LION alcolmeter 900 Instruction Manual

BS EN15964:2011 & IP54 Approved
lion alcolmeter® 900
Instrument Version: 900 English
*User Handbook*
Document Ref: RM92904 (Ver: DRAFT2)
© 2020, Lion Laboratories Limited

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 2

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 3
PLEASE...
Read this handbook carefully –BEFORE
using the instrument for the first time: then
retain it for future reference.

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 4
Contents
Contents .......................................................................................................4
Glossary ........................................................................................................5
Introduction..................................................................................................6
Principle of Operation...................................................................................6
Technical Details...........................................................................................8
Instrument Features ...................................................................................11
Subject Breath Test Procedure ...................................................................17
Last test recall.............................................................................................24
Other Instrument Functions .......................................................................25
Warning messages......................................................................................27
Care and maintenance................................................................................27
Dos and Don’ts............................................................................................31
Appendix I: Supplier’s Details ....................................................................32
Appendix II: Manufacturer’s Details ..........................................................33

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 5
Glossary
Terminology
BAC
Blood Alcohol Concentration
BrAC
Breath Alcohol Concentration
Calibration
Adjustment
Adjusts the calibration of the instrument
Calibration
Check
Checks the calibration of the instrument
Configuration
PC application that sets up the functionality of the
instrument
Deep-lung air
Air within the lungs that is in intimate contact with the
blood
Dry Gas
Dry alcohol gas of a known concentration provided from a
gas cylinder
Operator
Person who is carrying out the breath test
PIN
Personal Identification Number, used to unlock features
Sampling
Port
Metal protrusion in top channel of the instrument that fits
into the mouthpiece
Service lock
Time until the next service should be carried out
Subject
Person who is to have a breath test performed upon them
Suck back
Intake of breath by the subject during the sampling
routine
Suspicious
liquid
Liquid with an unknown alcohol content
Trigger Flow
Minimum air flow speed necessary for automatic
sampling
Wet vapour
Wet alcohol vapour of a known concentration provided by
a simulator

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 6
Introduction
The lion alcolmeter® 900 devices are hand-held portable instruments for
fast, on the spot determination of a subject’s breath (or blood) alcohol
concentration level. They are some of the latest in the range of lion
alcolmeter® instruments produced by Lion Laboratories Limited of Barry,
South Wales, in the United Kingdom. Lion Laboratories is a company
specialised in breath alcohol detection technology for applications in
traffic law enforcement, commercial health and safety, and in medicine.
This handbook explains the following aspects of the instrument, so please
read it carefully:
•Basic principles of operation.
•How to run a breath test.
•Interpretation of results.
•Care and basic maintenance.
•Calibration checking.
•Calibration adjustment.
•System configuration.
Principle of Operation
The lion alcolmeter®900 uses an electrochemical fuel cell sensor to
measure the concentration of alcohol vapour in the subject’s expired
breath, a measured amount of which is drawn from the mouthpiece into
the fuel cell by means of the sampling system.
In the fuel cell a voltage is generated in proportion to the breath alcohol
concentration: it is amplified and displayed in terms of the subject’s
alcohol level.

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 7
The breath sampling system is factory-set to sample deep-lung air; it is the
breath from the bottom of the lungs that truly reflects the amount of
alcohol in the body.
To ensure repeatable and accurate results across the temperature range,
and particularly at low temperatures the lion alcolmeter®900 uses a
heater to heat the metal sampling port and the inlet port of the fuel cell.
This stops the build up of condensation which can affect the alcohol
concentration in the sample.

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 8
Technical Details
Factor
Target
Comments
Specificity
Specific to alcohol only
When used correctly, not
sensitive to acetone, paint
and glue evaporation, food,
sweets, methane and all
materials without any
alcohol content which can be
found in human breath
Measurement
range
0.00mg/L breath (BrAC)
Note 2
Minimum Note 1
2.00mg/L breath (BrAC)
Note 2
Maximum
Measurement
Accuracy
±0.02mg/L ≤ 0.200mg/L
±10% above 0.20mg/L
In Active Mode only
Measurement
step
0.01mg/L
0.001mg/L in metrological
mode
Response time
less than 15 seconds
0.25mg/L alcohol level
Recovery time
less than 30 seconds
0.25mg/L alcohol level
Sampling
frequency
Less than 10 seconds
Negative test result
Operating
temperature
range
-5oC Note 2
Minimum
40oC Note 2
Maximum

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page 9
Ready time
15 seconds
Less than 60 seconds
From switch on at 20oC
From switch on at -5oC
Operating
humidity range
20% to 95%
Non-condensing
Storage
temperature
-20oC
Minimum
70oC
Maximum
Measurement
method
Active
Using mouthpiece, with
automatic sampling
Trigger Flow
9L/min Note 2
In active test mode only
Trigger Volume
1.2L Note 2
In active test mode only
Power
supply
4 x AA Alkaline
batteries
Refer to page 16
Number of
tests
More than 8000 on a
set of batteries Note 3
Zero alcohol repeat tests
Database
capacity
10000 records
When full, the database rolls
over and replaces the oldest
record with latest record
Instrument
weight
210g
Without any accessories or
batteries

lion alcolmeter® 900 User Handbook (Ref: RM92904)(ver: DRAFT2)
Page
10
Note 1: Below a value of 0.02mg/L the instrument will report zero.
Note 2: These values are configurable and may vary depending upon the
customer’s requirements.
Note 3: Testing was carried out by performing tests without alcohol
present sequentially at +5ºC ambient temperature.
Audible notes
Instrument will beep
Upon successful and
unsuccessful tests as a
minimum
Radio frequency
interference
Compliance
EN 61000-4-3:2020
Mechanical
stress
Compliance
EN 62262:1995 / A1:2002 (IK-03)
Dust and ingress
protection
Compliance
EN 60529:2001 (IP-54)
Calibration
frequency
Every 6
months
The instrument shall display, and lock
if the instrument is not calibrated
within this calibration time
Other manuals for alcolmeter 900
1
Table of contents
Other LION Test Equipment manuals