KNT 310
Data recorder and data reader for KISTOCK
dataloggers
Weight
340 g
Display
Graphic display 128 x 128 pixels
Dimensions: 50 x 54 mm
Blue bac light
keypad
Metal-coated
5 eys and 1 joystic
Conformity
Electromagnetical compatibility
(as per NF EN 61326-1 standard)
Material
Compatible with food-industry
environment
Shoc -proof ABS housing
Protection
IP54
Power supply
4 al aline batteries 1,5V LR6
Environment
Air and neutral gases
Auto shut-off
Adjustable from 0 to 120 mn
TECHNICAL FEATURES
Storage capacity 500 000 points or 50 data sets
Speed of data transfer From istoc to data recorder: 400 values/second
From data recorder to PC: 1000 values/second
perating temperature From 0 to 50°C
Storage temperature From-20 to +80°C
Radio span (in free field) 300 m
Frequency 868 MHz
Data storage In the data recorder until downloading on PC
FEATURES OF HOUSIN
- Compatible with wireless Kistocks, wired kistocks 2010 and previous
ranges
- Possibility to record up to 50 datasets
- Possibility to view real-time measurements from a istoc datalogger
• Side view• Front view
• Top view
DIMENSIONS (mm)
KEY POINTS
57.4
161.9
80.8
ELMTEC
Kattreppeln 28 • 38154 Königslutter • 05353 / 9545 - 0 • ¬05353 / 9545 - 45 •

USING DATA RECORDER
Before downloading data stored from a istoc datalogger, this istoc has to registered in the data
recorder.
See ilog RF user manual to register a datalogger in the data recorder.
➢ Turn on the data recorder.
➢ Go to “Kistock” with left and right arrows then press K.
➢ Select “RF Connection” with up and down arrows then press K (see fig.1).
A progress bar then the list of registered kistock are displayed.
➢ Select the Kistoc to download with up and down arrows then press K (see fig.2).
Connection screen then “Infos” screen are displayed. This last screen displays the
kistock serial number, dataset name and number of dataset point.
From this screen:
➢ Select “Discharge“ with left and right arrows then press K.
Discharge screen with its progress bar is displayed.
hen discharge is finished:
➢ Select “Display“ then press K to visualize main features of the dataset : type of
Kistoc , serial number, dataset name, dates of beginning and end, number of
measurements and of registered channel with minimum and maximum values for each
channel (see fig.3).
Or
➢ Select “New dataset“ then press K to launch a new dataset.
● Discharge date from a wired kistock to the data recorder
➢ Connect the data recorder and the istoc with the cable.
➢ Turn on the data recorder.
➢ Go to “Kistock“ with left and right arrows then press K.
➢ Select “Cable Connection/» with up and down arrows then press K (see fig.4).
A progress bar is displayed «Infos“ screen. This last screen displays the kistock serial
number, dataset name and number of dataset point.
From this screen:
➢ Select “Discharge“ with left and right arrows then press K.
Discharge screen with its progress bar is displayed.
hen discharge is finished:
➢ Select “Display“ then press K to visualize main features of the dataset : type of
Kistoc , serial number, dataset name, dates of beginning and end, number of
measurements and of registered channel with minimum and maximum values for each
channel (see fig.5).
Or
➢ Select “New dataset“ then press K to launch a new dataset.
Fig. 4
Fig. 2
Fig. 3
Fig. 1
● Data downloading from a wireless kistock to data recorder
Fig. 5
Kattreppeln 28 • 38154 Königslutter • 05353 / 9545 - 0 • ¬05353 / 9545 - 45 •

● Set the device
This part allows to adjust the following parameters of the data recorder:
- Date/Time
- Beep
- Auto shut-off
- Contrast
- Bac light
- Memory
- Language
➢ From main screen, go to “Params” then press K.
Set date and time/:
➢ Go to “Date and time” line with up and down arrows then press K.
➢ Set date and time with arrows.
➢ Press K to validate.
● Visualize data from a kistock in real-time on the data recorder
When a Kistoc is registered in the data recorder, it is possible to visualize its
measurements in real-time:
➢ Go to “Kistock” with left and right arrows then press K.
➢ Select “RF connection” with up and down arrows then press K.
➢ Select the Kistoc up and down arrows then press K.
➢ Select “Values” with left and right arrows.
Data recorder displays values measured by the Kistock. (see fig.7).
● Delete dataset stored in the data recorder
➢ Turn the data recorder.
➢ Go to “Memory” with left and right arrows.
The device as s to erase:
➢ Go to YES with up and down arrows to erase all datasets
or
➢ Go to N to cancel and bac to main screen or press Esc.
Fig. 6
Fig. 7
➢ Select an operator and/or a customer if necessary.
➢ Clic on “Next” button.
➢ Select the dataset file destination that will be created.
➢ Select the required option concerning the files management.
➢ Clic on “Finish” button.
● Discharge stored dataset from the data recorder to a computer
➢ Connect the data recorder on the base.
➢ Turn on the data recorder.
➢ Launch Kilog RF software.
➢ Select as communication support the data recorder.
➢ Clic on “Transmit data” button.
➢ Tic the dataset to transmit.
Kattreppeln 28 • 38154 Königslutter • 05353 / 9545 - 0 • ¬05353 / 9545 - 45 •

Set backlight/:
➢ Go to “Backlight” line with up and down arrows then press K.
➢ Set bac light between 1 and 9 or on Auto.
➢ Press K to validate.
Delete datasets/:
➢ See previous page.
Select the language of the device/:
➢ Go to “Language” line with up and down arrows then press K.
➢ Select language: French, English or another installed language.
➢ Press K to validate.
Set contrast/:
➢ Go to “Contrast” line with up and down arrows then press K.
➢ Set contrast between 0 and 9.
➢ Press K to validate.
Activate or deactivate beep/:
➢ Go to “BEEP” line with up and down arrows then press K.
➢ Select N to activate or FF to deactivate the beep of the eys.
➢ Press K to validate.
Activate or deactivate auto shut-off of the data recorder/:
➢ Go to “Extinction” line with up and down arrows then press K.
➢ Select OFFto deactivate it or the time before automatic shut-off (15, 30, 45 min or 1h).
➢ Press K to validate.
● About data recorder/:
➢ Turn on the device.
➢ Go to “Infos” with left and right arrows..
➢ Press K.
Data recorder serial number, number of stored datasets and firmware version are displayed.