A revolutionary, hand-held strength measurement system
Model 01163
THE LAFAYETTE MANUAL MUSCLE TEST SYSTEM
USER’S MANUAL

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
The Lafayette Manual Muscle Test System (MMT)
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
2
Table of Contents:
System Description 3
Features and Specifications 4
Precautions 4
Basics of Operation 5
Function Buttons 6
Reset Button 6
Range Button 6
Scroll/LB/KG Button 6
Enter/Store Button 6
Menu Button 6
Main Measurement Screen 7
Menu Descriptions 8
Main Menu 8
Show Data Menu 8
Function Descriptions 9
Download 9
Show Data 9
Version Identifier 10
Deleting Data Registers 10
Beep 11
Time 11
Auto Store 12
Data Storage 12
Reset 13
Range 13
Low Battery Detection 14
Battery Saver Mode 14
Battery Change 15
Stirrup Change 16
Error and Message Screen Summary 16
Low Battery Warning 16
Memory Full Warning 17
Diagnostic Alert Message 17
Replacement Parts 18
Appendix A -
Torque Measurements 18
Ordering Information 20

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
Model 01163 User Instructions 3
System Description:
The new Lafayette Manual Muscle Test System (MMT) is a hand-held device
for objectively quantifying eccentric muscle strength. Using the Lafayette MMT,
the peak force required to break an isometric contraction is measured as the
examiner applies force against the subject. The Lafayette MMT readily lends
itself to large, pre-participation screenings or field examinations where bulky,
more expensive systems do not. The microprocessor control allows for storage
of calibration values and automatic drift compensation, resulting in reliable,
accurate, and stable muscle strength readings. The MMT also features interactive
menus to allow a wide range of options such as data storage and preset test times
to be implemented. While powerful and versatile, the MMT is still small enough
to fit comfortably in the palm of the hand. Its ergonomic design allows for both
patient and tester comfort while easily conforming to most manual muscle testing
protocols.

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
The Lafayette Manual Muscle Test System (MMT)
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
4
Note: 1. When performing repeated tests, inconsistent placement of the MMT will affect scores.
2. Extreme temperature, especially heat, may affect the values obtained.
3. The MMT cannot tolerate the stress of being used as a floor scale.
4. Care should be taken not to drop the MMT, as it may affect the calibration.
5. Exceeding the force limit (300lbs./136kg) may permanently damage the MMT and/or
invalidate the calibration.
Features and Specifications:
System Features:
· Designed for high inter-instrument reproducibility
· Three, easy to change molded plastic stirrups with pads
· Force measurement in pounds or kilograms (user selectable)
· Measures peak force, time to reach peak force and total test time
· Data storage for up to 52 tests in on-board memory (peak force and time to
reach peak force)
· Manual or automatic storage of data
· Dual measurement range: 0-300lbs. (136.1kg) or 0-50lbs. (22.6kg)
· Selectable test time from 1-10 seconds
· Tone to indicate end of preset test time
· Microprocessor controlled
· Easy to read graphical LCD display
· Manual ON/OFF switch
· Manual or automatic reset
· Built-in stored data browsing capability
· Low battery detection indicated by tone and icon
· Automatic battery saving sleep mode
· Interactive menus which allow user to select device options
· Battery powered: (1) lithium battery
· Minimal measurement drift
System Specifications:
· Size: 3” x 4” x 1.5” (7.6cm x 10.2cm x 3.8cm)
· Weight: 10.6oz (300g)
· Range: 0-300lbs. (136.1kg) / 0-50 pounds (22.6kg)
· Accuracy: ± 1% over full scale (both ranges)
· Resolution: 0.4lb (0.2kg) high range / 0.1lb (0.1kg) low range
· Battery Life: 80-85 hours, 10-12 hours after low battery condition
· Timing Accuracy: ± 0.03%
· Data Storage Capacity: 52 tests
· Calibration Points: 0, 25 and 50lbs. (0,11.3 and 22.6kg)
· Preset Test Length: 1-10 seconds; in 1 second increments

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
Model 01163 User Instructions 5
(3) LCD Screen
Basics of Operation:
The Lafayette MMT is small enough to be held in
one hand and easily read. The size and weight of the
MMT permit the examiner to use the same
procedures and break test techniques
(described in the literature and taught
by academic institutions) without
any modification of technique or
positioning. The unit is simply placed
between the examiner’s hand and the
limb being tested. The examiner’s
downward force is transmitted to the
limb through the MMT unit. The MMT is
designed for one hand operation.
It can comfortably fit in the right or left hand.
The hand is placed under the strap and around
the body of the MMT. This allows easy access to the
RESET button (1) with the thumb. All of the other buttons are pressed using the opposite
hand.
The MMT is activated by sliding the power switch to the “ON” position (2).
Measurements are taken by pressing the padded stirrup against the muscle being tested on
the subject. The force and time data are displayed on the LCD screen (3) on the face of
the MMT.
(1) RESET Button
(2) ON / OFF
Switch

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
The Lafayette Manual Muscle Test System (MMT)
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
6
Function Buttons:
The MMT has five function buttons that control the menus and allow the selection of
options and settings.
Reset Button: The RESET button is located on the top of the MMT. It is placed for easy
access with the thumb regardless of right or left hand operation. The RESET button clears
the display of all data and sets the zero point for the measurements. Care should be taken
not to apply force while pressing the RESET button. This will cause the MMT to sound
two successive audible beeps and display the following message “Diagnostic Alert Please
Press Reset”. The message will continue to be displayed and the user will not be able to
perform another measurement until the RESET button has been pressed.
Range Button: The RANGE button toggles between high
and low measurement range. An “H” or “L” is displayed
on the Main Measurement Screen to indicate the MMT
rangesetting. Changing the range causes the MMT to
automatically reset and clear all time and force
values.
Scroll/LB/KG Button: The LB/KG button has
dual functionality depending on which screen is
accessed. In the Main Measurement Screen, the
LB/KG button toggles the force measurement scale
between pounds and kilograms. An indicator on
the Main Measurement Screen shows which scale
is selected. When a menu or data display screen
is active, this button becomes a SCROLL button.
The SCROLL button is used to select options by
advancing a cursor from one option to the next.
The SCROLL button is also used to step through the
stored data.
Enter/Store Button: The ENTER/STORE button
has dual functionality depending on which screen
is accessed. In the Main Measurement Screen,
the button is used as a STORE button. Pressing
the STORE button stores the current test data in
the MMT memory. Only the peak force and peak time are stored in memory. When the
menu screens are accessed, the button is used as an ENTER button. The ENTER button is
used to activate settings and menu options. It is also used to clear some errors and warning
messages.
Menu Button: The menu button is used to enter the Main Menu. The Main Menu is used
to select the MMT options and access the sub-menus.
1
2345
1
2
3
4
5

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
Model 01163 User Instructions 7
Main Measurement Screen:
The Main Measurement Screen shows all of the current measurement information. This screen is
shown whenever a measurement is in progress. This screen is also the default start-up display.
Peak Force: The peak measurement force is displayed in large numbers on the upper left of
the Main Measurement Screen. This value is the peak pounds or kilograms applied during a
test. Force values are displayed to tenths of a pound (or kilogram).
Peak Time: The peak time value is displayed in seconds on the bottom left of the screen
under the peak force. This number is the time during the test when the peak force occurred.
Time values are displayed to hundredths of a second.
Instantaneous Force: The instantaneous force
value is displayed on the upper right on the
Main Measurement Screen. This value is the
force applied to the MMT at any time.
Running Time: The running time value is
displayed on the lower right of the Main
Measurement Screen. This value shows the total
time that force has been applied. After a test has
ended, this value shows the total length of the
test in seconds.
Range Indicator (H or L): The range indicator
shows if the MMT is set in the high or low
measurement setting. If in kg’s, the high range
will be 0kg - 136.1kg and the low range will be
0kg - 22.6kg. If in lbs., the high range will be
0lbs. - 300lbs. And the low range will be 0lbs. - 50lbs. The range indicator is the upper most
indicator in the center of the Main Measurement Screen. The Range is changed by pressing
the RANGE button.
Scale Indicator (LB or KG): The scale indicator shows if the MMT is set to measure in
pounds or kilograms. The scale indicator is the second indicator in the center of the Main
Measurement Screen. The scale is changed by pressing the LB/KG button.
Low Battery Indicator: The low battery indicator is present when a low battery condition
exists. The low battery indicator is the third indicator in the center of the Main Measurement
Screen.
Time Units Label: The time units label is the “S” on the bottom center of the Main
Measurement Screen. This label indicates that the time measurements are in seconds.
This label does not change.
6
10
9
11
12
13
7
8
7
8
6
9
10
11 12

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
The Lafayette Manual Muscle Test System (MMT)
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
8
Menu Description:
Main Menu: The Main Menu is used to access the MMT options and sub-menus. The menu
options are:
1. Activate the Download feature
2. Enter the Show Data Menu
3. Activate the preset test time beep
4. Set the Time
5. Activate the Auto-Store feature
6. Enter the Calibration procedure (for manufacturer use only. Contact Lafayette Instrument
Co. for calibration.)
7. Exit the Menu
Menu items are selected by pressing the SCROLL button to move the cursor to the desired item
and pressing the ENTER button to activate the item. See Function Descriptions for details on
each option.
Download:
The Download function allows data to be
transferred to a computer for permanent
storage. The download is accomplished
via infrared connection to a USB
download device. This device can be
purchased separately from Lafayette
Instrument (Model 01163IR).
Function Descriptions:
Show Data Menu: The Show Data Menu allows the user to view stored data values or clear
the memory registers. The Show Data Menu is accessed by selecting SHOW DATA from the
main menu screen. Upon entering the menu, the user can
1. View the stored data.
2. Delete the memory registers.
3. View the version number.
3. Exit the Menu.
Menu items are selected by pressing the SCROLL button to move the cursor to the desired
item and pressing the ENTER button to activate the item. See the sections (on page 9 and 10)
regarding SHOW DATA, DELETING DATA REGISTERS, and VERSION IDENTIFIER for
details on the operation of those menu items.

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
Model 01163 User Instructions 9
To download:
1. Plug in the USB download device to the host computer
2. Run the download application software.
During download, the MMT will
attempt to establish communication with
the computer. The computer must have
the download software running and the
USB download peripheral connected for
communication to be established.
3. Point the Infrared port in the bottom of the MMT at the USB download device.
Since the connection is infrared, the line of sight between the computer and MMT must
be free from obstructions. The infrared transmitter on the bottom of the MMT must be
pointed directly at the USB download device. The MMT can not be more than 12 feet from
the download device. If the MMT is more than 12 feet away, errors may result in the data
transmission.
4. Enter the Main Menu by pressing the MENU button
5. Press ENTER to begin download.
If a connection can not be established, an error is given along with two audible beeps. An
option menu is then given which allows (1) a reattempt of communication or (2) canceling
the operation. The download device receives the data and transfers it to the computer
software. The software then checks the data for validity and reports the checksum back to the
MMT. Once the MMT receives the correct checksum, the data registers are cleared. If the
correct checksum is not received , the error message is given and the data is not cleared.
Download Function (continued):
Show Data:
The Show Data function allows the
stored data to be display and reviewed.
To review stored data:
1. Enter the Main Menu by pressing
the MENU button.
2. Press the SCROLL button one (1) time to move the cursor to SHOW DATA.
3. Press ENTER to display the SHOW DATA menu.
4. Press the SCROLL button one (1) time to move the cursor to SHOW DATA.
5. Press ENTER to show the data.
6. Press the SCROLL button to step through the data registers.
7. Press the RESET button to exit to the Main Measurement Screen.
Note: Only the Peak Force (in lbs. or kg) and the peak time are stored in memory. Empty
data registers are shown as “7. KL G . S”. The SHOW DATA function can also be
used to determine how many empty registers are available.

3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
The Lafayette Manual Muscle Test System (MMT)
3700 Sagamore Parkway North . PO Box 5729 . Lafayette, IN 47903 USA . Ph: 765-423-1505 Fax: 765-423-4111 . www.lafayetteinstrument.com . E-mail:[email protected]
10
Deleting Data Registers:
When the data memory is full, a
warning message is given to clear the
data registers or discontinue storage.
The data registers can be cleared by
manually clearing the registers from
the SHOW DATA menu.
To clear all data registers:
1. Enter the Main Menu by pressing the MENU button.
2. Press the SCROLL button one (1) time to move the cursor to SHOW DATA.
3. Press ENTER to turn the display to SHOW DATA menu.
4. Press the SCROLL button two (2) times to move the cursor to CLEAR ALL
DATA REGS.
5. Press ENTER to clear data registers.
6. Press ENTER to confirm delete operation.
7. MMT confirms that the delete has been completed.
Version Identifier:
The MMT has a version ID number to help track new features and changes to the device.
To view the ID number:
1. Enter the main menu by pressing the MENU button.
2. Press the SCROLL button one (1) time to move the cursor to SHOW DATA.
3. Press ENTER to display the
SHOW DATA menu.
4. The version ID is in the upper
right hand corner of the SHOW
DATA menu.
Table of contents