cart.
1.18. Guide the wire through the cable tray and to the relay at the back of the tube
stand.
1.19. Locate black draw wire sensor at the bottom of the tube stand.
1.20. Attached the wire to the bottom of the arm holder.
1.21. Mount the control panel through the 4 provided screws on the front of the arm.
1.22. Attach the bracket provided to the top of the tube stand.
1.23. Mount the front cladding (extrusion) to on the bracket.
1.24. Repeat 1.22 and 1.23 for the chest stand.
1.25. Attach the plastic part at the end of the cable tray to the cart (use the designated
supplied screws)
Side covers
Figure 5: showing the top cover, brake strips, side cover and corner.
Corners

Page 9of 15
Rev. 01
Tube , collimator, and generator installation
(all cables are going to the back of the tube stand and through the cable tray)
1.26. Place the tube on top of the arm.
1.27. Place the collimator ring at the bottom of the arm below the tube through the
arm
1.28. Place the collimator below the arm as shown in the picture, then screw the
collimator inside screws to fix the collimator to the collimator ring.
1.28.1. (refer to the collimator Manual)
Refer to the following pages for proper cable management and connections for the tube,
collimator and generator
2. Collimator, Tube, and generator installation:
2.1. Collimator: (refer to the collimator manual)
2.1.1. Align the tube using the screws found with the
metal plates inside of the blue binder. (note: there
are two different types of screws used according to
the type of X-ray tube).
2.1.2. Thereare two main tools used to install
collimator;
2.1.2.1. Allenkey for screws
2.1.2.2. Allenkey for the mounting plate

Page 10 of 15
Rev. 01
2.1.3. remove the collimator mounting plate
2.1.4. connect the collimator mounting plate to the tube using the screws previously
used to align and keep tube in place
2.1.5. place collimator on to mounting plate and tighten all four screws in order to
secure collimator on the tube arm
2.1.6. connect collimator to CPI generator cable then other collimator cable located at
the back of the table to foot pedal cable ( there is only one way to connect the
cables)
Figure 8: showing the 4 mounting holes at the bottom of the arm to mount the tube and the collimator ring
2.2. Tube: (refer to the generator and the tube manual)
2.2.1. Place tube on holder so that it is stable
2.2.2. Unscrew the tap (be gentle with cap screws as they are easy to strip)
2.2.3. Grey tube wire is connected to the tube with the wire side labeled 1,2,3, etc.
2.2.4. Connect the grey wire as follow inside the tube to its relative number: (refer to
figure 9)
2.2.4.1. Black wire is 1
2.2.4.2. Red wire is 2
2.2.4.3. Yellow wire is 3
2.2.4.4. White wire is 5
2.2.4.5. Blue wire is 6
2.2.5. Get the wire through the built in zip tie in order to adjust the fix the wgrey wire
in place inside the tube