DAB_A01-BA-def-1510 9
2. Temperature adjustment
Wird bei der Temperaturkalibrierung die zulässige Abweichung über-/unterschritten
kann die Temperatur des Instruments wie nachfolgend beschrieben justiert werden.
If the admissible deviation is exceeded / not reached during temperature calibration the
temperature of the instrument can be adjusted as described below.
1. Switch off instrument.
2. Press , quickly release, then press and at the same time.
„8.1x“ will be displayed. If not, separate device from power supply and restart
step two.
3. Adjust temperature using arrow keys . Each time you press the but-
ton the temperature will be increased or reduced by 1°C.
4. Confirm input by .
For further information on how to operate the moisture meter please refer to the
operating instructions enclosed with each model.