4545 : 28-09-2012
30 31
28 29
AC Apparent Power (VA, from 0.000VA to 9999 KVA)
Range
Resolution
Accuracy of Readings2
>20V & >20A <20V or <20A
10.0 - 999.9 VA 0.1VA ±2% ± 20dgts ±2% ± 40dgts
1.000 - 9.999 KVA
0.001 KVA
±2% ± 20dgts ±2% ± 40dgts
10.00 - 99.99 KVA 0.01 KVA ±2% ± 20dgts ±2% ± 40dgts
100.0 - 999.9 KVA 0.1 KVA ±2% ± 20dgts ±2% ± 40dgts
1000 - 9999 KVA 1 KVA ±2% ± 20dgts ±2% ± 40dgts
2For CTf1, the accuracy in percentage is the same (±2%).
But the additional VA should be multiplied by the CT
ratio.
For example, ±2.0VA becomes ±2.0VA * CT ratio
±4.0VA becomes ±4.0VA * CT ratio
AC Reactive Power (VAR, from 0.000 VAR to 9999 KVAR)
Range
Resolution
Accuracy of Readings3
>20V & >20A <20V or <20A
10.0 - 999.9 VAR 0.1VAR ±2% ± 30dgts ±3% ± 40dgts
1.000 - 9.999 KVAR
0.001 KVAR
±2% ± 30dgts ±3% ± 40dgts
10.00 - 99.99 KVAR
0.01 KVAR
±2% ± 30dgts ±3% ± 40dgts
100.0 - 999.9 KVAR 0.1 KVAR ±2% ± 30dgts ±3% ± 40dgts
1000 - 9999 KVAR 1 KVAR ±2% ± 30dgts ±3% ± 40dgts
3For CTf1, the accuracy in percentage is the same (±2%).
But the additional VAR should be multiplied by the CT
ratio.
For example, ±3.0VAR becomes ±3.0VAR * CT ratio
±4.0VAR becomes ±4.0VAR * CT ratio
Range of CT (Current Transformer) Ratio : 1 to 250
H.P. (Horse Power)
1 H.P. = 746 W
AC Active Energy ( WH, or KWH, from 0 WH to 999,999
KWH) WH = W * Time (in hours)
AC Current (50 or 60 Hz, Auto Range, True RMS, Crest
Factor < 4, CT=1)
(Overload Protection AC 2000A)
Range Resolution Accuracy of Readings4
10.0 - 1500.0 A 0.01 A ± 2% ± 5dgts
4For CT =1, the accuracy in percentage is the same (±2%).
But the additional digits should be multiplied by the CT
ratio.
For example, ±5 digits becomes ±5 digits * CT ratio
AC Voltage (50 or 60 Hz, Auto Range, True RMS, Crest Factor
< 4, Input Impedance 10 MV, Overload Protection AC 800V)
Range Resolution Accuracy of Readings
10.0 V - 600.0 V 0.1 V ± 0.5% ± 5dgts
Harmonics of AC Voltage in Percentage (1 to 99 order,
minimum voltage at the 50 or 60 Hz > AC 80V. If the voltage is
0 at 50 or 60 Hz, all the percentage (%) display is 0.)
Range Resolution Accuracy
1 - 20th 0.1% ±2%
21 - 49th 0.1% 4%of reading ± 2.0%
50 - 99th 0.1% 6%of reading ± 2.0%
Harmonics of AC Voltage in Magnitude (1 to 99th order,
minimum voltage at the 50 or 60 Hz > AC 80V)
Range Resolution Accuracy
1 - 20th 0.1% ± 2.0% ± 0.5V
21 - 49th 0.1% 4%of reading ± 0.5V
50 - 99th 0.1% 6%of reading ± 0.5V
Harmonics of AC Current in Percentage (1 to 99th order,
minimum current at the 50 or 60 Hz o 20 A. If the current is 0
at 50 or 60 Hz, all the percentage (%) display is 0.)
Range Resolution Accuracy
1 - 20th 0.1% ±2%
21 - 49th 0.1% 4% of reading ± 2.0%
50 - 99th 0.1% 6% of reading ± 2.0%
Harmonics of AC Current in Magnitude (1 to 99th order,
minimum current at the 50 or 60 Hz > 20A)
Range Resolution Accuracy
1 - 20th 0.1% ± 2% of reading ± 0.4A
21 - 49th 0.1% ± 4% of reading ± 0.4A
50 - 99th 0.1% ± 6% of reading ± 0.4A
Power Factor (PF)
Range Resolution Accuracy
>20V & >20A <20V or <20A
0.000 - 1.000 0.001 ± 0.04 ± 0.1
Phase Angle (u)
Range Resolution Accuracy
-180 to 18000.10±10
0 to 36000.10±10
Total Harmonic Distortion (THD-F with respect to the 50 or 60
Hz, min. value at the 50 or 60 Hz > 80V and > 20 A, 1 to 50
Harmonics. If the voltage or current is 0 at 50 or 60 Hz, all the
percentage (%) dispiay is 0)
Range Resolution Accuracy
0.0 - 20% 0.1% ±2.0%
20 - 100% 0.1% ± 6% of reading ±1%
100 - 999.9% 0.1% ±10% of reading ±1%
Peak Value of AC Voltage (peak value > 5V) or AC Current
(peak value > 20A)
Range Sampling Time Accuracy of Reading
50 Hz 39 ms ± 5% ± 30 digits
60 Hz 33 ms ± 5% ± 30 digits
Crest Factor (C.F.) of ACV (peak value > 5V) or ACA (peak
value > 20A)
Range Resolution Accuracy of Reading
1.00 - 99.99 0.01 ± 5% ± 30 digits
Resistance (V) and Continuity (Beep if less than 50V)
Range Resolution Accuracy
7.0 - 999.9 V0.1 V± 5 V
1000 - 1200 V1 V± 5 V
If reading is less than 7V, it is displayed as 0V
Frequency (RMS Value >10V) or ACA (RMS value >30A)
Range Resolution Accuracy of Readings
45 - 65 0.1 ± 0.2 Hz