Martel LC-110 User manual

LC-110/H
Reference Manual
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

1. Introduction ........................................1
1.1 Customer Service ................................................1
1.2 Standard Equipment ..............................................2
1.3 Safety information ................................................2
2. Calibrator Interface and Operation .....................5
2.1 Milliamp Source .................................................6
2.2 Milliamp Simulate.................................................7
2.3 Milliamp Measure without 24V power .................................8
2.4 Milliamp Measure with 24V power ...................................9
2.5 Volts Measure ...................................................9
3. Main Menu ........................................10
4. Calibrator Setup Menu...............................11
4.1 Setting Auto Ramp Time ..........................................11
4.2 Setting Auto Step Time ...........................................12
4.3 Enabling Valve Test ..............................................12
4.4 Enabling the HART Resistor........................................13
4.5 Selecting mA Span ..............................................13
4.6 Contrast Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.7 Configuring Auto Shutdown .......................................14
4.8 HART Write Enable...............................................14
5. Advanced Features..................................15
5.1 Step and Ramp Operation.........................................15
5.2 Valve Test ......................................................16
5.3 HART 250ΩResistor .............................................16
6. HART Device Communications ........................17
6.1. HART Connection Diagrams.......................................17
6.2 Communications Setup and Selection ...............................19
6.3 Connecting to a HART Device......................................21
6.4 Disconnect.....................................................23
6.5 Function Select Menu ............................................23
6.6 Display Setup & Data.............................................23
6.7 Modify Setup ...................................................24
6.8 Trim, Set, & Zero Menu ...........................................28
6.9 Device Diagnostic ...............................................32
6.10 Data Log & Config Log ..........................................33
6.11 HART Modem Simulation.........................................38
6.12 Error Displays..................................................39
7. Maintenance .......................................45
7.1 Replacing Batteries ..............................................45
7.2 Cleaning the Unit ................................................45
7.3 Fuse Protection .................................................45
8. Specifications .....................................46
9. Warranty ..........................................46
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

1
Introduction
The Martel LC-110 and LC-110H are mA (loop) calibrators designed
to be simple to use and to offer the highest accuracy of any loop
calibrator in their class. The LC-110H differs from the standard
LC-110 in that it incorporates HART communications and supports a
select subset of the HART universal and common practice
commands. This unique feature allows the LC-110H to be used as
both a loop calibrator and basic function communicator.
The calibrator supports the following functions:
•Currentmeasurementandsourcing,includingselectable24V
supply
•Voltagemeasurement
•Valvetestcapability
•SelectableHART250Ωloop resistor
•Outputstepandramp
The calibrator has the following features:
•Largeeasytoreaddisplayformeasurementsanddataentry
•Digitalknobwithselectabledecadecontrolforquickandeasy
data entry
•Interactivemenus
•Inputandoutputjacksprotectedbyselfresettingfuses
providingprotectionto240VAC
Inthecommunicatormodetheuserisabletoreadbasicdevice
information,performdiagnostictests,andtrimthecalibrationonmost
HARTenabledtransmitters.Inthepastthiscouldonlybedonewitha
dedicated communicator.
TheLC-110HcanalsobeusedasaHARTmodem,allowingPChost
applications to perform more sophisticated diagnostics and setup.
HARTformatcommandsfromthehostarepassedtothetransmitter,
andresponsesarepassedback.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

2
1.2 Standard Equipment
Checktoseeifyourcalibratoriscomplete.Itshouldinclude:
•Calibrator
•6xAAAbatteries(installed)
•Testleads
- The LC-110 is supplied with one set of leads with alligator
clips installed
-TheLC-110Hissuppliedwithtwosetsoftests.Oneset
with alligator clips and the other set with mini-grabber
style ends
•Owner’sManual
•NISTCertificate
1.3 Safety information
Symbols Used
ThefollowingtableliststheInternationalElectricalSymbols.Someor
all of these symbols may be used on the instrument or in this
manual.
SymbolDescription
AC (Alternating Current)
AC-DC
Battery
CEComplieswithEuropeanUnionDirectives
DC
DoubleInsulated
ElectricShock
Fuse
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

3
PEGround
HotSurface(BurnHazard)
ReadtheUser’sManual(ImportantInformation)
Off
On
CanadianStandardsAssociation
The following definitions apply to the terms “Warning” and “Caution”.
•“Warning”identifiesconditionsandactionsthatmayposehazards
to the user.
•“Caution”identifiesconditionsandactionsthatmaydamagethe
instrument being used.
Usethecalibratoronlyasspecifiedinthismanual,otherwiseinjury
and damage to the calibrator may occur.
Warning
To avoid possible electric shock or personal injury:
•Donotapplymorethantheratedvoltage.Seespecificationsfor
supported ranges.
•Followallequipmentsafetyprocedures.
•Nevertouchtheprobetoavoltagesourcewhenthetestleadsare
plugged into the current terminals.
•Donotusethecalibratorifitisdamaged.Beforeyouusethe
calibrator,inspectthecase.Lookforcracksormissingplastic.Pay
particular attention to the insulation surrounding the connectors.
•Selecttheproperfunctionandrangeforyourmeasurement.
•Makesurethebatterycoverisclosedandlatchedbeforeyou
operate the calibrator.
•Removetestleadsfromthecalibratorbeforeyouopenthebattery
door.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

4
•Inspectthetestleadsfordamagedinsulationorexposedmetal.
Checktestleadscontinuity.Replacedamagedtestleadsbefore
you use the calibrator.
•Whenusingtheprobes,keepyourfingersawayfromtheprobe
contacts. Keep your fingers behind the finger guards on the
probes.
•Connectthecommontestleadbeforeyouconnectthelivetest
lead.Whenyoudisconnecttestleads,disconnectthelivetestlead
first.
•Donotusethecalibratorifitoperatesabnormally.Protectionmay
beimpaired.Whenindoubt,havethecalibratorserviced.
•Donotoperatethecalibratoraroundexplosivegas,vapor,ordust.
•Disconnecttestleadsbeforechangingtoanothermeasureor
source function.
•Whenservicingthecalibrator,useonlyspecifiedreplacementparts.
•Toavoidfalsereadings,whichcouldleadtopossibleelectricshock
orpersonalinjury,replacethebatteryassoonasthebattery
indicator appears.
•Toavoidaviolentreleaseofpressureinapressurizedsystem,shut
off the valve and slowly bleed off the pressure before you attach
the pressure module to the pressure line.
Caution
To avoid possible damage to calibrator or to equipment under test:
•Usetheproperjacks,function,andrangeforyourmeasurementor
sourcing application.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

5
2. Calibrator Interface and Operation
MARTEL ELECTRONICS
LC-110H Precision Loop
Calibrator w/HART™
100%
25%
0%
MENU
EXIT
Source 0.00%
4.000mA
Simultaneous % Display
4mA = 0.00%
20mA = 100.00%
Cursor
100% Key: Sets Output to 20mA
25% Key: Steps 4, 8, 12, 16, 20mA
0% Key: Sets Output to 4mA
Digital Knob Sets Output Value
Press to Move Cursor
Menu
Press to Enter Setup Menu
Press Again to Exit
Step / Ramp Enable
Backlight
On / Off
Mode
Figure 1
Figure1istypicalofthemaindisplayforalloperatingmodes.It
shows the present measurement or output value. The individual
operating modes are described below.
UsetheMENU/EXITkeytodisplaythemainmenudescribedin
section3toselecttheprimaryoperatingmodeofthecalibrator,to
accessthecalibratorsetupfunctions,ortoenterHARTmode.
Inoutputmode:
•Presstheknobtomovethedecadecursor.
•Rotatetheknobtoincrementordecrementtheoutputinsteps
indicated by the selected decade.
•Usethe3percentkeystodirectlysettheoutputtopreset
values.
•Usethestep/rampkeytoselectandstoptheseadvanced
modeswhicharedescribedinsection5.1.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

6
2.1 Milliamp Source
InthemASourcemodethecalibratoroutputsasignalfrom0to
24mAintoaloadofupto1000ohms(750ohmsiftheinternalHART
resistor is switched on).
Figure2showsthemaindisplayandtypicalconnectionsforthis
mode.
MARTEL ELECTRONICS
LC-110H Precision Loop
Calibrator w/HART™
100%
25%
0%
MENU
EXIT
Source 0.00%
4.000mA
4-20mA INPUT DEVICE
RECORDER/INDICATOR
SCADA SYSTEM ETC.
(+)
(–)
4 to 20mA
Figure 2
•Presstheknobtomovethedecadecursor.
•Rotatetheknobtoincrementordecrementtheoutputinsteps
indicated by the selected decade.
•Usethe3percentkeystodirectlysettheoutputtopresetvalues.
•Usethestep/rampkeytoselectandstoptheseadvancedmodes
whicharedescribedinsection5.1.Whenautomaticsteporramp
is active one of the following is shown in the lower left corner:
Automatic step:
Automatic ramp: ∧
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

7
•PressMENU/EXITtoenterthemainmenudescribedinsection3.
•Thetext‘ValveTest’isshowninthelowercenterwhenthevalve
testfunctionhasbeenenabledontheSetupMenu,section4.3.
Valvetestisdescribedinmoredetailinsection5.2.
•Thetext‘250Ω’textisshowninthelowerrightcornerwhenthe
HARTresistorhasbeenenabledontheSetupMenu,section4.4.
2.2 Milliamp Simulate
InthemASimulatemodethecalibratorfunctionslikea2-wire
transmitterbycontrollingtheloopcurrentfromanexternalpower
supply. This function is a great way to test a loop with the transmitter
removed.
Figure3showsthemaindisplayandtypicalconnectionsforthis
mode.
MARTEL ELECTRONICS
LC-110H Precision Loop
Calibrator w/HART™
100%
25%
0%
MENU
EXIT
Simulate 0.00%
4.000mA
POWER SUPPLY
30 VDC MAX
(+)
(–)
4 to 20mA
4-20mA
INPUT DEVICE
(+)
(–)
Figure 3
•Presstheknobtomovethedecadecursor.
•Rotatetheknobtoincrementordecrementtheoutputinsteps
indicated by the selected decade.
•Usethe3percentkeystodirectlysettheoutputtopresetvalues.
•Usethestep/rampkeytoselectandstoptheseadvancedmodes
whicharedescribedinsection5.1.Whenautomaticsteporramp
is active one of the following is shown in the lower left corner:
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

8
Automatic step:
Automatic ramp: ∧
•PressMENU/EXITtoenterthemainmenudescribedinsection3.
•Thetext‘ValveTest’isshowninthelowercenterwhenthevalve
testfunctionhasbeenenabledontheSetupMenu,section4.3.
Valvetestisdescribedinmoredetailinsection5.2.
•Thetext‘250Ω’textisshowninthelowerrightcornerwhenthe
HARTresistorhasbeenenabledontheSetupMenu,section4.4.
2.3 Milliamp Measure without 24V power
InthemAMeasuremode,thecalibratordisplaystheloopcurrent.
Figure4showsthemaindisplayandtypicalconnectionsforthis
mode.
MARTEL ELECTRONICS
LC-110H Precision Loop
Calibrator w/HART™
100%
25%
0%
MENU
EXIT
Measure 0.00%
4.000mA
4-20mA
OUTPUT DEVICE
(+)
(–)
4 to 20mA
Figure 4
•PressMENU/EXITtoenterthemainmenudescribedinsection3.
•Thetext‘250Ω’textisshowninthelowerrightcornerwhenthe
HARTresistorhasbeenenabledontheSetupMenu,section4.4.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
This manual suits for next models
1
Table of contents
Other Martel Test Equipment manuals

Martel
Martel 3001 User manual

Martel
Martel T150 User manual

Martel
Martel LC100 User manual

Martel
Martel IVC-222HPII User manual

Martel
Martel BetaGauge II User manual

Martel
Martel BETAGAUGE 330 User manual

Martel
Martel TC100 User manual

Martel
Martel PSC4010 User manual

Martel
Martel 311A-Ex User manual

Martel
Martel MC1210 User manual