Integra FIREBOY eco User manual

144952_V01_QSG_FIREBOY_EN.docx 1/4
Quick Start Guide FIREBOY plus/eco Safety Bunsen Burner
This quick start guide is intended to provide a quick overview of your FIREBOY’s (FB)
key features and to offer basic instructions for getting started. For
detailed
information, please refer to the operating instructions (OI)
that can be found at
www.integra-biosciences.com in different languages.
Intended use
This is a general-purpose laboratory instrument. Any use of this instrument in a medical or IVD
setting is the sole responsibility of the user. FIREBOY plus/eco is an automatic safety Bunsen burner
for heating or flame-sterilizing suitable laboratory materials. For continuous operation the optionally
available long burner head must be used.
Safety information
Regardless of the listed safety notes, all locally applicable regulations must be observed.
1) FIREBOY may only be used by properly trained personnel under constant supervision in a
manner specified by INTEGRA Biosciences.
2) Observe the hazard warnings on the device. There is the risk of burns from the flame and the
still hot device.
3) Do not use the FIREBOY near flammable material or in explosive areas. Work only in well
ventilated areas.
4) Use only approved and tested safety gas tubing with threaded or hose connectors.
5) Firmly tighten all gas connections and verify gas tightness, e. g. with soap water.
6) Use only original INTEGRA Biosciences power supply pack or rechargeable batteries (FB plus
only).
7) Be sure to close off the gas supply before: Transport of the device, changing the nozzle, the
adapter, the gas cartridge or the battery, longer breaks, cleaning work.
8) Servicing work and repairs may only be performed by INTEGRA Biosciences or an authorized
after-sales service member.
Getting started
Set up the instrument on a planar surface according to the IQ/OQ document (PN 144953).
Supply voltage: 100 .zH06 – 05,CAV042 –
a. Burner head
b. Motion sensor (FB plus only)
c. LCD display (FB plus only)
d. GAS KNOB, to adjust the flame size
e. AIR KNOB, to adjust the flame temperature
f. MENU KNOB (FB plus) / PUSH BUTTON (FB eco), to
switch on/off, for flame ignition or extinguishing and for
menu navigation
Switch on/off:
ON: Press (f) briefly, OFF: Press (f) for 2 sec
Settings and language selection (FIREBOY plus only)
Rotate the MENU KNOB to switch between the operating modes and select them by pressing down.
Adjustable parameters are displayed on a black background. These can be changed by rotating the
MENU KNOB and confirmed by pressing.
a
c
e
d
f
b

2/4 144952_V01_QSG_FIREBOY_EN.docx
•SENSOR: flame is ignited, if an object passes the sensor.
•BUTTON: flame is ignited by pressing the MENU KNOB.
•FOOTSWITCH: flame is ignited via the foot or benchtop switch.
•SETTINGS: To set general system parameters.
For language selection rotate the MENU KNOB and navigate to the SETTINGS menu. Select a
language and confirm it by pressing. Set other system parameters to your requirements.
Fitting the nozzle for the gas in use
FIREBOY may only be operated with the nozzle that corresponds to the gas type in use.
Type of gas
Nominal = maximal
gas input pressure
Inscription nozzle
Natural (methane) gas
20 mbar
N80
Butane/propane gas
50 mbar
P60
The P60 nozzle for butane/propane gas is fitted ex-factory. If necessary, replace the nozzle. Unscrew
the brass screw on the lower side of the device with a small coin. Turn over the device and allow the
nozzle to drop out. Insert a new nozzle and retighten the brass screw.
Adapters for the gas connection
When replacing the adapters, always close off the gas supply, remove the gas cartridge,
switch off the device and disconnect the mains pug.
•Insert the appropriate adapter in the brass gas adapter opening on
the rear side, make sure it audibly clicks into position.
•To remove the adapter, use a pointed tool to press the grey adapter
mechanism on the lower side inward and pull the adapter to the rear.
Connecting the gas cartridge
The P60 nozzle and the gas cartridge adapter must be installed. Gas cartridges may only
be used at 15 .C°53 – Only replace a gas cartridge, if the adapter is connected to the
device and closed.
Connect the cartridges as described in the enclosed instructions for the adapters.
52 g butane cartridge adapter, e.g. for CV360:
•Ensure that the flat side of the cap (a) points towards the device.
•Turn the milled knob fully clockwise in OFF direction.
•Press the gas cartridge into the cap until its audibly clicks into
position.
•Turn the milled knop fully counterclockwise in ON direction.
a

144952_V01_QSG_FIREBOY_EN.docx 3/4
Connecting a tubing
A propane/butane gas cylinder must be equipped with a pressure regulator to not exceed
the maximum input pressure of 50 mbar.
•Connect the tubing from the central gas supply or from the gas
cylinder to the previously installed adapter and fix it with a clip, if
necessary.
Electricity supply and connecting a switch
•Insert the mains plug in the socket on the rear side of the FIREBOY
and plug it to the electricity mains. (FIREBOY plus can also be
operated with the optional battery.) Switch on the device.
•Insert the plug of the foot or benchtop switch in the socket on the
rear side, if required.
Operating
There a several options for flame ignition:
FIREBOY plus
Ignition
Burning time
SENSOR
Object passes the sensor
window (two times or once)
Timer with the option to extinguish by
pressing the MENU KNOB
BUTTON
By pressing the MENU KNOB
(continuous mode)
Extinguish by pressing the MENU KNOB,
maximum burning time defined
FOOTSWITCH
By pressing the foot or
benchtop switch
As long as pressed or fix by timer with the
option to extinguish by pressing the switch
or MENU KNOB
FIREBOY eco
Ignition
Burning time
Button
By pressing the MENU KNOB
(continuous mode)
Extinguish by pressing the MENU KNOB,
maximum burning time of 60 min
Foot/benchtop
By pressing the foot or
benchtop switch
As long as pressed
Working with liquids
•In order to protect the burner chamber from contamination when
working with liquids, the folding stand at the base can be used to
incline the FIREBOY.
•The inclination direction can be modified by removing the folding
stand after pressing it together and inserting it on the opposite side.
Maintenance
Allow the burner head to cool off before cleaning the device. Switch off the device and
disconnected it form the gas supply. Allow the device to dry before operating it again.
•Clean the FIREBOY with a cloth moistened with soapy water or with a 70 % ethanol solution.
•If the burner head get soiled, it can be pulled vertically out for disinfection.
•If the FIREBOY has been contaminated, wipe it with a lint-free cloth lightly soaked with a
suitable disinfectant and wipe dry directly.
•Check the tubing frequently for brittle or porous areas. Replace it, if damaged, or at the latest 8
years after production (year imprinted on the tubing).

4/4 144952_V01_QSG_FIREBOY_EN.docx
Equipment Disposal
FIREBOY must not be disposed of with unsorted municipal waste or incinerated.
Dispose of the FIREBOY and the optional, discharged NiMH battery separately in
accordance with the regulations in your area governing disposal of devices.
Manufacturer
INTEGRA Biosciences AG
CH-7205 Zizers, Switzerland
T +41 81 286 95 30
F +41 81 286 95 33
INTEGRA Biosciences Corp.
Hudson, NH 03051, USA
T +1 603 578 5800
F +1 603 577 5529
info@integra-biosciences.com
www.integra-biosciences.com
Other manuals for FIREBOY eco
2
This manual suits for next models
2
Table of contents
Other Integra Burner manuals