KNF NF5 User manual

KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions Keep for future reference!
DIAPHRAGM LIQUID PUMP
NF 5, NFB 5
Operating and
Installation
Instructions
Read and observe these Oper-
ating and Installation Instruc-
tions!
An additional letter prefixing
the NF model code is a coun-
try-
specific designation, with no
technical relevance.
Contents Page
1. About this document ................................................................. 2
2. Use ........................................................................................... 3
3. Safety ....................................................................................... 4
4. Technical data .......................................................................... 6
5. Assembly and function ............................................................. 9
6. Installation and connection ..................................................... 10
7. Operation ................................................................................ 15
8. Servicing ................................................................................. 19
9. Troubleshooting ...................................................................... 21
10. Accessories ............................................................................ 23
11. Decontamination declaration .................................................. 24
KNF Flodos AG
Wassermatte 2
6210 Sursee, Switzerland
Tel +41 (0)41 925 00 25
Fax +41 (0)41 925 00 35
www.knf-flodos.ch
info@knf-flodos.ch
NF 5 KP DC
-
M 12V
DC
-
M / DCB / DCB
-
4
KP / KT /
TT [Ch.
4
]
NF / NFB
Supply
voltage [Ch.
4
]
5
-
/ PMLxxxx /
PLxxxx [Ch.
1
]

Diaphragm liquid pump NF 5, NFB 5 About this document
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 2
1. About this document
1.1. Use of the Operating and Installation Instruc-
tions
The Operating and Installation Instructions are part of the pump.
Forward the Operating and Installation Instructions to any sub-
sequent owners of the pump.
Customer-specific project pumps (pump models which begin with
"PL" or "PML") may differ from the Operating and Installation In-
structions.
In the case of project pumps, take note of any additionally
agreed specifications.
1.2. Symbols and markings
Warning
WARNING
This symbol indicates a potential danger.
It also indicates the possible consequences of failure
to observe the warning. The signal word (e.g. "Warn-
ing") indicates the level of danger.
Here you will see actions for avoiding the danger
and potential consequences.
Danger levels
Signal word
Meaning Consequences if not observed
DANGER warns of immedi-
ate danger
Consequences are death or se-
rious injury and/or serious prop-
erty damage.
WARNING
warns of poten-
tial danger
Death or serious injury and/or
serious damage to property are
possible.
CAUTION warns of a poten-
tially dangerous
situation
Minor injury or damage to prop-
erty are possible.
Tab. 1
Other information and symbols
This indicates a required activity (step).
This indicates the first step of a required activity. Additional
consecutively numbered steps follow.
This symbol indicates important information.
Project pumps

Diaphragm liquid pump NF 5, NFB 5 Use
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 3
2. Use
2.1. Intended use
The pumps are intended for transferring and metering liquids.
Owner's responsibility
Only install and operate the pumps under the operating parameters
and conditions described in Chapter 4 About this document, Tech-
nical data.
Only completely installed pumps may be taken into service.
Before transferring or metering a medium, check whether the me-
dium can be transferred danger-free in the specific application
case.
Before using a medium, check the compatibility of the materials of
the pump head, pump housing, diaphragm and valves with the me-
dium.
The temperature of the medium must lie within the permissible
temperature range (see Chapter 4).
The transferred medium should not contain particles as these can
prevent the pump from working correctly. If this cannot be guaran-
teed, a filter < 50 m with sufficiently large filter area must be used
upstream of the pump.
Filters may be ordered as accessories, see Spare Parts and
Accessories (Chapter 10).
2.2. Improper use
DANGER
The pumps must not be operated in an explosive
atmosphere.
Please contact your local KNF partner for special designs that
are not included in the technical specification (www.knf.com).
Operating parameters and
conditions
Requirements for
transferred medium

Diaphragm liquid pump NF 5, NFB 5 Safety
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 4
3. Safety
Observe the safety precautions in Chapter s
6. Installation and connectionand 7.Operation.
The pumps are built according to the generally recognised rules of
technology and in accordance with the pertinent occupational
safety and accident prevention regulations. Nevertheless, dangers
may occur during their use which may lead to injuries to the user or
others, or to damage to the pump or other property.
Only use the pumps when they are in a good technical and proper
working order, in accordance with their intended use, observing the
safety advice within the Operating and Installation Instructions, at
all times.
Make sure that only trained and instructed personnel or specially
trained personnel work on the pumps. This especially applies to
assembly, connection and servicing work.
Make sure that all personnel have read and understood the Oper-
ating and Installation Instructions, and in particular the "Safety"
Chapter .
Always ensure adherence to all pertinent accident prevention and
safety regulations when working on and operating the pump.
When transferring dangerous media, observe the safety regula-
tions for handling such media.
Always ensure adherence to all information stickers on the pumps,
such as flow direction arrows and type plates, and keep stickers in
legible condition.
All replacement parts should be properly stored and disposed of in
accordance with the applicable environmental protection regula-
tions. Ensure adherence to the pertinent national and international
regulations. This especially applies to parts contaminated with toxic
substances.
Dispose of all packaging in an environmentally-appropriate
manner. The packaging materials are recyclable.
Dispose of end-of-life equipment in an environmentally
friendly manner. Use appropriate waste collection systems
for the disposal of end-of-life equipment. Used pumps con-
tain valuable recyclable materials.
Personnel
Working in a
safety-conscious manner
Handling dangerous media
Notes
Environmental protection
Disposal

Diaphragm liquid pump NF 5, NFB 5 Safety
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 5
The pumps comply with the fundamental requirements of Directive
2011/65/EU (RoHS2).
The pumps comply with the safety requirements regarding electro-
magnetic compatibility in Directive 2004/108/EC.
For the purposes of the Machinery Directive 2006/42/EC, pumps
are “partly completed machinery", and are therefore to be regarded
as not ready for use. Partly completed machinery may not be com-
missioned until such time as it has been determined that the ma-
chine in which the partly completed machinery is to be assembled
conforms to the provisions of the Machinery Directive 2006/42/EC.
The essential requirements of Annex I of Directive 2006/42/EC
(general principles) are applied and observed.
The following harmonised standards are met:
NF 5 DC-M
EN 55022
EN 55011
NF 5 DCB / DCB-4
EN 61000-6-2
EN 61000-6-3 (incl. EN 55022 / EN 55011)
NFB 5 DCB / DCB-4
EN 61000-6-2
EN 61000-6-3 (incl. EN 55022 / EN 55011)
All repairs to the pump(s) must be carried out by the relevant KNF
customer service team.
Only use genuine parts from KNF for servicing work.
EU directives/standards
Customer service and repairs

Diaphragm liquid pump NF 5, NFB 5 Technical data
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 6
4. Technical data
Pump materials
The pump type KP stands for:
Assembly Material1)
Pump head* PP
Anchor valves / seals EPDM
Diaphragm NF5 EPDM
Diaphragm NFB5 PTFE-coated
Tab. 2 1) according to DIN ISO 1629 and 1043.1
The pump type KT stands for:
Assembly Material1)
Pump head* PP
Anchor valves / seals FFKM
Diaphragm PTFE-coated
Tab. 3 1) according to DIN ISO 1629 and 1043.1
The pump type TT stands for:
Assembly Material1)
Pump head* PVDF
Anchor valves / seals FFKM
Diaphragm PTFE-coated
Tab. 4 1) according to DIN ISO 1629 and 1043.1
* The pump head (Fig. 1) comprises an intermediate plate and a connect-
ing plate.
Hydraulic ratings NF 5
Parameter Value
Flow rate DC-M [ml/min] 1), 2) 70
Flow rate DCB-4 [ml/min] 1), 2) 5-40
Flow rate DCB [ml/min] 1), 2) 40
Permissible pressure [bar g] 1
Suction head KP [mWG] 3
Suction head KT, TT [mWG] 2.5
Tab. 5

Diaphragm liquid pump NF 5, NFB 5 Technical data
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 7
Hydraulic ratings NFB 5
Parameter Value
Flow rate DCB-4 [ml/min] 1), 2) 2x 5-40
Flow rate DCB [ml/min] 1), 2) 2x 40
Permissible pressure [bar g] 1
Permissible pressure (with restricted oper-
ating time) [bar g]
4
Suction head [mWG] 3
Tab. 6
1) Measured with water under KNF standard conditions, against atmos-
pheric pressure. Pump head material: KP
2) Flow rates may vary from the values shown, depending on fluid viscos-
ity, ambient temperature, pump head material and the hoses / hose con-
nectors used.
Hydraulic connections NF 5
Parameter Value
Hose barb OD (mm) 3
Recommended hose size ID [mm] 3
Tab. 7
Hydraulic connections NFB 5
Parameter Value
Hose barb OD (mm) 3
Recommended hose size ID [mm] 3
Tab. 8
Specifications NF 5 DC-M
Motor voltage 12V 24V
Power consumption [W] 1.44 1.44
Max. I load [A] 0.12 0.06
Max. permissible current consumption [A] 0.17 0.09
Lead size [-] AWG22
Protection class [-] IP30
Weight1) [g] 56
Tab. 9
Specifications NF 5 DCB / DCB-4
Motor options DCB-4 DCB 12V DCB 24V
Supply voltage
[V] 10…26.4 12 24
Power con-
sumption [W] 1.1 1.0 1.2
I loadmax. [A] 0.04…0.09 0.08 0.05
Lead size [-] AWG28
Protection class
IP40
Weight 1) [g] 48
Tab. 10

Diaphragm liquid pump NF 5, NFB 5 Technical data
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 8
In the case of type NF 5 DCB-4 the supply voltage is not de-
signed for operation with pulse-width modulated voltages
(PWM).
However, a PWM signal (frequency 500Hz or higher) can be
used to control the control voltage.
Electrostatic sensitive components (ESD)
1) The weight may differ slightly from the stated value, depending on the
version.
Specifications NFB 5 DCB / DCB-4
Motor options DCB-4 DCB 12V DCB 24V
Supply voltage
[V]
10…26.4 12 24
Power con-
sumption [W] 2.1 1.6 1.9
I loadmax. [A] 0.08…0.19 0.13 0.08
Lead size [-] AWG28
Protection
class
IP40
Weight 1) [g] 75
Tab. 11
1) The weight may differ slightly from the stated value, depending on the
version.
In the case of type NFB 5 DCB-4 the supply voltage is not de-
signed for operation with pulse-width modulated voltages
(PWM).
However, a PWM signal (frequency 500Hz or higher) can be
used to control the control voltage.
Electrostatic sensitive components (ESD)
Other parameters
Parameter Value
Permissible
ambient temperature range (°C)
+ 5 to + 40
Permissible media temperature range (°C) + 5 to + 80
Permissible kinematic viscosity of medium
[cSt]
≤ 150
Tab. 12

Diaphragm liquid pump NF 5, NFB 5 Assembly and function
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 9
5. Assembly and function
Assembly
1 Outlet
2 Inlet
3 Connecting plate
4 Drive
5 Connecting leads
6 Head plate
7 Intermediate plate
Fig. 1: diaphragm liquid pump NF 5
1 Outlet valve
2 Inlet valve
3 Working chamber
4 Diaphragm
5 Eccentric
6 Connecting rod
7 Pump drive
Fig. 2: Pump assembly
diaphragm liquid pumps are based on reciprocating displacement
pump technology. The elastic diaphragm (4) is moved up and
down by the eccentric (5) and the connecting rod (6). In the down-
ward stroke it aspirates the medium to be transferred via the inlet
valve (2). In the upward stroke the diaphragm pushes the medium
out of the pump head via the outlet valve (1). The diaphragm her-
metically seals off the working chamber (3) from the pump drive
(7).
NFB 5 Version (double-headed pump, see Fig. 3)
This pump type has two NF5 pump heads that can be operated individu-
ally or or in parallel connection (see Chapter 6.3.2).
Fig. 3: NFB 5
(double-headed pump)

Diaphragm liquid pump NF 5, NFB 5 Installation and connection
KNF Flodos AG BA_NF5_NFB5_EN_01_172971.docx
Translated from the Original Operating and Installation Instructions 10
6. Installation and connection
Only install the pumps under the operating parameters and condi-
tions described in Chapter 4,Technical data.
Observe the safety precautions (see Chapter 3).
6.1. Installation
Before installation, store the pump at the installation location to
bring it up to ambient temperature.
Mounting dimensions (see Fig. 4 to Fig. )
Fig. 4: Mounting dimensions NF 5 DC-M
Fig. 5: Mounting dimensions NF 5 DCB
Mounting dimensions
Other manuals for NF5
1
This manual suits for next models
1
Table of contents
Other KNF Water Pump manuals
Popular Water Pump manuals by other brands

Sykes AmeriPumps
Sykes AmeriPumps GP100M Operation and maintenance instructions

DUROMAX
DUROMAX XP WX Series user manual

BRINKMANN PUMPS
BRINKMANN PUMPS SBF550 operating instructions

Franklin Electric
Franklin Electric IPS Installation & operation manual

Xylem
Xylem e-1532 Series instruction manual

Milton Roy
Milton Roy PRIMEROYAL instruction manual